Calix Spinal Implant System X034-0387Recalls & reported problems
Xtant Medical Holdings, Inc. · 26mm x 10mm x 6mm x 5° PLIF ASSY · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M697X03403871), received between Jul 19, 2023 and Jul 19, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 19, 2023Material Fragmentation; Material Fragmentation
THE MANUFACTURER WAS MADE AWARE OF A PRODUCT COMPLAINT ON 6/19/2023. IT WAS REPORTED THAT A 26MM X 10MM X 6MM X 5° PLIF IMPLANT MALFUNCTIONED WHEN BEING PLACED DURING A PROCEDURE. THERE WERE NO KNOWN PATIENT COMPLICATIONS OR DELAY IN TREATMENT ASSOCIATED WITH THIS COMPLAINT. THE MALFUNCTIONED IMPLANT WAS REMOVED, AND AN ALTERNATE AVAILABLE IMPLANT WAS SUCCESSFULLY UTILIZED TO COMPLETE THE SURGICAL PROCEDURE. THE COMPLAINT IMPLANT WAS DISCARDED AT THE SURGICAL FACILITY AND NOT AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR ASSESSMENT. THE COMPLAINANT PROVIDED PHOTOS OF THE COMPLAINT IMPLANT, WHICH SHOWED THE PROXIMAL END THAT ENGAGES WITH THE SYSTEM INSERTER HAD FRACTURED FROM THE BODY OF THE IMPLANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17355314 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page