Skip to content

iSy® Implants-Set P1410.4411Recalls & reported problems

ALTATEC GmbH · P1410.4411 iSy® Implants-Set, 4 units d 4.4 mm, L11 mm · Class 2 · Implant

All models of iSy Implants-Set (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI E219P141044111), received between Jun 26, 2024 and Jul 8, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate3

Patient problems

  • Insufficient Information3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 8, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE

    Patient problems: Insufficient Information

    MAUDE report 25748667 ↗
  • InjuryDec 9, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 20877959 ↗
  • InjuryJun 26, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 19619350 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page