Skip to content

PROFEMUR PPRCLS08Recalls & reported problems

Microport Orthopedics Inc. · Class 2 · Implant

All models of PROFEMUR (116)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M684PPRCLS081), received between Jul 6, 2020 and Jul 6, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Unspecified Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 6, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY BREAKAGE/SPLINTERING OF THE LINER (RIM SPLINTRED, CERAMIC PARTICLES WITHIN THE TISSUE). X-RAY BEFORE SURGERY SHOWED CERAMIC PARTICLES NEAR FEMORAL NECK. THE PATIENT WAS REVISED PRIMARILY DUE TO A LATE LOW-GRADE INFECTION (STAPH. CAPRAE), NOT DUE TO THE CERAMIC BREAKAGE. INTRAOPERATIVELY, CERAMIC INSERT SHOWED SMALL LESION ON RIM AT THREE PLACES. SOME CERAMIC PARTICLES REMOVED DURING EXPLANTATION. CUP AND STEM WERE WELL FIXED IN BONE. METALBACK DAMAGED DURING EXPLANTATION. (B)(6), (B)(4).

    Patient problems: Unspecified Infection

    MAUDE report 10233912 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page