Skip to content

Piccolo Comprehensive Metabolic Panel 400-0028Recalls & reported problems

Abaxis, Inc. · The Piccolo® Comprehensive Metabolic Panel, used with the Piccolo® blood chemistry analyzer or the Piccolo Xpress® chemistry analyzer, is intended to be used for the in vitro quantitative determination of alanine aminotransferase (ALT), albumin, alkaline phosphatase (ALP), aspartate aminotransferase (AST), calcium, chloride, creatinine, glucose, potassium, sodium, total bilirubin, total carbon dioxide, total protein, and blood urea nitrogen (BUN) in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location. The tests on this panel are waived under CLIA ’88 regulations. If a laboratory modifies the test system instructions, then the tests are considered high complexity and subject to all CLIA requirements. For CLIA waived labs, only lithium heparin whole blood may be tested. For use in moderate complexity labs, lithium heparinized whole blood, lithium heparinized plasma, or serum may be used. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI EABA40000281), received between Dec 13, 2019 and Sep 27, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “High Readings” (3 reports).

5 other recalls cover devices cleared under the same FDA submission (K010670, K934592, K942782, K950164, K992140, K993211); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Readings3
  • Low Readings3
  • Device Contamination with Body Fluid1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • Electrolyte Imbalance1
  • Hematuria1
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 27, 2024High Readings; High Readings

    (B)(6) 2024: ZOETIS ABAXIS UNION CITY TECHNICAL SUPPORT RECEIVED A CALL FROM A LABORATORY HEALTH PROFESSIONAL REPORTING HIGH POTASSIUM (K+) USING THE COMPREHENSIVE METABOLIC PANEL (CMP) LOT 4251AC2 ON THE PICCOLO XPRESS ANALYZER SERIAL NUMBER (B)(6). THE PATIENT WAS TREATED AND ADMITTED TO THE HOSPITAL FOR HYPERKALEMIA AND RHABDOMYOLYSIS.

    Patient problems: Electrolyte Imbalance; Hematuria

    MAUDE report 20323498 ↗
  • MalfunctionSep 12, 2024Low Readings; Low Readings

    ON 15 AUGUST 2024, ZOETIS ABAXIS UNION CITY TECHNICAL SUPPORT RECEIVED A CALL FROM A HEALTHCARE PROFESSIONAL REPORTING AN UNEXPECTED HIGH POTASSIUM (K+) RESULT USING THE PICCOLO XPRESS ANALYZER, SERIAL NUMBER (B)(6), AND THE PICCOLO COMPREHENSIVE METABOLIC PANEL (CMP), LOT 4251AC2. THE PATIENT, A 21-YEAR-OLD MALE, WAS TREATED BASED ON THE RESULTS OF THE PICCOLO XPRESS CHEMISTRY ANALYZER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20207488 ↗
  • MalfunctionApr 30, 2024Low Readings; Low Readings

    18 APR 2024: ABAXIS UNION CITY TECHNICAL SUPPORT RECEIVED A CALL FROM A LABORATORY HEALTH PROFESSIONAL REPORTING LOW POTASSIUM (K+) USING THE COMPREHENSIVE METABOLIC PANEL (CMP) ON THE PICCOLO XPRESS ANALYZER. THE PATIENT WAS TREATED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19216258 ↗
  • MalfunctionNov 23, 2022High Readings; High Readings

    A LABORATORY HEALTHCARE PROFESSIONAL REPORTED UNEXPECTED HIGH CREATININE (CRE) RESULTS. PER REPORT, A PATIENT WAS TREATED BASED ON SYMPTOMS OF APPENDICITIS - NOT DUE TO THE PICCOLO XPRESS ANALYZER RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15854127 ↗
  • MalfunctionJun 14, 2022Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; High Readings; Output Problem

    THE CUSTOMER A HEALTHCARE PROFESSIONAL REPORTED UNEXPECTED HIGH SODIUM (NA+). THE PATIENT WAS TREATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14686943 ↗
  • MalfunctionMar 20, 2020Device Contamination with Body Fluid; Device Contamination with Body Fluid

    CUSTOMER SAYS THEY RECEIVED REAGENT DISCS THAT APPEAR ALREADY USED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9862762 ↗
  • MalfunctionDec 13, 2019Low Readings; Low Readings

    LOW POTASSIUM (K+) PATIENT RESULTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9469240 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K010670, K934592, K942782, K950164, K992140, K993211. One submission often covers many models, so check the product and lot details in each record.

  • Z-2361-2020 · Dec 11, 2019 · Terminated
    Piccolo Xpress chemistry analyzer, Model No. 1100-1001, Software version 2.1.55 (NGA)/3.1.35(Jen II)
  • Z-1066-2011 · Nov 19, 2010 · Terminated
    Abaxis brand Piccolo Comprehensive Metabolic Panel; Used with the Piccolo Blood Chemistry Analyzer or Piccolo xpress Chemistry Analyzer; Catalog/Part Number: 400-0028; Product is
  • Z-1155-2010 · Nov 16, 2009 · Terminated
    Piccolo Comprehensive Metabolic Panel Reagent Disc, 10-pack box, for use with the Piccolo Blood Chemistry Analyzer or Piccoloxpress Chemistry Analyzer, Catalog No. 400-0028, manufa
  • Z-2287-2008 · Apr 21, 2008 · Terminated
    Piccolo Chemistry Analyzer software version 6.121 with Electrolyte Panel Lot #8041AC5, model number 400-1022 in polyfoil pouch. Piccolo Electrolyte Panel is a single-use, self-c
  • Z-2286-2008 · Apr 21, 2008 · Terminated
    Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Number 400-1028 in polyfoil pouch. Piccolo Comprehensive Metabolic Panel is a s

Other Mercuric Thiocyanate, Colorimetry, Chloride devices

All Mercuric Thiocyanate, Colorimetry, Chloride devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page