Microport Orthopedics Inc. · ADVANCE® PRIMARY FEMORAL SIZE 3 RIGHT NONPOROUS · Class 2 · Implant

All models of ADVANCE (147)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI M684KFTCNP3R1), received between Jun 25, 2019 and Apr 19, 2024 — classified as 12 injuries.

The most frequently reported specific device problem is “Unstable” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable5
  • Insufficient Device Problem Information4
  • Device Dislodged or Dislocated3
  • Malposition of Device3
  • Fracture2
  • Loss of Osseointegration1

Patient problems

  • Bone Fracture(s)5
  • No Information4
  • Insufficient Information2
  • Arthritis1
  • Osteolysis1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 19, 2024Loss of Osseointegration; Loss of Osseointegration

    ALLEGEDLY, THE PATIENT WAS DIAGNOSED WITH RADIOLOGY. EVIDENT LOOSENING OF THE PROSTHESIS WAS DETECTED, THE TIBIAL BASE WAS REPLACED. THE INITIAL IMPLANTATION WAS PERFORMED IN 2015. FIRST REVISION: IT WAS DETECTED SHORTLY AFTER IMPLANTATION THAT THERE WAS INFECTION AND AN INSERT CHANGE WAS PERFORMED. PE WAS NOT FIRMLY ATTACHED TO THE TIBIAL COMPONENT AS INTENDED, BUT BECAME DETACHED FROM IT. AS A RESULT, THERE WAS WEAR OF THE UNDERSIDE OF THE INSERT WITH POLYETHYLENE GRANULOMAS. THERE WERE POLYETHYLENE GRANULOMAS THROUGHOUT THE JOINT AND COMPLETE LOOSENING OF THE PROSTHESIS. THE JOINT WAS RECONSTRUCTED WITH A COMPLEX COUPLED, MODULAR, STEM-GUIDED PROSTHESIS AND ALL GRANULOMAS WERE REMOVED IN APPROXIMATELY 3 HOURS OF REVISION SURGERY. THE EXPLANTS WERE GIVEN TO THE PATIENT: PRODUCT ID: KIMP312R ADVANCE® II MEDIAL PIVOT TIBIAL INSERT SZ 3 RIGHT 12MM / LOT NO.: 1687317/ QTY: (B)(4). PRODUCT ID: KFTCNP3R ADVANCE® PRIMARY FEMORAL SIZE 3 RIGHT NONPOROUS / LOT NO.: 1707334/ QTY: (B)(4). PRODUCT ID: KTCCNP30 ADVANCE® II COCR TIBIAL BASE NONPOROUS SZ 3 STANDARD/ LOT NO.: 1644600/ QTY: (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19146252 ↗
  • InjuryFeb 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, LYSIS FEMUR/PERIPROSTHETIC FRACTURE SIDE: R REVISION NJR: (B)(6). GENDER: F EXPLANTED: FEMORAL COMPONENT, TIBIAL COMPONENT, TIBIAL LINER

    Patient problems: Bone Fracture(s); Osteolysis

    MAUDE report 18796415 ↗
  • InjuryFeb 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, OTHER. SIDE: R. REVISION NJR: 854772. GENDER: F. EXPLANTED: ALL.

    Patient problems: Insufficient Information

    MAUDE report 18796292 ↗
  • InjuryAug 3, 2022Fracture

    ALLEGEDLY, PATIENT WAS REVISED DUE TO PERI-PROSTHETIC FRACTURE. REVISION NJR NUMBER: (B)(4). SIDE: R . PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING. COMPONENT NOT REVISED: PRODUCT: ADVANCE II MEDIAL PIVOT TIBIA INSERT SZ 3 RIGHT 10MM PRODUCT ID: KIMP310R LOT # : 125303897 QTY: 1.

    Patient problems: Bone Fracture(s)

    MAUDE report 15160770 ↗
  • InjuryMay 12, 2021Fracture

    ALLEGEDLY, PATIENT WAS REVISED DUE TO PERI-PROSTHETIC FRACTURE. (B)(4). SIDE: R. PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING. (B)(6) REFERENCE NO: (B)(4).

    Patient problems: Bone Fracture(s)

    MAUDE report 11812925 ↗
  • InjuryDec 4, 2020Unstable; Unstable

    ALLEGEDLY, PATIENT WAS REVISED DUE TO ASEPTIC LOOSENING TIBIA; INSTABILITY; WEAR OF POLYETHYLENE COMPONENT; PROGRESSIVE ARTHRITIS. REVISION NJR NUMBER: (B)(4). SIDE: R. PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING. (B)(6) REFERENCE NO: (B)(4).

    Patient problems: Arthritis

    MAUDE report 10945709 ↗
  • InjuryJun 29, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED DUE TO PERIPROSTHETICFRACTURE. COMPONENT NOT REVISED: ADVANCE II COCR TIBIA BASE NONPOROUS SZ 3 STANDARD PRODUCT ID: KTCCNP30 LOT ID: 1736502. ADVANCE II MEDIAL PIVOT TIBIA INSERT SZ 3 RIGHT 12MM PRODUCT ID: KIMP312R LOT ID: 1512029. REVISION NJR NUMBER: 4373718. SIDE: R. PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING.

    Patient problems: Bone Fracture(s)

    MAUDE report 10207157 ↗
  • InjuryJun 26, 2020Unstable; Unstable

    ALLEGEDLY, PATIENT WAS REVISED DUE TO INSTABILITY. REVISION NJR NUMBER: (B)(4); SIDE: R; PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING.

    Patient problems: No Information

    MAUDE report 10203386 ↗
  • InjuryJun 23, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED DUE TO PERI PROSTHETIC FRACTURE. COMPONENT NOT REVISED: ADVANCE II COCR TIBIA BASE NONPOROUS SZ 3 STANDARD PRODUCT ID: KTCCNP30 LOT #: 020153197. REVISION NJR NUMBER: 4348091. SIDE: R. PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING.

    Patient problems: Bone Fracture(s)

    MAUDE report 10184206 ↗
  • InjuryJun 25, 2019Unstable; Malposition of Device; Device Dislodged or Dislocated; Unstable; Malposition of Device; Device Dislodged or Dislocated

    ALLEGEDLY, PATIENT WAS REVISED DUE TO DISLOCATION SUBLUXATION; INSTABILITY; MALALIGNMENT REVISION NJR NUMBER: (B)(4), SIDE: R, PRIMARY ASA: P3 - INCAPACITATING SYSTEMIC DISEASE.

    Patient problems: No Information

    MAUDE report 8732940 ↗
  • InjuryJun 25, 2019Unstable; Malposition of Device; Device Dislodged or Dislocated; Unstable; Malposition of Device; Device Dislodged or Dislocated

    ALLEGEDLY, PATIENT WAS REVISED DUE TO DISLOCATION SUBLUXATION; INSTABILITY; MALALIGNMENT REVISION NJR NUMBER: (B)(4). SIDE: R. PRIMARY ASA: P3 - INCAPACITATING SYSTEMIC DISEASE.

    Patient problems: No Information

    MAUDE report 8732939 ↗
  • InjuryJun 25, 2019Unstable; Malposition of Device; Device Dislodged or Dislocated; Unstable; Malposition of Device; Device Dislodged or Dislocated

    ALLEGEDLY, PATIENT WAS REVISED DUE TO DISLOCATION SUBLUXATION; INSTABILITY; MALALIGNMENT REVISION NJR NUMBER: (B)(6). SIDE: R. PRIMARY ASA: P3 - INCAPACITATING SYSTEMIC DISEASE.

    Patient problems: No Information

    MAUDE report 8732937 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →