BioFlo 46470Recalls & reported problems

Navilyst Medical, Inc. · BioFlo Midline with ENDEXO Technology Maximal Barrier Nursing Kit · Class 2

All models of BioFlo (87)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI H965464700), received between Aug 23, 2017 and Feb 12, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment Of Device Component1
  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Fracture1
  • Hole In Material1
  • Material Frayed1
  • Separation Failure1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Separation Failure

    AN END-USER EXPERIENCED AN ISSUE WITH A GUIDEWIRE FROM A BIOFLO MIDLINE 4F SL-20CM MAX STERILE BARRIER NURSING KIT W/ TWO NITINOL GUIDEWIRE PG. IT WAS REPORTED THAT, AFTER MIDLINE INSERTION, RESISTANCE WAS FELT WHILE PULLING THE GUIDEWIRE BACK THROUGH THE SHEATH. THE ENTIRE SHEATH WAS PULLED FROM THE PATIENT, AND NO PRODUCT WAS RETAINED. UPON INSPECTION OF THE COMPONENTS REMOVED FROM THE PATIENT, THE GUIDEWIRE COULD NOT BE ADVANCED BACK THROUGH THE SHEATH AND HAD CAUSED DAMAGE TO THE INTRODUCER DURING ATTEMPTED WITHDRAWAL. THERE WERE NO ADVERSE EFFECTS OR PATIENT HARM REPORTED BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322902 ↗
  • MalfunctionNov 25, 2025Material Frayed; Material Frayed

    AN END-USER EXPERIENCE AN ISSUE WITH A GUIDEWIRE FROM A BIOFLO MIDLINE 4F SL-20CM MAX STERILE BARRIER NURSING KIT W/ TWO NITINOL GUIDEWIRE PG. IT WAS REPORTED THAT, DURING REMOVAL, FRAYING OF THE GUIDEWIRE WAS OBSERVED, AND FRAGMENTS WERE RETAINED WITHIN THE PATIENT'S TISSUE. THE PHYSICIAN WAS ABLE TO RETRIEVE THE RETAINED FRAGMENTS. THERE WERE NO ADVERSE EFFECTS OR PATIENT HARM REPORTED BECAUSE OF THIS INCIDENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 23653002 ↗
  • MalfunctionOct 28, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: HUB OF BIOFLO MIDLINE SEPARATED FROM THE CATHETER 1 HOUR AFTER INITIAL PLACEMENT. CATHETER WAS REMOVED AND REPLACED WITH A NEW OF THE SAME DEVICE. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS AVAILABLE FOR RETURN TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9244378 ↗
  • InjuryDec 17, 2018Fracture; Fracture

    AS REPORTED, DURING PLACEMENT OF BIOFLO MIDLINE CATHETER, THE GUIDEWIRE BECAME SEPARATED AND REQUIRED THE PATIENT TO HAVE A CUT-DOWN IN IR TO REMOVE THE WIRE.

    Patient problems: Foreign Body In Patient

    MAUDE report 8169468 ↗
  • InjuryMar 8, 2018Difficult to Remove; Difficult to Remove

    AS REPORTED: "MIDLINE PLACEMENT. REGISTER NURSE TRIED TO INSERT SHEATH OVER THE WIRE AND IT WOULD NOT ADVANCE. WHEN TRYING TO REMOVE THE WIRE, WIRE COULD NOT BE REMOVED FROM PATIENT'S ARM. PATIENT WAS SENT TO HAVE WIRE REMOVED BY A PHYSICIAN."

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 7324345 ↗
  • MalfunctionFeb 15, 2018Detachment Of Device Component; Detachment Of Device Component

    AS REPORTED, THE PICC HUB DETACHED FROM THE EXTENSION LEG. THE DEVICE WAS REMOVED AND THE CATHETER WAS REPLACED. NO COMPLICATIONS TO THE PATIENT WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7274490 ↗
  • MalfunctionAug 23, 2017Hole In Material; Hole In Material

    AS REPORTED, A 4F MIDLINE CATHETER WHICH HAD BEEN PLACED ON (B)(6) 2017 WAS FLUSHED PRIOR TO THE PATIENT GOING TO CT ON (B)(6) 2017. A PINHOLE LEAK WAS NOTED ON THE EXTENSION LEG AND THE CATHETER WAS REMOVED. THE PATIENT CONDITION WAS REPORTED AS "UNAFFECTED." IT HAS BEEN INDICATED THAT THE USED CATHETER WILL BE RETURNED TO ANGIODYNAMICS FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6813964 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →