Skip to content

CONSERVE® 38CH3800Recalls & reported problems

Microport Orthopedics Inc. · CONSERVE® TOTAL 38MM BCH® FEMORAL HEAD · Class 2 · Implant

All models of CONSERVE (74)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M68438CH38001), received between Feb 17, 2020 and Feb 22, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Break1

Patient problems

  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Loss of Range of Motion1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 22, 2024Break; Break

    ALLEGEDLY, ON (B)(6) 2015, THE PATIENT HAD A TOTAL LEFT HIP REPLACEMENT AND RECEIVED A WRIGHT PROFEMUR STEM, THE SIZE 9, STANDARD OFFSET DYNASTY SHELL SIZE 54, CLASS A LINER, SIZE 38 AND 6.5 X 35 MM. CANCELLOUS SCREW, SLEEVE FOR THE SHORT NECK AND THE CERAMIC 38 MM. HEAD.ON (B)(6) 2023 ( LESS THAN 8 YRS. AFTER THE SURGERY) AT 6:00 AM LYING IN BED THERE WAS A LOUD POP AND THE PATIENT COULD NOT USE HIS LEFT LEG. HE WENT TO THE LOCAL EMERGENCY ROOM WHERE THEY TOOK AN X-RAY AND STATED "YOUR HIP DEVICE HAS FAILED". THE X-RAY REVEALED A FRACTURE IN THE CERAMIC HEAD OF THE HIP DEVICE. BY THE TIME THE PATIENT COULD GET A HIP REVISION SURGERY ON (B)(6), THE CERAMIC HEAD HAD BROKEN IN TO 15-20 SHARDS WHICH THE SURGEON HAD TO REMOVE. THE PATIENT HAS NOW RECOVERED AND IS PHYSICALLY BETTER. THE PATIENT IS WRITING BECAUSE HE BELIEVES HE IS ENTITLED TO SOME SORT OF COMPENSATION FOR THE FAILURE OF THE HIP DEVICE AND CONSEQUENT PHYSICAL AND EMOTIONAL DISTRESS.

    Patient problems: Insufficient Information

    MAUDE report 18760326 ↗
  • InjuryJan 31, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, COMPONENTS REVISED FOR UNKNOWN REASONS. HIGH MBI.

    Patient problems: Insufficient Information

    MAUDE report 13397179 ↗
  • InjuryJul 15, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED DO TO UNKNOWN REASONS, AFTER A LEFT HIP ARTHROPLASTY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12175868 ↗
  • InjuryFeb 17, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED DUE TO ELEVATED COBALT ION LEVELS, PAIN AND STIFFENS OF THE HIP. IT WAS DISCOVERED THAT THE DEVICE FAILED DUE TO CORROSION OF THE OBLONG TAPER OF MODULAR NECK WHERE IT SEATED IN THE POCKED OF THE TITANIUM STEM.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 9717029 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page