CONSERVE® 38AM4600Recalls & reported problems
Microport Orthopedics Inc. · CONSERVE® A-CLASS® BFH® HEAD MEDIUM NECK 46MM · Class 3 · Implant
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI M68438AM46001), received between Jun 24, 2019 and Feb 1, 2024 — classified as 8 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information5
- Material Disintegration3
- Corroded1
Patient problems
- Pain4
- Reaction3
- Fluid Discharge1
- Insufficient Information1
- Metal Related Pathology1
- No Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryFeb 1, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, AN MRI OF THE PATIENT'S LEFT HIP TAKEN DUE TO PAIN ON (B)(6) 2021, THIS SHOWED A PERIARTICULAR CYST, SUGGESTIVE OF METALLOSIS. THE METAL ION LEVELS EXTRACTED ON (B)(6) 2022 WERE FOUND TO BE CR = 3.4 AND CO = 3.8 WHICH INCREASED WITH RESPECT TO THE METAL IONS EXTRACTED ON MAY 14, 2021, WHICH WERE CR= 2.2 AND CO= 2.9.
Patient problems: Metal Related Pathology; Pain
MAUDE report 18627704 ↗InjuryDec 22, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, THE HEAD AND THE NECK ON THE LEFT SIDE WERE REVISED DUE TO PAIN. NON EXPLANTED PRODUCTS: PRODUCT ID: PRLX5812 PROFEMUR® LX STEM SZ 12 5/8 PS COATED/ LOT NUMBER: 0711374029/ QTY: 1 PRODUCT ID: 38024652 CONSERVE® THIN SHELL 46MM ID 52MM OD BEADED/ LOT NUMBER: 0611361525/ QTY: 1.
Patient problems: Pain
MAUDE report 18387039 ↗InjuryDec 22, 2020Material Disintegration; Material Disintegration
ALLEGEDLY, PATIENT WAS REVISED DUE TO METAL DEBRIS.
Patient problems: Insufficient Information
MAUDE report 11059348 ↗InjuryMar 26, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY. THE IMPLANTS GENERATE METAL DEBRIS, CORROSION AND METAL IONS, WHICH CAUSE DANGEROUSLY ELEVATED BLOOD LEVELS OF COCRIONS ADVERSE TISSUE REACTIONS, PSEUDOTUMORS, NECROSIS, BONE LOSS .
Patient problems: Reaction
MAUDE report 9882381 ↗InjuryDec 10, 2019Corroded; Material Disintegration; Corroded; Material Disintegration
ALLEGEDLY, THE PATIENT HAD PAIN IN HER LEFT HIP, SUBJECT HAS A HISTORY OF ELEVATED COBALT AND CHROMIUM LEVELS. SUBJECT HAS A TOTAL OF FIVE TOTAL JOINTS IN HER BODY, INCLUDING BOTH HIPS. THE PHYSICIAN ORDERED METAL ION TESTING FROM JOINT FLUID FROM BOTH HIPS. NOT ENOUGH FLUID WAS OBTAINED FROM THE RIGHT HIP TO PERFORM TESTING. COBALT AND CHROMIUM METAL IONS WERE PRESENT IN THE LEFT HIP FLUID. UPDATE DECEMBER 5 2019. THE CONSERVE TOTAL A-CLASS FEMORAL HEAD WAS REMOVED. CORROSION OF THE FEMORAL HEAD-NECK/ TRUNNION JUNCTION WAS NOTED. THE MODULAR NECK AND FEMORAL STEM WERE WELL FIXED AND WERE NOT REVISED. THE SUBJECT WAS DISCHARGED THE SAME DAY.
Patient problems: Pain
MAUDE report 9444145 ↗InjuryJul 29, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY THE SHELL APPEARS TO BE LOOSE. CUP WAS CUT OUT AND REPLACED WITH STRYKERS CUP, LINER AND DUAL MOBILITY. OUR 28MM LONG HEAD WAS USED WITH THE DUAL MOBILITY . OUR HEDA CAME OUT (46MM).
Patient problems: Reaction
MAUDE report 8838380 ↗InjuryJun 26, 2019Material Disintegration; Material Disintegration
ALLEGEDLY THE PATIENT HAS PREVIOUSLY UNDERWENT A LEFT TOTAL HIP REPLACEMENT ON (B)(6) 2008. SHE WAS PRESENTED WITH LEFT HIP PAIN. HER MRI REVEALS SOME EXTENSIVE MULTI-LOCULATED FLUID COLLECTION. HER SERUM CHROMIUM AND COBALT LEVELS ARE OVER 20. A REVISION TO A TITANIUM FEMORAL NECK WAS RECOMMENDED. SHE HAS A BODY WEIGHT OF (B)(6) POUNDS. REVISION SURGERY IS SET FOR (B)(6) 2019.
Patient problems: Fluid Discharge; Pain; Reaction
MAUDE report 8736887 ↗InjuryJun 24, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, PATIENT SCHEDULED TO BE REVISED DUE TO UNKNOWN COMPLICATIONS.
Patient problems: No Information
MAUDE report 8728002 ↗
Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices
- Omnifit 2001-2246Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained (metal cemented acetabular component)1 recall0 reports
- Taperloc Complete Micro Primary Femoral 51-107150Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)41 reports
- Taperloc Complete Micro Primary Femoral 51-107140Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)36 reports
- Dynasty DSBFGD52Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)12 reports
- CONSERVE® 38AM4804Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)7 reports
- CONSERVE® 38AM4404Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)6 reports
- Dynasty DSBFGF56Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)5 reports
- CONSERVE® 38AM4635Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.