Profemur PPW38024Recalls & reported problems
Microport Orthopedics Inc. · PROFEMUR 175MM STEM BOWED/TAPER SIZE 14 · Class 3 · Implant
All models of PROFEMUR (116)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M684PPW380241), received between Oct 7, 2021 and Jan 5, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Corroded1
- Difficult to Remove1
Patient problems
- Insufficient Information1
- Metal Related Pathology1
- Osteolysis1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 5, 2023Break; Corroded; Difficult to Remove; Break; Corroded; Difficult to Remove
ALLEGEDLY, ON (B)(6) 2010, PATIENT HAD A WRIGHT MEDICAL PROFEMUR R BOWED DISTAL STEM AND PROXIMAL BODY, MATED WITH A PROFEMUR PLUS COCR MODULAR NECK, IMPLANTED IN HIS LEFT HIP. ON (B)(6) 2020, PATIENT'S PROFEMUR R BOWED DISTAL STEM SUDDENLY AND WITHOUT WARNING FAILED, FRACTURING IN TWO NEAR THE TOP OF THE STEM, BELOW THE MORSE TAPER. ON (B)(6) 2020, PATIENT HAD A REVISION SURGERY PERFORMED ON HIS LEFT. IN THE COURSE OF THE (B)(6) 2020, REVISION SURGERY THE DOCTOR NOTED THAT THE PROXIMAL BODY AND HEAD WERE GROSSLY LOOSE AND WERE EXTRACTED BY HAND. SURGEON PERFORMED AN EXTENDED TROCHANTERIC OSTEOTOMY TO REMOVE THE DISTAL REMNANT OF THE FRACTURED PROFEMUR R BOWED DISTAL STEM. IN THE (B)(6) 2020, REVISION SURGERY THE WRIGHT MEDICAL PROFEMUR R BOWED DISTAL STEM, PROFEMUR PROXIMAL BODY, PROFEMUR PLUS COCR MODULAR NECK, AND CERAMIC FEMORAL HEAD WERE REMOVED FROM THE BODY OF THE PATIENT, AND REPLACED WITH STRYKER ARTIFICIAL DEVICES, INCLUDING A STRYKER RESTORATION MODULAR SIZE 15 195MM CURVED STEM, 25MM +0 PROXIMAL BODY AND A 32MM +4 DELTA CERAMIC HEAD. IN THE COURSE OF THE (B)(6) 2020, REVISION SURGERY DOCTOR FOUND THE ACETABULAR COMPONENT (THE CUP) TO BE WELL FIXED WITH A CERAMIC LINER A
Patient problems: Metal Related Pathology; Osteolysis; Pain
MAUDE report 16108265 ↗InjuryOct 7, 2021Break; Break
ALLEGEDLY, THE PATIENT HAD PREVIOUSLY BEEN OPERATED ON TWICE, IN 2013 AND 2014. THIS TIME HE SUFFERED A STEM BREAK, HAVING TO UNDERGO SURGERY AGAIN FOR ITS REMOVAL. ADDITIONAL INFORMATION RECEIVED ON 03/01/2022 INCLUDING NEW PRODUCTS THAT WERE REVISED.
Patient problems: Insufficient Information
MAUDE report 12591208 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page