Skip to content

iTotal® Hip Replacement System HBS033000001Recalls & reported problems

Conformis, Inc. · HIP-PATIENT SPECIFIC STEM LEFT · Class 2 · Implant

All models of iTotal Hip Replacement System (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Conformis iTotal hip replacement stem is a patient-specific metal hip implant made to match an individual patient's anatomy. The 33 reports mostly describe the stem coming loose or dislocating after surgery, along with sizing and fit issues; 26 reports involved injuries and no deaths were reported. Some describe the stem not seating fully or a trial neck bending during surgery, leading to revision operations. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
26
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI M572HBS0330000011), received between Nov 21, 2018 and Jul 10, 2025 — classified as 26 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (16 reports).

4 other recalls cover devices cleared under the same FDA submission (K162719); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated16
  • Inadequacy of Device Shape and/or Size5
  • Appropriate Device Problem Term/Code Not Available4
  • Adverse Event Without Identified Device or Use Problem2
  • Crack1
  • Device Markings/Labelling Problem1
  • Fitting Problem1
  • Loose or Intermittent Connection1

Patient problems

  • No Known Impact Or Consequence To Patient10
  • Insufficient Information6
  • No Code Available5
  • Joint Dislocation3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Fall2
  • Limb Fracture2
  • Bone Fracture(s)1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 10, 2025Wrong Label; Wrong Label

    THE DEVICE FOR THE CASE WAS INTENDED TO BE A PATIENT-SPECIFIC SIZE 10 CORDERA HIPRX BUT IT INCLUDED THE STEM FOR A PATIENT-SPECIFIC SIZE 14 CORDERA HIPRX. THE PATIENT'S FEMUR SPLIT BECAUSE IT WAS ONLY PREPARED FOR THE SIZE 10 AND HAD TO BE CABLED FOR COMPLETION OF THE PROCEDURE. THIS IS A PATIENT-SPECIFIC CASE AND THE ONLY IMPACTED CASE WAS THE DEVICE DESCRIBED IN THIS REPORT. NO OTHER IMPACTED DEVICES IN THE FIELD.

    Patient problems: Limb Fracture; Periprosthetic fracture

    MAUDE report 22467708 ↗
  • InjuryFeb 23, 2022Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE SURGEON WASN'T ABLE TO GET A STABLE CONSTRUCT WITH THE PATIENT SPECIFIC STEM COUPLED WITH ANY TRIALED HEAD OFFSET. THE SURGEON ENDED UP REMOVING THE STEM AND IMPLANTING THE SIZE 12 127, 29 STEM. THE SURGEON DIDN'T HAVE A STEM EXTRACTOR AND HE WASN'T HAPPY WITH THE 3D PRINTED TRIAL NECK PERFORMANCE AS IT BENT DURING USE.

    Patient problems: Insufficient Information

    MAUDE report 13586717 ↗
  • InjuryFeb 2, 2022Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DR. CHRIS WILKINSON HAS A PATIENT WHO HAS DISLOCATED A COUPLE OF TIMES AND IS NEEDING TO REVISE HIM. REVISION SURGERY IS PLANNED FOR (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 13429841 ↗
  • InjuryJul 22, 2021Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WILL UNDERGO A THA REVISION. THE REASON FOR THE REVISION IS DUE TO DISLOCATION. PRIMARY SURGERY OCCURRED ON (B)(6) 2020. REVISION SURGERY IS PLANNED FOR (B)(6) 2021.

    Patient problems: Insufficient Information

    MAUDE report 12214978 ↗
  • InjuryJul 2, 2021Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT WAS REVISED TO A DEPUY CORAIL STEM. THE PATIENT-SPECIFIC STEM WAS NOT ABLE TO GET FULLY SEATED AND SAT ABOUT 6-7MM PROUD WHILE ALSO ADDING ADDITIONAL LEG LENGTH. THE MAIN FACTOR FOR THE REVISION WAS SIGNIFICANT SCOLIOSIS AND PELVIC OBLIQUITY IN THE PATIENT WHICH WASN'T FULLY APPRECIATED BY THE OFFICE X-RAYS, I-VIEW, OR THE SURGEON PREOPERATIVELY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12110222 ↗
  • InjuryJun 17, 2021Device Dislodged or Dislocated

    IT WAS REPORTED THAT A REVISION IS PLANNED FOR A PATIENT WITH A POST-OPERATIVE DISLOCATION. IT IS UNKNOWN AT THIS TIME WHAT THE CAUSE FOR THE DISLOCATION WAS.

