Profemur PLS0R411Recalls & reported problems
Microport Orthopedics Inc. · PROFEMUR RENAISSANCE STEM REDUCED SZ 11 · Class 3 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M684PLS0R4111), received between Jul 15, 2021 and Oct 11, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Difficult to Remove1
- Unstable1
Patient problems
- Metal Related Pathology1
- Osteopenia/ Osteoporosis1
- Unspecified Tissue Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 11, 2022Difficult to Remove; Unstable; Device Dislodged or Dislocated; Difficult to Remove; Unstable; Device Dislodged or Dislocated
ALLEGEDLY, ON OR ABOUT APRIL 30,204 PATIENT HAD A PROFEMUR TOTAL HIP SYSTEM IMPLANTED IN HER RIGHT HIP. ON OR ABOUT (B)(6) 2019 PATIENT REPORTED FOR REVISION SURGERY DUE TO ELEVATED COBAL AND CHROMIUN LEVELS, PAIN AND EVIDENCE OF A PSEUDOTUMOR. PATIENT'S FIRST REVISION WAS NECESSARY BECAUSE THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION. IT WAS DISCOVERED CORROSION OF THE PROFEMUR COCR MODULAR NECK WHERE IT SEATED IN THE POCKET OF THE MODULAR STEM. THE IMPLANTED NECK AND HEAD WERE REVISED. AFTER THE REVISION SURGERY PATIENT REPORTED AT LEAST 3 DISLOCATIONS. THE PREVIOUS EVENTS WERE CAPTURED UNDER MICROPORT ORTHOPEDICS GROUP NUMBER: 20060165. ADDITIONAL INFORMATION RECEIVED (B)(6) 2022: ALLEGEDLY, ON OR ABOUT (B)(6) 2021, PATIENT REPORTER FOR A REVISION SURGERY OF THE RIGHT TOTAL HIP ARTHROPLASTY, REVISION OF BOTH THE ACETABULAR AND FEMORAL COMPONENTS. SURGEON RECOMMENDED THIS SECOND REVISION SURGERY DUE TO RECURRENT INSTABILITY WITH DISLOCATIONS. DURING REVISION SURGERY, SURGEON DETERMINED THE ONLY WAY TO SAFELY REMOVE THE MODULAR STEM WAS TO PERFORM AN EXTENDED TROCHANTERIC OSTEOTOMY(ETO), WHICH REQUIRED FOUR DALL MILES CABLES WITH THE USE OF A 240 MM REVISION STEM.
Patient problems: Metal Related Pathology; Osteopenia/ Osteoporosis
MAUDE report 15583485 ↗InjuryJul 15, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
ALLEGEDLY, PATIENT REVISED DUE TO COPIOUS AMOUNT OF BLACK DISCOLORED TISSUE, BLACK DISCOLORATION OF THE JOINT CAPSULE CONSISTENT WITH METALLOSIS AND DISLOCATION. (LEFT HIP)
Patient problems: Unspecified Tissue Injury
MAUDE report 12178681 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.