ADVANCE® KIMP312RRecalls & reported problems
Microport Orthopedics Inc. · ADVANCE® II MEDIAL PIVOT TIBIAL INSERT SZ 3 RIGHT 12mm · Class 2 · Implant
All models of ADVANCE (147)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M684KIMP312R1), received between Jun 20, 2019 and Apr 19, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Unstable” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable2
- Loss of Osseointegration1
- Naturally Worn1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Insufficient Information1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 19, 2024Loss of Osseointegration; Naturally Worn; Loss of Osseointegration; Naturally Worn
ALLEGEDLY, THE PATIENT WAS DIAGNOSED WITH RADIOLOGY. EVIDENT LOOSENING OF THE PROSTHESIS WAS DETECTED, THE TIBIAL BASE WAS REPLACED. THE INITIAL IMPLANTATION WAS PERFORMED IN 2015. FIRST REVISION: IT WAS DETECTED SHORTLY AFTER IMPLANTATION THAT THERE WAS INFECTION AND AN INSERT CHANGE WAS PERFORMED. PE WAS NOT FIRMLY ATTACHED TO THE TIBIAL COMPONENT AS INTENDED, BUT BECAME DETACHED FROM IT. AS A RESULT, THERE WAS WEAR OF THE UNDERSIDE OF THE INSERT WITH POLYETHYLENE GRANULOMAS. THERE WERE POLYETHYLENE GRANULOMAS THROUGHOUT THE JOINT AND COMPLETE LOOSENING OF THE PROSTHESIS. THE JOINT WAS RECONSTRUCTED WITH A COMPLEX COUPLED, MODULAR, STEM-GUIDED PROSTHESIS AND ALL GRANULOMAS WERE REMOVED IN APPROXIMATELY 3 HOURS OF REVISION SURGERY. THE EXPLANTS WERE GIVEN TO THE PATIENT: PRODUCT ID: KIMP312R ADVANCE® II MEDIAL PIVOT TIBIAL INSERT SZ 3 RIGHT 12MM / LOT NO: 1687317/ QTY: (B)(4). PRODUCT ID: KFTCNP3R ADVANCE® PRIMARY FEMORAL SIZE 3 RIGHT NONPOROUS / LOT NO: 1707334/ QTY: (B)(4). PRODUCT ID: KTCCNP30 ADVANCE® II COCR TIBIAL BASE NONPOROUS SZ 3 STANDARD/ LOT NO: 1644600/ QTY: (B)(4).
Patient problems: Insufficient Information
MAUDE report 19147710 ↗InjuryOct 13, 2021Unstable; Unstable
ALLEGEDLY, PATIENT WAS REVISED DUE TO INSTABILITY. REVISION NJR NUMBER: (B)(4). SIDE: R. PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING. COMPONENTS NOT REVISED: ADVANCE PRIMARY FEMORAL SIZE 3 RIGHT NONPOROUS PRODUCT NAME: KFTCNP3R, LOT NUMBER: 7484802. ADVANCE II COCR TIBIA BASE NONPOROUS SZ 3 STANDARD PRODUCT NAME: KTCCNP30, LOT NUMBER: 7350410.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12625078 ↗InjuryJun 20, 2019Unstable; Unstable
ALLEGEDLY, RIGHT "PATELLA OA" IS REASON FOR REVISION, PAIN & INSTABILITY. POLY X 12MM WAS PULLED AND REPLACED WITH A 3X17MM, STRYKER PATELLA WAS IMPLANTED. COMPONENTS NOT REVISED: ADVANCE FEMUR SIZE 3 RIGHT KFTCPC3R LOT 038577450, ADVANCE TIB BASE SIZE 3 KTSSFM31 LOT 127503866, ADVANCE MODULAR KEEL SPKT0023 LOT 018484504, CANCELLOUS BONE SCREW 75520040 X2 LOT 028547359, CANCELLOUS BONE SCREW 75520025 X2 LOT 028551177 & 02851176.
Patient problems: Pain
MAUDE report 8716449 ↗
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page