Skip to content

Dynasty DLXPGC32Recalls & reported problems

Microport Orthopedics Inc. · DYNASTY® A-CLASS® 32MM STD GROUP C CROSSLINKED POLY LINER · Class 2 · Implant

All models of Dynasty (44)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M684DLXPGC321), received between Jun 4, 2020 and Oct 11, 2022 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated3
  • Unstable1

Patient problems

  • Metal Related Pathology1
  • No Information1
  • Osteopenia/ Osteoporosis1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 11, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    ALLEGEDLY, ON OR ABOUT (B)(6) 2004 PATIENT HAD A PROFEMUR TOTAL HIP SYSTEM IMPLANTED IN HER RIGHT HIP. ON OR ABOUT (B)(6) 2019 PATIENT REPORTED FOR REVISION SURGERY DUE TO ELEVATED COBAL AND CHROMIUN LEVELS, PAIN AND EVIDENCE OF A PSEUDOTUMOR. PATIENT'S FIRST REVISION WAS NECESSARY BECAUSE THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION. IT WAS DISCOVERED CORROSION OF THE PROFEMUR COCR MODULAR NECK WHERE IT SEATED IN THE POCKET OF THE MODULAR STEM. THE IMPLANTED NECK AND HEAD WERE REVISED. AFTER THE REVISION SURGERY PATIENT REPORTED AT LEAST 3 DISLOCATIONS. THE PREVIOUS EVENTS WERE CAPTURED UNDER MICROPORT ORTHOPEDICS GROUP NUMBER: 20060165. ADDITIONAL INFORMATION RECEIVED 7/21/2022: ALLEGEDLY, ON OR ABOUT (B)(6) 2021, PATIENT REPORTER FOR A REVISION SURGERY OF THE RIGHT TOTAL HIP ARTHROPLASTY, REVISION OF BOTH THE ACETABULAR AND FEMORAL COMPONENTS. SURGEON RECOMMENDED THIS SECOND REVISION SURGERY DUE TO RECURRENT INSTABILITY WITH DISLOCATIONS. DURING REVISION SURGERY, SURGEON DETERMINED THE ONLY WAY TO SAFELY REMOVE THE MODULAR STEM WAS TO PERFORM AN EXTENDED TROCHANTERIC OSTEOTOMY(ETO), WHICH REQUIRED FOUR ALL MILES CABLES WITH THE USE OF A 240 MM REVISION STEM. FIRS

    Patient problems: Metal Related Pathology; Osteopenia/ Osteoporosis

    MAUDE report 15583533 ↗
  • InjuryJul 15, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY, PATIENT REVISED DUE TO COPIOUS AMOUNT OF BLACK DISCOLORED TISSUE, BLACK DISCOLORATION OF THE JOINT CAPSULE CONSISTENT WITH METALLOSIS AND DISLOCATION. (LEFT HIP)

    Patient problems: Unspecified Tissue Injury

    MAUDE report 12178675 ↗
  • InjuryJun 4, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY THE PATIENT WAS REVISED DUE TO MULTIPLE HIP DISLOCATIONS, HAVING A FAILED ACETABULAR. (RIGHT) ADDITIONAL INFORMATION RECEIVED ON 12-SEP-2019: SAE# (B)(6). ALLEGEDLY, PERIPROSTHETIC FRACTURE OF THE ACETABULAR COMPONENT PATIENT WAS SEEN IN CLINIC POST SX BETWEEN JUL 2012-JAN 2013. IMPROVEMENTS WERE SLOW AND THE PATIENT WAS NOT FULLY WEIGHT BEANING. PATIENT WAS FOUND TO HAVE A FAILED ACETABULAR ON (B)(6) 2013, AND THE PATIENT UNDERWENT A REVISION WITH CAGE RECONSTRUCTION.

    Patient problems: No Information

    MAUDE report 10119075 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →