DuraFlow 10301202Recalls & reported problems
Angiodynamics, Inc. · DuraFlow 2 - 15.5F x 28cm Chronic Hemodialysis Catheter Set (with Cuff 23cm from Tip) · Class 2 · Implant
All models of DuraFlow (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H787103012020), received between Jan 28, 2022 and Jan 28, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K110936); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 18, 2024Class IIOpen, Classified
Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
Z-1761-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 28, 2022Fracture; Fracture
AN OR MANAGER REPORTED AN ISSUE WITH A DURAFLOW 2, 28CM STRAIGHT BASIC KIT. DURING A LAPAROSCOPIC PERITONEAL DIALYSIS CATHETER INSERTION PROCEDURE, THE SURGEON ATTEMPTED TO PULL THE GUIDEWIRE OUT OF THE VESSEL AND WAS MET WITH RESISTANCE DUE TO A KINK. WHEN THE GUIDEWIRE WAS PULLED OUT, IT BECAME UNRAVELLED AND A PIECE FELL OFF INSIDE OF THE PATIENT. THE SURGEON WAS ABLE TO REMOVE THE FRAGMENT FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A NEW KIT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. TO NOTE, THE SURGEON WAS LATER INFORMED BY AN ANGIODYNAMICS REP THAT THIS DEVICE IS FOR HEMODIALYSIS USE ONLY, NOT PERITONEAL DIALYSIS.
Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions
MAUDE report 13385468 ↗
FDA clearance and classification
Cleared through 510(k) K110936 on Jun 10, 2011. Submitted by AngioDynamics, Inc. as “Duraflow 2 Chronic Hemodialysis Catheter And Procedure Kit”. The FDA review took 67 days. FDA record ↗
The FDA regulates product code MSD (catheter, hemodialysis, implanted) under 21 CFR 876.5540 as a Class II device. 21 CFR 876.5540 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Recalls under the same FDA submission
These recalls involve devices cleared under K110936. One submission often covers many models, so check the product and lot details in each record.
- Z-1764-2024 · Mar 18, 2024 · Open, ClassifiedDURAFLOW 2, 24CM STRAIGHT VASC-PAK PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012185
- Z-1763-2024 · Mar 18, 2024 · Open, ClassifiedDURAFLOW 2, 55CM STRAIGHT BASIC KIT PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012115
- Z-1760-2024 · Mar 18, 2024 · Open, ClassifiedDURAFLOW 2, 24CM STRAIGHT BASIC KIT -Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012015
- Z-1765-2024 · Mar 18, 2024 · Open, ClassifiedDURAFLOW 2, 28CM STRAIGHT VASC-PAK PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012195
- Z-1762-2024 · Mar 18, 2024 · Open, ClassifiedDURAFLOW 2, 32CM STRAIGHT BASIC KIT PG-Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012035
Common questions about the DuraFlow 10301202
Has the DuraFlow 10301202 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1761-2024 (Class II), began on Mar 18, 2024: Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the DuraFlow 10301202?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 28, 2022. That report lists device problem “Fracture” (1); patient problem “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DuraFlow 10301202?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DuraFlow 10301202 FDA approved?
It was cleared through 510(k) K110936 on Jun 10, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the DuraFlow 10301202 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the DuraFlow 10301202, and is it still sold?
The FDA's device database lists Angiodynamics, Inc. as the company responsible for it (brand DuraFlow, model 10301202). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths293 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death142 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page