EverPatch+ 1044002Recalls & reported problems
Corneat Vision Ltd · Permanent Scleral Reinforcement Matrix. Two shapes (one rectangular, one shield-shaped) are provided in each unit, to be selected according to physician preference. · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI G16010440020), received between Apr 21, 2025 and Aug 22, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Improper or Incorrect Procedure or Method1
Patient problems
- Wound Dehiscence2
- Erosion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 22, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
CORNEAT RECEIVED INFORMATION FROM AN UNINVOLVED THIRD-PARTY SUGGESTING THAT A COLLEAGUE HAD IMPLANTED 3 EVERPATCH+ DEVICES AND OBSERVED WOUND DEHISCENCE IN ALL 3 CASES. ADDITIONAL INFORMATION WAS REQUESTED FROM THE THIRD PARTY AND THIS ENABLED CORNEAT TO IDENTIFY THE IMPLANTING SURGEON WHO WAS LATER CONTACTED. THE IMPLANTING SURGEON CONFIRMED OBSERVING WOUND DEHISCENCE IN 3 EVERPATCH+ CASES, CLARIFYING THAT THESE PATIENTS SELECTED FOR HIS FIRST USE OF THE DEVICE WERE SOME OF HIS WORST CANDIDATES IN TERMS OF THEIR OCULAR SURFACE HEALTH. THE SURGEON DESCRIBED HIS SURGICAL TECHNIQUE AS A LIMBAL-BASED APPROACH WITH THE SURGICAL INCISION 5 MM POSTERIOR TO THE LIMBUS. IMPORTANTLY, ALL SURGICAL WOUNDS WERE CLOSED DIRECTLY OVER THE EVERPATCH+, DEVIATING FROM THE EVERPATCH+ INSTRUCTIONS FOR USE WHICH SPECIFIES "THE DEVICE SHOULD NOT BE IMPLANTED DIRECTLY UNDER A CONJUNCTIVAL INCISION". TENON WAS NOT CLOSED INDEPENDENTLY, BUT RATHER SECURED TO THE PERIPHERAL SCLERA. TWO CASES EXHIBITED EXPOSURE IMMEDIATELY DUE TO WOUND DEHISCENCE AND ONE EXPOSURE WAS OBSERVED 6 WEEKS POST IMPLANTATION. TWO PATIENTS UNDERWENT SURGICAL REVISION AND REVISION SURGERY IS PLANNED FOR THE THIRD PATIENT. CORNEAT REQ
Patient problems: Wound Dehiscence
MAUDE report 22870084 ↗InjuryApr 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2025, A PATIENT WITH GLAUCOMA UNDERWENT IMPLANTATION OF THE AHMED GLAUCOMA VALVE IN THE RIGHT EYE AND THE EVERPATCH+ WAS USED TO COVER THE TUBE. A FORNIX-BASED FLAP WAS CREATED AND THE GLAUCOMA VALVE WAS IMPLANTED IN THE SUPERIOR TEMPORAL QUADRANT; CAUTERY WAS USED INTRAOPERATIVELY TO CONTROL BLEEDING AND MINIMIZE SCARRING. SURGERY WAS UNEVENTFUL AND THE WOUND WAS CLOSED WITH VICRYL CONTINUOUS SUTURES. TOPICAL STEROIDS WERE PRESCRIBED POSTOPERATIVELY (PER STANDARD OF CARE). WITHIN THE FIRST POSTOPERATIVE MONTH, THE PATIENT PRESENTED WITH EXPOSURE OF THE CONJUNCTIVA OVER THE EVERPATCH+; HOWEVER, THERE WAS NO WOUND LEAK (I.E., SEIDEL NEGATIVE). ON (B)(6) 2025, TUBE REVISION SURGERY WAS PERFORMED WHERE THE EVERPATCH+ WAS EXPLANTED AND REPLACED WITH A TUTOPLAST. DURING EXPLANTATION, THE EVERPATCH+ WAS NOT WELL ADHERENT AND WAS EASILY REMOVED. THE SURGEON SUSPECTS THAT THE PATIENT'S VERY THIN CONJUNCTIVA (NO TENON'S) LIKELY CONTRIBUTED TO THE CONJUNCTIVAL EROSION. THE PATIENT HAS NOT EXPERIENCED A LOSS OF 2 OR MORE LINES OF BEST CORRECTED VISUAL ACUITY (BCVA) COMPARED TO BASELINE.
Patient problems: Erosion; Wound Dehiscence
MAUDE report 21877966 ↗
Other Prosthesis, Eyelid Spacer/Graft, Polymer devices
- CorNeat EverPatch 1007691Corneat Vision Ltd · Prosthesis, Eyelid Spacer/Graft, Polymer13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page