DuraMax 10302804Recalls & reported problems
Angiodynamics, Inc. · 15.5F x 28cm Chronic Hemodialysis Catheter Set (with Cuff 23cm from Tip) · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI H787103028040), received between Jan 31, 2023 and Jun 22, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
29 other recalls cover devices cleared under the same FDA submission (K101843); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 18, 2024Open, Classified
Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
Z-1768-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Deformation Due to Compressive Stress2
- Fracture1
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 22, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
AN END USER EXPERIENCED WITH A CATHETER FROM A DURAMAX STACKED TIP 28CM STR. BASIC KIT. WHEN THE PATIENT PRESENTED FOR A DIALYSIS TREATMENT USING THE IMPLANTED CATHETER, THE PROVIDER WAS UNABLE TO PERFORM DIALYSIS AS THE CLAMP KINKED THE CATHETER AND DIALYSIS WAS UNABLE TO BE RUN. THE PATIENT HAD TO RETURN TO THE HOSPITAL TO HAVE THE CATHETER REPLACED. ONCE REPLACED, THE PATIENT WAS ABLE TO RETURN TO THE DIALYSIS CENTER AND HAVE THE FULL TREATMENT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM DUE TO THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17184940 ↗MalfunctionJun 22, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
AN END USER EXPERIENCED WITH A CATHETER FROM A DURAMAX STACKED TIP 28CM STR. BASIC KIT. WHEN THE PATIENT PRESENTED FOR A DIALYSIS TREATMENT USING THE IMPLANTED CATHETER, THE PROVIDER WAS UNABLE TO PERFORM DIALYSIS AS THE CLAMP KINKED THE CATHETER AND DIALYSIS WAS UNABLE TO BE RUN. THE PATIENT HAD TO RETURN TO THE HOSPITAL TO HAVE THE CATHETER REPLACED. ONCE REPLACED, THE PATIENT WAS ABLE TO RETURN TO THE DIALYSIS CENTER AND HAVE THE FULL TREATMENT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM DUE TO THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17184930 ↗MalfunctionApr 19, 2023Fracture; Fracture
A DISTRIBUTOR REPORTED AN END USER EXPERIENCED AN ISSUE WITH SEVERAL DURAMAX CATHETERS. PER THE USER, THEY HAVE RECENTLY HAD EIGHT INCIDENTS OF BROKEN CLAMPS OF PERMACATH CATHETERS. ALL INCIDENTS OCCURRED WITH THE SAME SCENARIO. THE CLAMPS WERE BROKEN OUTSIDE THE PATIENTS RESULTING IN THE CATHETER HAVING TO BE REMOVED AND REPLACED. THE PATIENTS DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM DUE TO THESE INCIDENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16771259 ↗MalfunctionJan 31, 2023Infusion or Flow Problem; Infusion or Flow Problem
A TERRITORY MANAGER REPORTED AN END USER EXPERIENCED AN ISSUE WHEN USING A CATHETER FROM A DURAMAX STACKED TIP 28CM STR. BASIC KIT. IT WAS REPORTED THE CATHETER'S BLUE PORT HAD A SLOW FLOW AND RESISTANCE. AS IT WAS INDICATED THE DEVICE IS AVAILABLE FOR RETURN, IT IS REASONABLE TO CONCLUDE THE CATHETER WAS REMOVED DUE TO THIS ISSUE. IT IS UNKNOWN IF IT WAS REPLACED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION BECAUSE OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16273418 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K101843. One submission often covers many models, so check the product and lot details in each record.
- Z-1578-2026 · Feb 13, 2026 · Open, ClassifiedDuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028045/A, H787103028051/A, H787103028055/A,
- Z-1767-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 24CM STR. BASIC KIT, 5-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028035
- Z-1766-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 24CM STR. BASIC KIT, 1-UP PG Product Number: H787103028031
- Z-1770-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 32CM STR. BASIC KIT, 5-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028055
- Z-1773-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 32 CM STRAIGHT VASC-PAK, 1-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028201
- Z-1769-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 28CM STR. BASIC KIT, 5-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028045
- Z-1771-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 28 CM STRAIGHT VASC-PAK, 1-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028191
- Z-1772-2024 · Mar 18, 2024 · Open, ClassifiedDURAMAX STACKED TIP 28 CM STRAIGHT VASC-PAK, 5-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028195
- Z-0731-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax 22 cm STRAIGHT --- 1-15.5F x 22cm DuraMax Hemodialysis Catheter (with Cuff 17cm from Tip), 2- Injection Caps --- Catalog No. [REF] 10302902 --- MANUFACTURED I
- Z-0727-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax VP Vascpak Catheter Kit 15.5F x 28cm Chronic Hemodialysis VascPak (with Cuff 23cm from Tip) --- Catalog No. [REF] 10302819 --- MANUFACTURED IN USA AngioDynami
- Z-0719-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax 15.5F x 22 cm Chronic Hemodialysis Catheter Set (with Cuff 17cm from Tip) --- Catalog No. [REF] 10302802 --- MANUFACTURED IN USA AngioDynamics, Inc., 603 Quee
- Z-0721-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax 15.5F x 28 cm Chronic Hemodialysis Catheter Set (with Cuff 23cm from Tip) --- Catalog No. [REF] 10302804 --- MANUFACTURED IN USA AngioDynamics, Inc., 603 Quee
- Z-0733-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax 55 cm STRAIGHT --- 1-15.5F x 55cm DuraMax Hemodialysis Catheter (with Cuff 50cm from Tip), 2- Injection Caps --- Catalog No. [REF] 10302909 --- MANUFACTURED I
- Z-0729-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax VP Vascpak Catheter Kit 15.5F x 40cm Chronic Hemodialysis VascPak (with Cuff 35cm from Tip) --- Catalog No. [REF] 10302822 --- MANUFACTURED IN USA AngioDynami
- Z-0720-2012 · Dec 30, 2011 · TerminatedAngiodynamics DuraMax 15.5F x 24 cm Chronic Hemodialysis Catheter Set (with Cuff 19cm from Tip) --- Catalog No. [REF] 10302803 --- MANUFACTURED IN USA AngioDynamics, Inc., 603 Quee
…and 14 more.
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.