    Patient problems: Joint Dislocation

    MAUDE report 12021513 ↗
  • InjuryMay 18, 2021Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT SUFFERED A DISLOCATION OF THEIR HIP IMPLANT. REVISION SURGERY OCCURRED ON (B)(6) 2021 WHERE AN EXCHANGE TO DUAL MOBILITY AND HEAD EXCHANGE WERE PERFORMED. THERE WAS NO DAMAGE TO ANY DEVICE DISCOVERED. THE SURGEON BELIEVES THE DISLOCATION WAS RELATED TO A SPINE DEFORMITY THAT THE PATIENT HAS.

    Patient problems: Insufficient Information

    MAUDE report 11843174 ↗
  • InjuryMay 18, 2021Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT SUFFERED A DISLOCATION OF THEIR HIP IMPLANT. REVISION SURGERY OCCURRED ON (B)(6) 2021 WHERE A LINER EXCHANGE AND HEAD EXCHANGE WAS PERFORMED. THERE WERE NO LOOSENING OF COMPONENTS DISCOVERED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11843160 ↗
  • MalfunctionMay 10, 2021Fitting Problem; Positioning Problem

    IT WAS REPORTED THAT THE SURGEON HAD TROUBLES WITH THE FIT AND POSITIONING OF THE PATIENT-SPECIFIC STEM. THE SURGEON WAS UNABLE TO FULLY SEAT THE SIZE 13 STEM WHICH SAT 10 MM PROUD. HE COMPLETED THE SURGERY WITH A SMALLER STEM FROM ANOTHER VENDOR WHILE STILL UTILIZING THE CONFORMIS CUP, LINER, AND HEAD.

    Patient problems: Insufficient Information

    MAUDE report 11802851 ↗
  • MalfunctionApr 14, 2021Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE STEM HUNG UP ABOUT 4 MM PROUD AND THE SURGEON NEEDED AN EXTRACTION INSTRUMENT TO GET THE IMPLANT OUT. THE SURGEON TRAILED WITH SHORTER HEADS, BUT ULTIMATELY STAYED WITH THE PLANNED LENGTH DUE TO OFFSET CONSEQUENCES. THE PATIENT LEG LENGTH WAS LEFT LONGER THAN THE PLAN AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11668014 ↗
  • MalfunctionMar 24, 2021Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE SURGEON WENT TO UTILIZE THE PATIENT-SPECIFIC STEM AND SAID IT WAS TOO SHORT. THE SURGEON USED A STEM FROM A DIFFERENT COMPANY TO COMPLETE THE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 11566261 ↗
  • InjurySep 24, 2020Loose or Intermittent Connection

    IT WAS REPORTED THAT THE PATIENT FELL IN THE SHOWER AND BROKE THEIR FEMUR. AS A RESULT, THE FEMORAL STEM IS LOOSE. REVISION SURGERY IS PLANNED TO ADDRESS THE ISSUE.

    Patient problems: Fall; Limb Fracture

    MAUDE report 10583810 ↗
  • MalfunctionAug 25, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD A POST-OPERATIVE HIP DISLOCATION SHORTLY AFTER THEIR PRIMARY SURGERY. THE SURGEON SUCCESSFULLY CORRECTED THE ISSUE VIA A CLOSED REDUCTION. THE SURGEON BELIEVES THE ANESTHESIA LASTED LONGER THAN USUAL FROM THE PRIMARY SURGERY AND CAUSED THE PATIENT TO STEP INCORRECTLY, CAUSING THE DISLOCATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10450899 ↗
  • InjuryJul 23, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD A POST-OPERATIVE HIP DISLOCATION. THE PATIENT RECEIVED A CLOSED REDUCTION TO ADDRESS THE ISSUE.

    Patient problems: No Code Available

    MAUDE report 10317862 ↗
  • InjuryJul 15, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD A POST-OPERATIVE HIP DISLOCATION. THE PATIENT RECEIVED A CLOSED REDUCTION TO ADDRESS THE ISSUE.

    Patient problems: No Code Available

    MAUDE report 10280087 ↗
  • MalfunctionJun 30, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT A PATIENT WENT TO THE ER WITH RIGHT HIP PAIN AFTER SHE FELT A POP. AN ORTHO DISLOCATION REDUCTION PROCEDURE WAS PERFORMED UNDER SEDATION AND THE PATIENT WAS DISCHARGED WITH A KNEE IMMOBILIZER.

    Patient problems: Pain

    MAUDE report 10215459 ↗
  • InjuryMay 15, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAS A HIP DISLOCATION. REPLACEMENT COMPONENTS HAVE BEEN ORDERED FOR REVISION SURGERY.

    Patient problems: Joint Dislocation

    MAUDE report 10063332 ↗
  • InjuryApr 21, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD A FALL RESULTING IN A HIP DISLOCATION. LINERS AND HEADS HAVE BEEN ORDERED FOR REVISION SURGERY.

    Patient problems: Fall

    MAUDE report 9982611 ↗
  • InjuryApr 8, 2020Crack

    IT WAS REPORTED THAT THE PATIENT'S GREATER TROCHANTER WAS CRACKED DURING A TOTAL HIP ARTHROPLASTY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9935391 ↗
  • InjuryFeb 21, 2020Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE END OF SURGERY, A CALCAR CRACK WAS NOTICED IN THE PATIENT. THE SURGEON UTILIZED A CABLE TO STABILIZE THE CRACK AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Code Available

    MAUDE report 9742043 ↗
  • InjuryJan 28, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAS A POST-OPERATIVE HIP DISLOCATION. REVISION SURGERY IS PLANNED TO EXCHANGE THE LINER AND HEAD.

    Patient problems: No Code Available

    MAUDE report 9638017 ↗
  • InjuryJan 8, 2020Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE PLANNED STEM SIZE FOR THE PATIENT WAS TOO LARGE TO ACCOMMODATE THE FEMORAL CANAL. THE SURGEON MADE A CONVERSION TO AN OFF-THE-SHELF HIP REPLACEMENT AND COMPLETED THE SURGERY SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9565682 ↗
  • InjuryJan 3, 2020Device Dislodged or Dislocated

    IT WAS REPORTED THAT A HIP PATIENT HAS A POST-OPERATIVE DISLOCATION. REPLACEMENT COMPONENTS HAVE BEEN ORDERED FOR REVISION SURGERY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9546354 ↗
  • InjuryNov 27, 2019Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE CONFORMIS CUSTOM STEM DID NOT FIT PROPERLY DURING SURGERY. THE SURGEON HAD TO MAKE AN INTRAOPERATIVE CONVERSION TO AN OFF-THE-SHELF STEM TO COMPLETE THE PROCEDURE.

    Patient problems: Failure of Implant

    MAUDE report 9385568 ↗
  • InjuryNov 15, 2019Malposition of Device

    IT WAS REPORTED THAT THE SURGEON CONVERTED FROM A CONFORMIS CUSTOM STEM AND HEAD TO AN OFF-THE-SHELF STEM AND HEAD. THE CONFORMIS STEM CCD ANGLE OF 125 WAS TOO VARUS AND CREATED TOO MUCH OFFSET WITH THE XL HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9329194 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162719. One submission often covers many models, so check the product and lot details in each record.

  • Z-0063-2020 · Aug 22, 2019 · Terminated
    iTotal Hip Replacement System, Model HBS-033-0015-020101, Size 15 Hip-Patient Specific Stem Right
  • Z-0064-2020 · Aug 22, 2019 · Terminated
    iTotal Hip Replacement System, Model HBS-033-0013-020101, Size 13 Hip-Patient Specific Stem Right
  • Z-1775-2019 · Mar 29, 2019 · Terminated
    Conformis iTotal Hip Replacement System: HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM LEFT)
  • Z-0731-2019 · Oct 16, 2018 · Terminated
    iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101

Other Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented devices

All Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page