SuperLine II FXS5008CRecalls & reported problems
Dentium USA · Endosseous dental implant · Class 2 · Implant
All models of SuperLine II (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 23
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI D755FXS5008C0), received between May 20, 2022 and May 13, 2026 — classified as 23 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate21
- Loss of Osseointegration2
Patient problems
- Inadequate Osseointegration23
- Failure of Implant20
- Pain3
- Unspecified Infection2
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
InjuryMay 13, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, 2 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE I. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. THE IMPLANT WAS DISLODGED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 25166348 ↗InjuryMay 13, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 3 MONTHS AFTER IMPLANT PLACEMENT. THE PATIENT'S X-RAY WAS PROVIDED. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. MOBILITY (SPINNING) WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 25166327 ↗InjuryApr 20, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 5 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 24933409 ↗InjuryApr 3, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 2 MONTHS AFTER IMPLANT PLACEMENT. PATIENT'S X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS NOT REPORTED. BONE RESORPTION WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTEVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 24774860 ↗InjuryMar 31, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 7 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. BONE RESORPTION AND PAIN WERE OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain
MAUDE report 24750081 ↗InjuryMar 24, 2026Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION APPROXIMATELY 1 YEAR AND 8 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 24678287 ↗InjuryJan 19, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 10 DAYS AFTER IMPLANT PLACEMENT. THE X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. PAIN WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain
MAUDE report 24110242 ↗InjuryJan 2, 2026Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. THE X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. ABNORMAL SENSATION AROUND IMPLANT PLACEMENT WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 23952412 ↗InjuryJan 2, 2026Loss of Osseointegration
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION AFTER 1 YEAR, APPROXIMATELY 3 YEARS AFTER IMPLANT PLACEMENT. THE X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 23952410 ↗InjuryAug 12, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 5 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS NOT REPORTED. PAIN WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain
MAUDE report 22785035 ↗InjuryJun 10, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 2 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. IMPLANT CAME OUT ON ITS OWN DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 22178398 ↗InjuryMay 1, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, AROUND 3 MONTHS AFTER IMPLANT PLACEMENT. THE X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 21946646 ↗InjuryMar 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 3 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 21627482 ↗InjuryMar 4, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 3 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. LOCAL INFECTION WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection
MAUDE report 21520970 ↗InjurySep 19, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 4 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 20261542 ↗InjuryOct 19, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 5 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS REPORTED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 17971167 ↗InjuryAug 22, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 2 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS REPORTED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Inadequate Osseointegration
MAUDE report 17599995 ↗InjuryAug 22, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS REPORTED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.
Patient problems: Inadequate Osseointegration
MAUDE report 17599787 ↗InjuryJul 5, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO POOR BONE QUALITY, 8 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 17260318 ↗InjuryApr 14, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE LESS THAN 1 MONTH AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. THE DOCTOR REPORTED BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 16748088 ↗InjuryAug 15, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 4 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED ON IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 15230418 ↗InjuryAug 15, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE IV, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED ON IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 15230415 ↗InjuryMay 20, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN ONE YEAR, AROUND 4 MONTHS AFTER IMPLANT PLACEMENT SURGERY. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS REPORTED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 14459263 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page