Prima Implant System 15627KRecalls & reported problems
Keystone Dental, Inc. · Implant · Class 2 · Implant
All models of Prima Implant System (58)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI D76815627K0), received between Feb 3, 2020 and Oct 18, 2024 — classified as 16 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (12 reports).
7 other recalls cover devices cleared under the same FDA submission (K051614, K072768); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate12
- Loss of Osseointegration4
Patient problems
- Inadequate Osseointegration8
- Osteopenia/ Osteoporosis7
- Failure of Implant3
- Bruxism2
- Implant Pain2
- Pain2
- Unspecified Infection1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryOct 18, 2024Failure to Osseointegrate; Failure to Osseointegrate
InjuryFeb 8, 2024Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION.
Patient problems: Implant Pain; Inadequate Osseointegration; Osteopenia/ Osteoporosis
MAUDE report 18667145 ↗InjuryJan 9, 2024Failure to Osseointegrate; Failure to Osseointegrate
FAILED TO OSSEOINTEGRATE.
Patient problems: Bruxism; Inadequate Osseointegration
MAUDE report 18479521 ↗InjuryAug 11, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION.
Patient problems: Bruxism; Implant Pain; Inadequate Osseointegration
MAUDE report 17511137 ↗InjuryMay 26, 2023Failure to Osseointegrate; Failure to Osseointegrate
FAILED TO OSSEOINTEGRATE.
Patient problems: Inadequate Osseointegration; Osteopenia/ Osteoporosis
MAUDE report 17009448 ↗InjuryFeb 2, 2023Failure to Osseointegrate; Failure to Osseointegrate
FAILED TO OSSEOINTEGRATE.
Patient problems: Inadequate Osseointegration; Osteopenia/ Osteoporosis
MAUDE report 16287497 ↗InjuryJan 18, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION.
Patient problems: Inadequate Osseointegration; Osteopenia/ Osteoporosis
MAUDE report 16195032 ↗InjuryAug 22, 2022Failure to Osseointegrate; Failure to Osseointegrate
FAILED TO OSSEOINTEGRATE.
Patient problems: Failure of Implant; Inadequate Osseointegration
MAUDE report 15267156 ↗InjuryFeb 1, 2022Failure to Osseointegrate; Failure to Osseointegrate
InjuryDec 29, 2021Failure to Osseointegrate; Failure to Osseointegrate
InjuryNov 11, 2021Failure to Osseointegrate; Failure to Osseointegrate
InjuryMay 10, 2021Failure to Osseointegrate; Failure to Osseointegrate
InjuryMar 18, 2021Failure to Osseointegrate; Failure to Osseointegrate
InjuryMar 4, 2020Failure to Osseointegrate; Failure to Osseointegrate
InjuryFeb 5, 2020Loss of Osseointegration; Loss of Osseointegration
InjuryFeb 3, 2020Failure to Osseointegrate; Failure to Osseointegrate
Recalls under the same FDA submission
These recalls involve devices cleared under K051614, K072768. One submission often covers many models, so check the product and lot details in each record.
- Z-2162-2018 · May 1, 2018 · TerminatedPrima Plus 4.1 x 11.5mm Implant, Catalog Number 15737K Keystone Dental implants are intended for use in partially or fully edentulous mandibles and maxillae, in support of single
- Z-1179-2018 · Oct 16, 2017 · TerminatedSTAGE-1 RDS COC Abutment Plastic Sleeve, Non-Lock, Multi-Unit; Catalog Number: S2437-01-1K
- Z-0727-2018 · Aug 25, 2017 · TerminatedStage-1 RDS Indexed Abutment for RDS Stage-1 Single Stage Implants Keystone s implants are intended for use in partially or fully edentulous mandibles and maxillae, in support
- Z-0608-2018 · Jun 20, 2017 · TerminatedMulti-Unit Abutment Angled, 17, 4mm Cuff 4.1/4.5mm TiLobe Implants, Catalog Number 45324K pRODUCT Usage: Use in partially or fully edentulous mandibles and maxillae, in support of
- Z-1818-2017 · Mar 9, 2017 · TerminatedKeystone Dental Inc. PrimaConnex Straight Implant TC, WD Internal Connection 5.0 x 11.5mm. Catalog Number(s): 15636K
- Z-0299-2017 · Sep 15, 2016 · TerminatedKeystone Dental PrimaConnex Straight Implant TC, RD Internal Connection 4.0 x 11.5mm. Product code:15613K Implants are intended for use in partially or fully edentulous mandibles a
- Z-0973-2013 · Mar 11, 2013 · TerminatedPrimaConnex Tapered Implant TC, SD Internal Connection 3.5 x 10mm Catalog Number: l5613K. Implants in support for fixed bridgework.
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page