FRIALIT® 32362461Recalls & reported problems
Dentsply Implants Manufacturing GmbH · FRIALIT® plus Implant D5.5/L10 · Class 2 · Implant
All models of FRIALIT (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI EFRI323624612), received between Sep 12, 2023 and Apr 22, 2026 — classified as 4 injuries.
The most frequently reported specific device problem is “Loss of Osseointegration” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration4
Patient problems
- Failure of Implant4
- Osteolysis2
- Pocket Erosion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 22, 2026Loss of Osseointegration
IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.
Patient problems: Failure of Implant; Osteolysis
MAUDE report 24960311 ↗InjuryMar 6, 2026Loss of Osseointegration
IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS AND PERIIMPLANTITIS.
Patient problems: Failure of Implant
MAUDE report 24536746 ↗InjuryJul 11, 2025Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.
Patient problems: Failure of Implant; Osteolysis; Pocket Erosion
MAUDE report 22487381 ↗InjurySep 12, 2023Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.
Patient problems: Failure of Implant
MAUDE report 17728713 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page