Axle Interspinous Fusion System X060-0080Recalls & reported problems
Xtant Medical Holdings, Inc. · Axle Locking Plate Assy, 36, 8mm · Class 2 · Implant
All models of Axle Interspinous Fusion System (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M697X06000801), received between Oct 20, 2021 and Oct 20, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 20, 2021Material Fragmentation; Material Fragmentation
THE COMPANY WAS NOTIFIED ON 9/28/2021 OF AN IMPLANT MALFUNCTION THAT OCCURRED AFTER PLACING AN INTERSPINOUS PLATE CONSTRUCT. AFTER IMPLANT PLACEMENT, THE PHYSICIAN ELECTED TO MAKE A POSITION ADJUSTMENT OF THE IMPLANTS. WHEN THE SET SCREW OF A LOCKING PLATE ASSEMBLY WAS BEING LOOSENED, IT SHEARED OFF. THE IMPLANT CONSTRUCT WAS REMOVED AND SUCCESSFULLY REPLACED WITH ALTERNATE IMPLANTS WITHOUT ANY FURTHER ISSUES. THERE WERE NO KNOWN PATIENT COMPLICATIONS ASSOCIATED WITH THIS COMPLAINT. THE COMPLAINT DEVICE ARRIVED AT THE MANUFACTURER FOR ASSESSMENT ON 10/01/2021.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12668531 ↗
Other Spinous process plate devices
- UniVise 48590035Stryker Corporation · Spinous process plate1 report
- ROMEO 2 PAD POSTERIOR AXIAL DEVICE PAD-IM TI 08-SSpineArt SA · Spinous Process Plate1 report
- Axle Interspinous Fusion System X060-0241Xtant Medical Holdings, Inc. · Spinous process plate1 report
- ZIP ULTRAAurora Spine, Inc. · Spinous Process Plate1 report
- ZIP LPAurora Spine, Inc. · Spinous Process Plate1 report
- ZIP LPAurora Spine, Inc. · Spinous Process Plate1 report
- CoFix Implant 10 mm RQI00010Paradigm Spine GmbH · Spinous Process Plate1 report
- SP-FIX 588.026Globus Medical, Inc. · Spinous process plate1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.