Skip to content

TC cavity burs cone square E013834101200Recalls & reported problems

MAILLEFER INSTRUMENTS HOLDING Sàrl · Dental burs · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI J003E0138341012001), received between Jul 17, 2024 and Jun 24, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Device Embedded In Tissue or Plaque2
  • Insufficient Information2
  • Laceration(s)2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 24, 2026Break

    IN THIS EVENT IT IS REPORTED THAT A CARB.B.CAV.CONE SQUARE FG 012 BUR BROKE DURING USE AND COLLAPSED ONTO THE PATIENT'S TONGUE, CAUSING THE PATIENT TO BLEED. THE DOCTOR APPLIED PRESSURE HEMOSTASIS ON THE PATIENT AND PERFORMED FOLLOW UP TREATMENT AFTER THE PATIENT HAD NO DISCOMFORT. NO INJURY.

    Patient problems: Laceration(s)

    MAUDE report 25609776 ↗
  • InjuryDec 4, 2025Break; Break

    IN THIS EVENT IT IS REPORTED THAT A CARB.B.CAV.CONE SQUARE FG 012 BUR THAT WAS USED IN THE TREATMENT OF A PATIENT DAMAGED SOFT TISSUE OF THE PATIENT'S LATERAL EDGE OF THEIR TONGUE DURING USE OF THE BUR. REPORTEDLY, THIS LEAD TO AN ULCERATION THAT WAS TREATED WITH IODINE GLYCERIN. FURTHER INFORMATION REQUESTED.

    Patient problems: Laceration(s)

    MAUDE report 23716272 ↗
  • MalfunctionNov 12, 2025Break; Break

    IN THIS EVENT IT IS REPORTED THAT A CARB.B.CAV.CONE SQUARE FG 012 BUR BROKE DURING USE. THE OUTCOME OF THIS EVENT IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23532952 ↗
  • MalfunctionOct 22, 2025Break; Break

    IN THIS EVENT IT IS REPORTED THAT A CARB.B.CAV.CONE SQUARE FG 012 BUR BROKE DURING USE. REPORTEDLY, NO INJURY BUT THE OUTCOME IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356686 ↗
  • MalfunctionOct 22, 2025Break; Break

    IN THIS EVENT IT IS REPORTED THAT A A CARB.B.CAV.CONE SQUARE FG 012 BUR BROKE DURING USE. REPORTEDLY, NO INJURY BUT THE OUTCOME IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356634 ↗
  • MalfunctionJun 16, 2025Break; Break

    IN THIS EVENT IT IS REPORTED THAT A CARB.B.CAV.CONE SQUARE FG 012 BROKE DURING USE. THE OUTCOME OF THIS EVENT IS UNKNOWN AS OF THIS MDR. REPORTEDLY, THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22210468 ↗
  • MalfunctionJul 17, 2024Break; Break

    IN THIS EVENT IT IS REPORTED THAT A CARB.B.CAV.CONE SQUARE FG 012 BROKE DURING USE. REPORTEDLY, UNBEARABLE PAIN WAS CAUSED TO THE PATIENT. THE OUTCOME IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 19762486 ↗

Other BUR, DENTAL devices

All BUR, DENTAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page