Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M684PHA002341), received between Feb 20, 2020 and Oct 3, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Corroded1
- Insufficient Device Problem Information1
- Material Disintegration1
Patient problems
- Pain2
- Failure of Implant1
- Metal Related Pathology1
- Osteolysis1
- Reaction1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, PATIENT UNDERWENT REVISION SURGERY ON RIGHT HIP DUE TO PAIN FROM UNDERLYING MECHANICAL ASSISTED CREVICE CORROSION FROM MODULAR, COBALT CHROME RIGHT HIP REPLACEMENT WITH AN ELEVATED COBALT LEVELS AND EVIDENCE OF PSEUDOTUMOR ON MRI AND METALLOSIS. NON MPO PRODUCTS: ZIMMER BIOMET RB 50 MM ACETABULAR COMPONENT, 50/36 POLYLINER, 25 MM ACETABULAR SCREW.
Patient problems: Metal Related Pathology; Pain
MAUDE report 20360482 ↗InjuryAug 3, 2022Break; Break
ALLEGEDLY, FOUR YEARS AFTER THE LEFT THA, RADIOGRAPHS SHOW STEM FRACTURE IN DISTAL PORTION, REVISION HAS BEEN SCHEDULED (B)(6) 2022. THE DISTAL PORTION WILL BE LEFT AS IT IS, AND A CEMENTED STEM FROM ANOTHER COMPANY WILL BE INSERTED. (B)(4).
Patient problems: Failure of Implant
MAUDE report 15160554 ↗InjuryFeb 20, 2020Corroded; Material Disintegration; Corroded; Material Disintegration
ALLEGEDLY, PATIENT PRESENTED ELEVATED COBALT ION LEVELS AND PAIN. THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE, CORROSION ON THE OBLONG TAPER WHERE THE NECK WAS SEATED IN THE POCKET OF THE PROFEMUR TITANIUM STEM, WHICH CAUSED ADDUCTOR DAMAGE, SOFT TISSUE REACTION AND PSEUDOTUMOR REACTION. DURING THE REVISION THE SURGEON REMOVED THE FAILED COMPONENTS OF THE PROFEMUR TOTAL HIP SYSTEM.
Patient problems: Osteolysis; Pain; Reaction
MAUDE report 9733090 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page