BioFlo 02-805Recalls & reported problems
Navilyst Medical, Inc. · 15.5Fx32cm BioFlo DuraMax with ENDEXO Technology Chronic Hemodialysis Catheter Basic Kit · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI H965103028050), received between Oct 16, 2020 and Mar 30, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Crack” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Crack3
- Gas/Air Leak2
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2023Crack; Gas/Air Leak; Crack; Gas/Air Leak
A CLINICAL SPECIALIST MANAGER REPORTED A DISTRIBUTOR HAD BEEN NOTIFIED OF SIX EVENTS ASSOCIATED WITH BIOFLO DURAMAX DIALYSIS CATHETERS. THEY RELATED THAT SOME OF THE DIALYSIS SITES WHERE PATIENTS PRESENT IN VIENNA AUSTRIA HAVE EXPERIENCED "CRACKED/BROKEN HUBS" AND "AIR IN THE BLOODLINE" SUCH THAT THE CATHETERS HAD TO BE REPLACED. THEY INDICATED THERE HAVE BEEN 6 CATHETERS WITH THIS PROBLEM. NONE OF THE PATIENTS EXPERIENCED ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THE INCIDENTS. IT WAS NOTED THAT THESE INSTANCES ONLY OCCURRED AT SITES AWAY FROM THE MAIN FACILITY WHERE THE PATIENTS RECEIVE HEMODIALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16646054 ↗MalfunctionMar 30, 2023Crack; Gas/Air Leak; Crack; Gas/Air Leak
A CLINICAL SPECIALIST MANAGER REPORTED A DISTRIBUTOR HAD BEEN NOTIFIED OF SIX EVENTS ASSOCIATED WITH BIOFLO DURAMAX DIALYSIS CATHETERS. THEY RELATED THAT SOME OF THE DIALYSIS SITES WHERE PATIENTS PRESENT IN VIENNA AUSTRIA HAVE EXPERIENCED "CRACKED/BROKEN HUBS" AND "AIR IN THE BLOODLINE" SUCH THAT THE CATHETERS HAD TO BE REPLACED. THEY INDICATED THERE HAVE BEEN 6 CATHETERS WITH THIS PROBLEM. NONE OF THE PATIENTS EXPERIENCED ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THE INCIDENTS. IT WAS NOTED THAT THESE INSTANCES ONLY OCCURRED AT SITES AWAY FROM THE MAIN FACILITY WHERE THE PATIENTS RECEIVE HEMODIALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16645932 ↗MalfunctionOct 16, 2020Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
A DISTRIBUTOR REPRESENTATIVE REPORTED TO AN ANGIODYNAMICS REPRESENTATIVE AN ONGOING ISSUE AN END USER HAS BEEN EXPERIENCING WITH BIOFLO DIALYSIS CATHETERS. PER THE DISTRIBUTOR: YESTERDAY I WAS VISITING A CUSTOMER WHO HAS PROBLEMS WITH THE BIOFLO DURAMAX CATHETER. THEY HAVE HAD SEVERAL INCIDENTS WITH LEAKING CATHETERS. THE LEAK WAS OFTEN NOT VISIBLE - SOME WERE HAIRLINE CRACKS OTHER HOLES COULD BE SEEN, ALWAYS NEXT TO THE LUER. THE LEAKS WERE DISCOVERED BECAUSE BLOOD LEAKED AND OTHER TIMES THE DIALYSIS MACHINE ALARMED BECAUSE AIR WAS ASPIRATED INTO THE EXTENSION LEG TUBING. IN SEVERAL CATHETERS THE LEAK HAS BEEN REPAIRED (INSTEAD OF CHANGED). HOWEVER, THE DEFECTIVE CATHETERS HAVE NOT BEEN SAVED, SO I CANNOT SEND THEM TO YOU. WE HAVE NO LOT/BATCH NUMBERS AVAILABLE. THERE HAS BEEN NO REPORTS OF PATIENT HARM OR INJURIES DUE THESE REPORTED MALFUNCTIONS. IT IS REASONABLE TO CONCLUDE THE DEVICES WERE REMOVED AND REPLACED IN ORDER TO CONTINUE TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10690643 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page