Skip to content

PROFEMUR® PHAC1244Recalls & reported problems

Microport Orthopedics Inc. · PROFEMUR® NECK A/R 15DG LONG COBALT CHROME · Class 2 · Implant

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI M684PHAC12441), received between Nov 21, 2019 and Apr 3, 2025 — classified as 6 injuries.

The most frequently reported specific device problem is “Corroded” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded3
  • Insufficient Device Problem Information2
  • Device Dislodged or Dislocated1
  • Naturally Worn1

Patient problems

  • Pain3
  • Insufficient Information1
  • Loss of Range of Motion1
  • Metal Related Pathology1
  • No Information1
  • Reaction1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, ON (B)(6) 2024, THE PATIENT REPORTED COMPLICATIONS FROM THE REVISION DEVICES. SECOND REVISION SURGERY EXPECTED. NO OTHER INFORMATION AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 21769049 ↗
  • InjurySep 26, 2024Corroded; Naturally Worn; Corroded; Naturally Worn

    ALLEGEDLY, THE DEVICES FAILED DUE TO CORROSION OF THE MODULAR NECK AT THE HEAD-NECK JUNCTION AND WHERE THE NECK SEATED IN THE POCKET OF THE STEM, AND EARLY EXCESSIVE WEAR OF THE LINER. ALSO, PRODUCTION OF METAL DEBRIS RESULTING IN METALLOSIS, POLYETHYLENE WEAR DEBRIS, ADVERSE LOCAL TISSUE REACTION TO PARTICLES, WEAKNESS, AND PAIN.

    Patient problems: Pain

    MAUDE report 20311960 ↗
  • InjuryApr 27, 2022Corroded; Corroded

    ALLEGEDLY, THE PATIENT PRESENTED WITH PAIN AND LACK OF MOBILITY. THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION. AT REVISION SURGERY, IT WAS DISCOVERED THAT THE DEVICE FAILED DUE TO CORROSION OF THE OBLONG TAPER OF THE PROFEMUR COCR MODULAR NECK WHERE IT SEATED IN THE POCKET OF THE PROFEMUR TITANIUM STEM.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 14218899 ↗
  • InjuryApr 21, 2021Corroded; Corroded

    ALLEGEDLY, THE PATIENT HAD A SECOND REVISION SINCE THE COCR MODULAR NECK FAILED. THE PATIENT PRESENTED PAIN, ELEVATED COBALT ION LEVELS, ADVERSE TISSUE REACTION AND CORROSION.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 11703680 ↗
  • InjuryNov 21, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY, THE PATIENT WAS REVISED DUE TO DISLOCATION. THE DOCTOR PULLED THE NECK, HEAD, SLEEVE, LINER, AND 3 SCREWS AND CEMENTED IN A BIOMET CONSTRAINED LINER. THE PATIENT HAD A PREVIOUS REVISION FOR MOM, INCIDENT GROUP 19060012, A SECOND REVISION FOR INFECTION, INCIDENT GROUP 19060187, AND A THIRD REVISION FOR DISLOCATION, INCIDENT GROUP 19060015. ITEMS NOT REVISED: DYNASTY BIOFOAM SHELL DSBFGG60 LOT 079859623, PROFEMUR PLASMA Z STEM PHA00270 LOT 067445032.

    Patient problems: No Information

    MAUDE report 9358820 ↗
  • InjuryNov 21, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED FROM A TOTAL HIP DUE TO A PSUEDO-TUMOR.

    Patient problems: Reaction

    MAUDE report 9354515 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.