Dentium USA · Endosseous dental implant · Class 2 · Implant

All models of SuperLine II (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Dentium SuperLine II dental implant is an endosseous dental implant. Reports mostly describe implants failing to bond with bone and being removed, sometimes with infection or pain. There are 41 reports, no deaths and 41 injuries, and no recalls name this identifier. Reports fell from 11 to 5 in the last 12 months compared with the previous 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
0
Injuries reported
41
Recalls naming it
0

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI D755FXS5010C0), received between Aug 25, 2021 and Jun 11, 2026 — classified as 41 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (40 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 4 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate40
  • Loss of Osseointegration1

Patient problems

  • Inadequate Osseointegration41
  • Failure of Implant40
  • Unspecified Infection17
  • Pain9

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2026Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 4 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 25480310 ↗
  • InjuryJun 11, 2026Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 14 DAYS AFTER IMPLANT PLACEMENT. PATIENT'S X-RAY WAS PROVIDED. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS NOT REPORTED. LOCAL INFECTION AND PAIN WERE FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 25480284 ↗
  • InjuryMay 13, 2026Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 3 MONTHS AFTER IMPLANT PLACEMENT. THE PATIENT'S X-RAY WAS PROVIDED. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 25166325 ↗
  • InjuryFeb 13, 2026Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. NO SIGNIFICANT FINDINGS WERE OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 24339748 ↗
  • InjuryDec 22, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 22 DAYS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS NOT REPORTED. LOCAL INFECTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 23870733 ↗
  • InjuryJul 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, AROUND 3 MONTHS AFTER IMPLANT PLACEMENT. THE PATIENT'S X-RAY WAS PROVIDED. THE BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATED. LATERAL WALL GRAFT, WHICH WAS DONE PRIOR, WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 22639750 ↗
  • InjuryJul 25, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, AROUND 10 MONTHS AFTER IMPLANT PLACEMENT. THE PATIENT'S X-RAY WAS PROVIDED. THE BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 22618665 ↗
  • InjuryJul 15, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 10 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 22515601 ↗
  • InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INFECTION, 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. NO SIGNIFICANT FINDINGS WERE FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 22288343 ↗
  • InjuryMay 13, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO INFECTION, 18 DAYS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS MODERATE. LOCAL INFECTION AND PAIN WERE FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT NEEDS ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 22015379 ↗
  • InjuryMay 6, 2025Loss of Osseointegration; Loss of Osseointegration

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO LOSS OF OSSEOINTEGRATION, 2 YEARS AND 1 MONTH AFTER IMPLANT PLACEMENT. X-RAY WAS PROVIDED. BONE QUALITY AROUND THE AREA WAS TYPE III, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21973017 ↗
  • InjuryApr 3, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 3 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 21769905 ↗
  • InjuryMar 31, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. NO SIGNIFICANT FINDING WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 21735501 ↗
  • InjuryDec 31, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 2 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 21050439 ↗
  • InjuryDec 11, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 1 MONTH AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE II. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. NO SIGNIFICANT FINDING WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20909153 ↗
  • InjuryNov 19, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR, APPROXIMATELY 2 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS NOT REPORTED. LOCAL INFECTION WAS FOUND DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 20727580 ↗
  • InjuryOct 14, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. PAIN WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain

    MAUDE report 20442546 ↗
  • InjurySep 27, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IT WAS RECORDED THAT THE IMPLANT WAS PLACED ON (B)(6) 2011 AND EXPLANTED ON (B)(6) 2023. HOWEVER, THE RECORD DID NOT MATCH ANY SALES HISTORY OF THE DEVICE. THE OFFICE WAS CONTACTED TO GET MORE INFORMATION ON THE SURGERY INFORMATION. HOWEVER, IT WAS NOT ABLE TO BE EXPLAINED. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. LOCAL INFECTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 20327943 ↗
  • InjurySep 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. BONE RESORPTION WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20284175 ↗
  • InjurySep 18, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 6 MONTHS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 20247700 ↗
  • InjuryMay 1, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 7 DAYS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE II, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. LOCAL INFECTION AND PAIN WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 19226885 ↗
  • InjuryApr 9, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, AROUND 4 MONTHS AFTER IMPLANT PLACEMENT. THE BONE QUALITY WAS TYPE III. THE ORAL HYGIENE AROUND IMPLANT SITE WAS NOT REPORTED. PERI-IMPLANTITIS WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 19076660 ↗
  • InjuryApr 9, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE WITHIN 1 YEAR AND INSUFFICIENT PRIMARY STABILITY, 11 DAYS AFTER IMPLANT PLACEMENT. THE BONE QUALITY (TYPE) WAS NOT REPORTED. THE ORAL HYGIENE AROUND IMPLANT SITE WAS GOOD. NO SIGNIFICANT FINDING WAS OBSERVED DURING THE REMOVAL SURGERY. BONE AUGMENTATION MATERIAL WAS NOT USED IN IMPLANT PLACEMENT SURGERY. THE PATIENT HAS SINCE RECOVERED.

    Patient problems: Failure of Implant; Inadequate Osseointegration

    MAUDE report 19076654 ↗
  • InjuryApr 2, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 6 DAYS AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. PAIN AND LOCAL INFECTION WERE OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection

    MAUDE report 19028876 ↗
  • InjuryApr 2, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE DOCTOR REPORTED THE IMPLANT WAS REMOVED DUE TO OSSEOINTEGRATION FAILURE, 1 MONTH AFTER IMPLANT PLACEMENT. BONE QUALITY AROUND THE AREA WAS TYPE I, AND ORAL HYGIENE AROUND THE IMPLANT WAS MODERATE. PAIN WAS OBSERVED DURING THE IMPLANT REMOVAL SURGERY. THE PATIENT WILL NEED ANOTHER SURGICAL INTERVENTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain