Skip to content

PROFEMUR® PHAC1214Recalls & reported problems

Microport Orthopedics Inc. · PROFEMUR® NECK A/R VAR/VAL 2 LONG COBALT CHROME · Class 2 · Implant

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI M684PHAC12141), received between Feb 20, 2020 and Feb 28, 2024 — classified as 6 injuries.

The most frequently reported specific device problem is “Corroded” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded4
  • Insufficient Device Problem Information2
  • Material Disintegration1

Patient problems

  • Pain3
  • Unspecified Infection2
  • No Known Impact Or Consequence To Patient1
  • Reaction1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, INFECTION SIDE: L REVISION NJR: (B)(4). GENDER: M EXPLANTED: FEMORAL STEM, ACETABULAR CUP, MODULAR HEAD, ACETABULAR LINER

    Patient problems: Unspecified Infection

    MAUDE report 18798368 ↗
  • InjuryFeb 7, 2022Corroded; Corroded

    ALLEGEDLY, THE PATIENT WAS REVISED DUE TO CORROSION OF THE MODULAR NECK AND PAIN(LEFT HIP).

    Patient problems: Pain

    MAUDE report 13470311 ↗
  • InjuryApr 21, 2021Corroded; Corroded

    ALLEGEDLY, PATIENT EXPERIENCED PAIN DUE TO CORROSION AT THE COCR MODULAR NECK.

    Patient problems: Pain

    MAUDE report 11703685 ↗
  • InjurySep 8, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED DUE TO DEEP WOUND INFECTION. LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 10503134 ↗
  • InjuryAug 13, 2020Corroded; Corroded

    ALLEGEDLY, PATIENT WAS REVISED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10401964 ↗
  • InjuryFeb 20, 2020Corroded; Material Disintegration; Corroded; Material Disintegration

    ALLEGEDLY, PATIENT PRESENTED ELEVATED COBALT ION LEVELS AND PAIN. THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE, CORROSION ON THE OBLONG TAPER WHERE THE NECK WAS SEATED IN THE POCKET OF THE PROFEMUR TITANIUM STEM, WHICH CAUSED ADDUCTOR DAMAGE, SOFT TISSUE REACTION AND PSEUDOTUMOR REACTION. DURING THE REVISION THE SURGEON REMOVED THE FAILED COMPONENTS OF THE PROFEMUR TOTAL HIP SYSTEM.

    Patient problems: Pain; Reaction

    MAUDE report 9733174 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.