Profemur PHA01234Recalls & reported problems
Microport Orthopedics Inc. · PROFEMUR NECK 8DG A/R LONG · Class 3 · Implant
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI M684PHA012341), received between Dec 5, 2019 and Nov 4, 2024 — classified as 8 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Insufficient Device Problem Information2
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Corroded1
- Loose or Intermittent Connection1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- No Information4
- Unspecified Infection3
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryNov 4, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, PATIENT WAS REVISED DUE TO INFECTION. SRNJR NUMBER: (B)(6). SIDE: R. GENDER: F. NON MPO REVISED PRODUCTS: PRODUCT ID: PT106056, REGEN/RNGLC+ MULTI 56MM SZ 24. LOT: UNKNOWN. QTY: 1. PRODUCT ID: EP043654, E1 ABT STANDARD ACETAB LINER 36/54MM (SZ 24). LOT: UNKNOWN. QTY: 1.
Patient problems: Unspecified Infection
MAUDE report 20598923 ↗InjuryAug 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, THIS PATIENT WAS REVISED DUE TO INFECTION AND NOT A FAILURE OF THE NECK. ALL OF THE COMPONENTS WERE REVISED DUE TO INFECTION, NOT DUE TO COMPONENT FAILURE. IT HAD BEEN IMPLANTED FOR A NUMBER OF YEARS. THE NECK WAS LEFT ATTACHED TO THE REMOVAL DEVICE AND WAS LEFT IN THE INSTRUMENT TRAY AS AN OVERSIGHT FOLLOWING THE END OF THE SURGERY. ADDITIONAL INFORMATION RECEIVED ON 02/01/2024: SHELL AND LINER LOOKED TO BE ANOTHER SUPPLIER.
Patient problems: Unspecified Infection
MAUDE report 20044374 ↗InjuryOct 30, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ALLEGEDLY, PATIENT WAS REVISED DUE TO DISLOCATION , SUBLUXATION AND MALALIGNMENT STEM. REVISION NJR NUMBER: 4502611. SIDE:R. PRIMARY ASA: P2 - MILD DISEASE NOT INCAPACITATING. NO INFORMATION PROVIDED FOR STEM. ADDITIONAL INFORMATION RECEIVED ON 10/08/2020 FROM (B)(6): ADDING A NEW INCIDENT FOR THE STEM WITH PARTIAL PRODUCT INFORMATION. ADDITIONAL INFORMATION RECEIVED ON 11/09/2020 FROM (B)(6) : ADDING STEM PRODUCT NUMBER AND LOT BATCH. MHRA REFERENCE NO: (B)(4).
Patient problems: No Information
MAUDE report 10765619 ↗InjuryAug 4, 2020Corroded; Corroded
ALLEGEDLY , THE SUSPICION OF ADVERSE REACTION TO METAL DEBRIS MAY BE ESPECIALLY BETWEEN NECK AND HEAD. DR. WANTS TO INVESTIGATE THE EXISTENCE OF ADVERSE REACTION TO METAL DEBRIS AS FOLLOWING. USED OUR PROFEMURXM, CHANGEABLE NECK, METAL HEAD ONLY AND USED DEPUY PINA CUP AND MARASON LINER 28MM DIAMETER FLAT. A FEW DAYS AGO, THE PATIENT COMPLAINED OF ABNORMAL NOISE FROM HIS JOINT SPINNING PREVENTION TAB OF 6 PIN CUP MARATHON LINER 4 OUT OF 6 DAMAGED. REPLACE WITH MARATHON LINER 32MM FLAT. OUR STEAM IS ALSO REPLACED WITH A CHANGEABLE NECK AR8 LONG AND A DELTA HEAD 32 MM LONG. (B)(6).
Patient problems: No Information
MAUDE report 10361476 ↗InjuryMar 31, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ALLEGEDLY, PATIENT WAS REVISED DUE TO PROSTHESIS DISLOCATION.
Patient problems: No Information
MAUDE report 9907455 ↗InjuryMar 6, 2020Loosening of Implant Not Related to Bone-Ingrowth
ALLEGEDLY, PATIENT WAS REVISED DUE TO SEPTIC LOOSENING.
Patient problems: Unspecified Infection
MAUDE report 9797946 ↗InjuryFeb 18, 2020Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available
ALLEGEDLY, PATIENT WAS REVISED DUE TO PAIN WITHOUT LOOSENING. ITEMS NOT REVISED: COTYLE "ANCA FIT?" SANS TROUS A/REVET. HAP 50, CATALOG # PPR67350, LOT # S03101591, QTY 1. INSERT CERAM "ANCA FIT?" 28/40 50-52-54/28 AL2.O3 BIO. FORTE, CATALOG # PPR67510, LOT # S02102352, QTY 1.
Patient problems: Pain
MAUDE report 9722966 ↗InjuryDec 5, 2019Break; Break
ALLEGEDLY, PATIENT REVISED DUE TO FEMORAL NECK FRACTURE.
Patient problems: No Information
MAUDE report 9415914 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.