PROFEMUR® PRGL0003Recalls & reported problems
Microport Orthopedics Inc. · PROFEMUR® GLADIATOR® PLASMA STEM SZ 3 · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M684PRGL00031), received between May 4, 2022 and May 4, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Corroded1
Patient problems
- Metal Related Pathology1
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 4, 2022Corroded; Corroded
ALLEGEDLY, THE PATIENT PRESENTED WITH PAIN AND ELEVATED COBALT AND CHROMIUM LEVELS. THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION. AT REVISION SURGERY, IT WAS DISCOVERED THAT THE DEVICE FAILED DUE TO CORROSION OF THE OBLONG TAPER OF THE PROFEMUR COCR MODULAR NECK AND THE DOCTOR IDENTIFIED IN THE DIAGNOSES PREOPERATIVE OF THE REVISION SURGERY 'PAINFUL LEFT TOTAL HIP ARTHOPLASTY WITH METALLOSIS.
Patient problems: Metal Related Pathology; Pain
MAUDE report 14278092 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.