BioFlo 03-819Recalls & reported problems
Navilyst Medical, Inc. · 15.5Fx28cm BioFlo DuraMax with ENDEXO Technology Chronic Hemodialysis Catheter VascPak Kit · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI H965103038190), received between Jun 12, 2020 and Jan 13, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Fluid/Blood Leak1
- Fracture1
- Material Puncture/Hole1
Patient problems
- Device Embedded In Tissue or Plaque4
- Hemorrhage/Blood Loss/Bleeding1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
1ST DEVICE, SAME PATIENT: A SENIOR MANAGER OF PATIENT SAFETY REPORTED AN ISSUE WITH A BIOFLO DIALYSIS CENTRAL VENOUS CATHETER. THE FOLLOWING WAS REPORTED: "CENTRAL VENOUS CATHETER PLACED ON (B)(6) 2022. ON (B)(6) 2022, CATHETER WAS LEAKING AT THE ARTERIAL LIMB AND REPAIR OF CVC TOOK PLACE ON (B)(6) 2022. (B)(4). PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS/HARM AS A RESULT OF THIS ISSUE (BLOOD LOSS, PROLONGED HOSPITALIZATION ETC. PT PRESENTED TO HEMODIALYSIS FACILITY ON (B)(6) 20/22 FOR REGULARLY SCHEDULED TREATMENT. TREATMENT INITIATED AT 557 AM. TREATMENT WAS UNEVENTFUL UNTIL 8:56 AM WHEN PATIENT ALERTED STAFF HE WAS BLEEDING. STAFF NOTED THE RED HUB OF THE CVC HAD DETACHED FROM THE LUMEN. TREATMENT WAS STOPPED. (B)(4). PT WAS TRANSPORTED TO THE HOSPITAL WHERE THE CLAMP AND HUB OF THE CVC WERE REPLACED. THE PATIENT LATER UNDERWENT A CATHETER EXCHANGE ON (B)(6) 2022".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16157057 ↗MalfunctionJan 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
2ND DEVICE, SAME PATIENT: A SENIOR MANAGER OF PATIENT SAFETY REPORTED AN ISSUE WITH A BIOFLO DIALYSIS CENTRAL VENOUS CATHETER. THE FOLLOWING WAS REPORTED: "- CENTRAL VENOUS CATHETER PLACED (B)(6) 2022. - REPAIR OF CVC TOOK PLACE 9/27/22 (B)(4). - PT PRESENTED TO HEMODIALYSIS FACILITY ON (B)(6) 2022 FOR REGULARLY SCHEDULED TREATMENT. TREATMENT INITIATED AT 557 AM. - TREATMENT WAS UNEVENTFUL UNTIL 856 AM WHEN PATIENT ALERTED STAFF HE WAS BLEEDING. - STAFF NOTED THE RED HUB OF THE CVC HAD DETACHED FROM THE LUMEN. TREATMENT WAS STOPPED. (B)(4). - PT WAS TRANSPORTED TO THE HOSPITAL WHERE THE CLAMP AND HUB OF THE CVC WERE REPLACED. - THE PATIENT LATER UNDERWENT A CATHETER EXCHANGE ON (B)(6) 2022 -PRE-EVENT HEMOGLOBIN 10.8 G/DL, POST EVENT HEMOGLOBIN 8.9 G/DL"
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 16157056 ↗MalfunctionDec 6, 2021Fracture; Fracture
AN END USER REPORTED AN ISSUE WITH A BIOFLO CHRONIC DIALYSIS CATHETER 15.5F 28CM SINGLE VALVE SHEATH VASCPAK KIT. DURING A PROCEDURE, THE PHYSICIAN WAS TRYING TO PLACE THE DEVICE IN THE PATIENT AND EXPERIENCED ISSUES WITH THE INTRODUCER AND THE WHITE PLASTIC TRANSITION PIECE, FROM THE TUNNELER TO THE CATHETER. THE INTRODUCER SHEATH BUCKLED; THEREFORE, ANOTHER SAME DEVICE WAS OPENED. THIS PART OF THE DEVICE, FROM THE SECOND KIT, BROKE INTO A FEW PIECES INSIDE THE PATIENT; HOWEVER, THE PHYSICIAN WAS ABLE TO FISH THEM OUT AND THEN USE A THIRD KIT TO PLACE THE CATHETER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 12933179 ↗MalfunctionJun 12, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
EVENT #2 AN IMAGING SUPERVISOR REPORTED THAT THE TUNNELER SLEEVE COMPONENT OF THIS BIOFLO CHRONIC DIALYSIS CATHETER SLID OFF THE TUNNELER AND BECOME DETACHED IN THE PATIENT WHILE TUNNELING WAS BEING PERFORMED. THE SLEEVE WAS ABLE TO BE REMOVED AFTER IT WAS NOTICED TO BE MISSING FROM THE TUNNELER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10148790 ↗MalfunctionJun 12, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
EVENT #3: AN IMAGING SUPERVISOR REPORTED THAT THE TUNNELER SLEEVE COMPONENT OF THIS BIOFLO CHRONIC DIALYSIS CATHETER SLID OFF THE TUNNELER AND BECOME DETACHED IN THE PATIENT WHILE TUNNELING WAS BEING PERFORMED. THE SLEEVE WAS ABLE TO BE REMOVED AFTER IT WAS NOTICED TO BE MISSING FROM THE TUNNELER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10148789 ↗MalfunctionJun 12, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
EVENT #1: AN IMAGING SUPERVISOR REPORTED THAT THE TUNNELER SLEEVE COMPONENT OF THIS BIOFLO CHRONIC DIALYSIS CATHETER SLID OFF THE TUNNELER AND BECOME DETACHED IN THE PATIENT WHILE TUNNELING WAS BEING PERFORMED. THE SLEEVE WAS ABLE TO BE REMOVED AFTER IT WAS NOTICED TO BE MISSING FROM THE TUNNELER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10148787 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page