Skip to content

iTotal Identity PS TPS-101-111211Recalls & reported problems

Conformis, Inc. · ITOTAL IDENTITY PS FEMORAL IMPLANT (COCR, CEMENTED), RIGHT or left · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M572TPS1011112111), received between Aug 5, 2022 and Jul 16, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Adverse Event Without Identified Device or Use Problem1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Joint Laxity2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Septic loosening1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 16, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT REQUIRES REVISION SURGERY DUE TO PATIENT INFECTION AFTER TOTAL KNEE REPLACEMENT. THE COMPONENTS WERE MANUFACTURED, PACKAGED, AND STERILIZED CORRECTLY AND IT IS NOT BELIEVED THAT THE IMPLANT CAUSED THE INFECTION AND SUBSEQUENT REVISION.

    Patient problems: Septic loosening

    MAUDE report 25840725 ↗
  • InjuryNov 18, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PS FEMORAL IMPLANT PACKAGING HAD A HOLE THROUGH BOTH LAYERS OF PACKAGING, BREACHING THE STERILE BARRIER. THIS WAS DISCOVERED DURING SURGERY; THE SURGEON CHOSE TO RE-STERILIZE THE IMPLANT. THIS RE-STERILIZATION PROCESS TAKES AROUND 1-2 HOURS AND NEEDED TO BE COMPLETED WHILE THE PATIENT WAS UNDER ANESTHESIA. THIS WAS A MAJOR DELAY TO THE SURGERY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15829374 ↗
  • InjuryOct 25, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT REPLACEMENT POLY INSERTS HAVE BEEN REQUESTED FOR REVISION SURGERY. REASON FOR REVISION IS LAXITY. PRIMARY SURGERY OCCURRED ON (B)(6) 2022. REVISION SURGERY IS PLANNED FOR (B)(6) 2022.

    Patient problems: Joint Laxity

    MAUDE report 15670651 ↗
  • InjuryAug 5, 2022Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT REPLACEMENT POLY INSERTS HAVE BEEN REQUESTED FOR REVISION SURGERY. REASON FOR REVISION IS LAXITY. PRIMARY SURGERY OCCURRED ON (B)(6) 2021. REVISION SURGERY IS PLANNED FOR (B)(6) 2022.

    Patient problems: Joint Laxity

    MAUDE report 15180477 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page