The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI M684PHAC12241), received between Mar 4, 2020 and Apr 10, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Corroded2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Noise, Audible1
  • Unstable1

Patient problems

  • Pain4
  • Insufficient Information3
  • Metal Related Pathology3
  • Loss of Range of Motion1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 10, 2026Break; Noise, Audible

    ALLEGEDLY, ON (B)(6) 2024, THE PATIENT WAS STANDING IN HIS LIVING ROOM AND BENT OVER TO PICK UP A PHONE CHARGER WHEN HE FELT A SUDDEN POP IN HIS RIGHT HIP AND IMMEDIATELY FELL TO THE FLOOR. HE EXPERIENCED PAIN AND WAS UNABLE TO BEAR WEIGHT ON HIS RIGHT LEG. X-RAYS REVEALED AN ANGULATED FRACTURE OF THE FEMORAL COMPONENT OF HIS RIGHT HIP ARTHROPLASTY HARDWARE, LATER CONFIRMED AS A FRACTURED MODULAR NECK STEM, WHICH NECESSITATED REVISION SURGERY PERFORMED ON (B)(6) 2024. NON-REVISED PRODUCTS: PRODUCT ID: DSBFGG58 DYNASTY® BF SHELL 58MM GROUP G/ LOT NO.: 1512229/ QTY: (B)(4). PRODUCT ID: 18080304 CANCELLOUS SELF-TAPPING 6.5MM BONE SCREW 3.5CM LENGTH/ LOT NO.: 1462885/ QTY: (B)(4).

    Patient problems: Pain

    MAUDE report 24841247 ↗
  • InjuryFeb 28, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY, DISLOCATION/SUBLUXATION. SIDE: L. REVISION NJR: 2865282. GENDER: F. EXPLANTED: FEMORAL STEM.

    Patient problems: Insufficient Information

    MAUDE report 18796275 ↗
  • InjuryApr 17, 2023Corroded; Corroded

    ALLEGEDLY, PROFEMUR® TOTAL HIP SYSTEM IMPLANTED IN THE LEFT SIDE HAS FAILED DUE TO CORROSION RELATED METALLOSIS/ADVERSE TISSUE REACTIONS. PATIENT PRESENTED WITH SEVERE PAIN, LACK OF MOBILITY, AND ELEVATED COBALT LEVELS.

    Patient problems: Loss of Range of Motion; Metal Related Pathology; Pain

    MAUDE report 16750891 ↗
  • InjuryFeb 27, 2023Break; Break

    ALLEGEDLY, FRACTURE OF A MODULAR NECK (AR VV1 LONG) ON (B)(6) 2023. REVISION SURGERY ON (B)(6) 2023 (AT CLINIC (B)(6), DR. (B)(6)). ADDITIONAL INFORMATION 02/05/2023: ALLEGEDLY, PATIENT SUFFERED A SPONTANEOUS FRACTURE OF THE MODULAR NECK WHILE WALKING. THERE WERE NO PRIOR INDICATIONS OF THIS OCCURRENCE, NO PAIN OR RESTRICTION OF MOVEMENT. SINCE THE BROKEN NECK COULD NOT BE REMOVED DURING THE REOPERATION, THE PROSTHESIS SHAFT HAD TO BE REVISED AND THE FEMUR OPENED. A REVISION HIP STEM WAS USED. THE SURGEON DR. (B)(6) HAS BEEN USING MPO (FORMERLY WRIGHT) MODULARITY FOR ALMOST 30 YEARS AND AFFIRMED THAT HE HAD FOLLOWED ALL SPECIFICATIONS REGARDING IMPLANT "MODULARITY". THE INCIDENT IS REPORTED BY DR. (B)(6) REPORTED TO THE RESPONSIBLE AUTHORITIES (FEDERAL INSTITUTE FOR DRUGS AND MEDICAL DEVICES, BFARM). SINCE THE PATIENT WAS AND IS COMPLIANT, THE IMPLANT FAILURE IS INEXPLICABLE FOR THE PATIENT, AND HE INDICATED TO THE SURGEON THAT LEGAL ACTION SHOULD BE TAKEN AGAINST THE SURGEON AND THE MANUFACTURER. AFTER THE PATIENT'S REHAB, AT THE END OF FEBRUARY, HE ALSO EXPRESSED CONTACT WITH HIS LEGAL COUNSEL. HE ALSO INDICATED THAT HE COULD IMAGINE AN OUT-OF-COURT SETTLEMENT.

    Patient problems: Insufficient Information

    MAUDE report 16443317 ↗
  • InjuryNov 7, 2022Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, PATIENT WAS REVISED DUE TO ADVERSE SOFT TISSUE REACTION TO PARTICULATE DEBRIS REVISION NJR NUMBER: (B)(4). SIDE:R. PRIMARY ASA: P1 - FIT AND HEALTHY.

    Patient problems: Metal Related Pathology

    MAUDE report 15743495 ↗
  • InjuryJun 7, 2022Corroded; Corroded

    ALLEGEDLY, ON (B)(6) 2015, PATIENT HAD A RIGHT TOTAL HIP ARTHROPLASTY. PATIENT RECEIVED A PROFEMUR® LONG COBALT CHROME MODULAR NECK (PHAC1224) AND A PROFEMUR® Z FEMORAL STEM MADE OF TITANIUM ALLOY. ON (B)(6) 2020 DUE TO THE FAILURE OF THE DEVICE, PATIENT'S DEVICE WAS SURGICALLY REMOVED BY JAKUB TATKA, M.D., AT NEW YORK PRESBYTERIAN LAWRENCE HOSPITAL IN BRONXVILLE, NY IN A SURGICAL PROCEDURE COMMONLY CALLED A "REVISION." DURING HER REVISION SURGERY, DR. TATKA NOTED "METALLOSIS AND CORROSION CHANGES ON THE TAPER OF THE NECK AND BODY OF THE STEM TAPER" AND EXCISED METALLOSIS AND A PSEUDOTUMOR. DUE TO THE FAILURE OF HER PROFEMUR® TOTAL HIP SYSTEM, DR. TATKA NEEDED TO REMOVE THE WELL-FIXED FEMORAL STEM. PATIENT HAS ALSO SUSTAINED PAIN. THE CUP AND LINER HAVE NOT BEEN INDICATED IN THE REPORT TO BE MICROPORT PRODUCTS . THE CUP, LINER AND HEAD HAVE NOT BEEN INDICATED AS REVISED COMPONENTS.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 14621992 ↗
  • InjuryFeb 24, 2021Break; Break

    ALLEGEDLY, LEFT NECK BREAKAGE. BASG REF.# (B)(4). ADDITIONAL INFORMATION RECEIVED ON 3/1/2021: INCIDENT DESCRIPTION, PATIENT INFORMATION, ADDITIONAL COMPONENTS ID'S AND LOT NUMBERS, AND IMPLANT AND EXPLANT DATE. ALLEGEDLY, THE PATIENT WALKS ACROSS THE STREET ON (B)(6) 2021 AND REPORTS THAT HE FELL WITHOUT ANY EXTERNAL CAUSE. AFTERWARDS THE LEG NO LONGER RESILIENT.

    Patient problems: Insufficient Information

    MAUDE report 11371637 ↗
  • InjuryMar 4, 2020Break; Unstable; Break; Unstable

    ALLEGEDLY, PROFEMUR COCR NECK FRACTURED - RIGHT HIP. MECHANICAL COMPLICATION, PAIN AND INSTABILITY ON THE HIP DUE TO THE BROKEN NECK ARE REPORTED.

    Patient problems: Pain

    MAUDE report 9787957 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →