Skip to content

Profemur PPW38002Recalls & reported problems

Microport Orthopedics Inc. · PROFEMUR 135MM STEM STRAIGHT/TAPER SIZE 12 · Class 3 · Implant

All models of PROFEMUR (116)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M684PPW380021), received between Sep 18, 2021 and Sep 18, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Loss of Osseointegration1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2021Break; Loss of Osseointegration; Break; Loss of Osseointegration

    ALLEGEDLY, STEM IS BROKEN AND SEEMS LOOSE.

    Patient problems: Insufficient Information

    MAUDE report 12493248 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page