Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M684PHA002661), received between Apr 25, 2022 and Apr 25, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Corroded1

Patient problems

  • Loss of Range of Motion1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2022Corroded; Corroded

    ALLEGEDLY, PLAINTIFF PROFEMUR TOTAL HIP SYSTEM IMPLANTED ON (B)(6) 2016 IN HER RIGHT HIP, ON (B)(6) 2021 PLAINTIFF PRESENTED PAIN AND LACK OF MOBILITY, PLAINTIFF NEEDED A REVISION SURGERY BECAUSE THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE. AT REVISION SURGERY, IT WAS DISCOVERED THAT THE DEVICE FAILED DUE TO CORROSION OF THE OBLONG TAPER OF THE PROFEMUR COCR MODULAR NECK WHERE IT SEATED IN THE POCKET OF THE PROFEMUR. ACCORDING TO THE INFORMATION RECEIVED IN THE REPORT, WE DO NOT KNOW IF THE SCREWS, CUP, LINER AND HEAD WERE REVISED.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 14197220 ↗