Skip to content

CONSERVE® 38AM4600Recalls & reported problems

Microport Orthopedics Inc. · CONSERVE® A-CLASS® BFH® HEAD MEDIUM NECK 46MM · Class 3 · Implant

All models of CONSERVE (74)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI M68438AM46001), received between Jun 24, 2019 and Feb 1, 2024 — classified as 8 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information5
  • Material Disintegration3
  • Corroded1

Patient problems

  • Pain4
  • Reaction3
  • Fluid Discharge1
  • Insufficient Information1
  • Metal Related Pathology1
  • No Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 1, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, AN MRI OF THE PATIENT'S LEFT HIP TAKEN DUE TO PAIN ON (B)(6) 2021, THIS SHOWED A PERIARTICULAR CYST, SUGGESTIVE OF METALLOSIS. THE METAL ION LEVELS EXTRACTED ON (B)(6) 2022 WERE FOUND TO BE CR = 3.4 AND CO = 3.8 WHICH INCREASED WITH RESPECT TO THE METAL IONS EXTRACTED ON MAY 14, 2021, WHICH WERE CR= 2.2 AND CO= 2.9.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 18627704 ↗
  • InjuryDec 22, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, THE HEAD AND THE NECK ON THE LEFT SIDE WERE REVISED DUE TO PAIN. NON EXPLANTED PRODUCTS: PRODUCT ID: PRLX5812 PROFEMUR® LX STEM SZ 12 5/8 PS COATED/ LOT NUMBER: 0711374029/ QTY: 1 PRODUCT ID: 38024652 CONSERVE® THIN SHELL 46MM ID 52MM OD BEADED/ LOT NUMBER: 0611361525/ QTY: 1.

    Patient problems: Pain

    MAUDE report 18387039 ↗
  • InjuryDec 22, 2020Material Disintegration; Material Disintegration

    ALLEGEDLY, PATIENT WAS REVISED DUE TO METAL DEBRIS.

    Patient problems: Insufficient Information

    MAUDE report 11059348 ↗
  • InjuryMar 26, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY. THE IMPLANTS GENERATE METAL DEBRIS, CORROSION AND METAL IONS, WHICH CAUSE DANGEROUSLY ELEVATED BLOOD LEVELS OF COCRIONS ADVERSE TISSUE REACTIONS, PSEUDOTUMORS, NECROSIS, BONE LOSS .

    Patient problems: Reaction

    MAUDE report 9882381 ↗
  • InjuryDec 10, 2019Corroded; Material Disintegration; Corroded; Material Disintegration

    ALLEGEDLY, THE PATIENT HAD PAIN IN HER LEFT HIP, SUBJECT HAS A HISTORY OF ELEVATED COBALT AND CHROMIUM LEVELS. SUBJECT HAS A TOTAL OF FIVE TOTAL JOINTS IN HER BODY, INCLUDING BOTH HIPS. THE PHYSICIAN ORDERED METAL ION TESTING FROM JOINT FLUID FROM BOTH HIPS. NOT ENOUGH FLUID WAS OBTAINED FROM THE RIGHT HIP TO PERFORM TESTING. COBALT AND CHROMIUM METAL IONS WERE PRESENT IN THE LEFT HIP FLUID. UPDATE DECEMBER 5 2019. THE CONSERVE TOTAL A-CLASS FEMORAL HEAD WAS REMOVED. CORROSION OF THE FEMORAL HEAD-NECK/ TRUNNION JUNCTION WAS NOTED. THE MODULAR NECK AND FEMORAL STEM WERE WELL FIXED AND WERE NOT REVISED. THE SUBJECT WAS DISCHARGED THE SAME DAY.

    Patient problems: Pain

    MAUDE report 9444145 ↗
  • InjuryJul 29, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY THE SHELL APPEARS TO BE LOOSE. CUP WAS CUT OUT AND REPLACED WITH STRYKERS CUP, LINER AND DUAL MOBILITY. OUR 28MM LONG HEAD WAS USED WITH THE DUAL MOBILITY . OUR HEDA CAME OUT (46MM).

    Patient problems: Reaction

    MAUDE report 8838380 ↗
  • InjuryJun 26, 2019Material Disintegration; Material Disintegration

    ALLEGEDLY THE PATIENT HAS PREVIOUSLY UNDERWENT A LEFT TOTAL HIP REPLACEMENT ON (B)(6) 2008. SHE WAS PRESENTED WITH LEFT HIP PAIN. HER MRI REVEALS SOME EXTENSIVE MULTI-LOCULATED FLUID COLLECTION. HER SERUM CHROMIUM AND COBALT LEVELS ARE OVER 20. A REVISION TO A TITANIUM FEMORAL NECK WAS RECOMMENDED. SHE HAS A BODY WEIGHT OF (B)(6) POUNDS. REVISION SURGERY IS SET FOR (B)(6) 2019.

    Patient problems: Fluid Discharge; Pain; Reaction

    MAUDE report 8736887 ↗
  • InjuryJun 24, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT SCHEDULED TO BE REVISED DUE TO UNKNOWN COMPLICATIONS.

    Patient problems: No Information

    MAUDE report 8728002 ↗

FDA clearance and classification

The FDA regulates product code JDL (prosthesis, hip, semi-constrained (metal cemented acetabular component)) under 21 CFR 888.3320 as a Class III device. 21 CFR 888.3320 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CONSERVE® 38AM4600

Has the CONSERVE® 38AM4600 been recalled?

No FDA recall record in our data names this exact device identifier (DI M68438AM46001), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CONSERVE® 38AM4600?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Jun 24, 2019 and Feb 1, 2024. In those reports, the most common specific device problems are “Material Disintegration” (3) and “Corroded” (1); the most common patient problems are “Pain” (4), “Reaction” (3) and “Fluid Discharge” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CONSERVE® 38AM4600?

None of the reports are classified as deaths, 8 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CONSERVE® 38AM4600 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the CONSERVE® 38AM4600, and is it still sold?

The FDA's device database lists Microport Orthopedics Inc. as the company responsible for it (brand CONSERVE®, model 38AM4600). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices

All Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page