Xpress Minimally Invasive Pedicle Screw System X073-5550Recalls & reported problems
Xtant Medical Holdings, Inc. · XPRESS SYSTEM PEDICLE SCREW ASSY, 5.5 x 50mm · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M697X07355501), received between Feb 6, 2018 and Feb 6, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
1 other recall cover devices cleared under the same FDA submission (K152132); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2018Fracture; Fracture
ON (B)(4) 2017, X-SPINE SYSTEMS, INC. RECEIVED EMAIL COMMUNICATIONS REPORTING BROKEN PEDICLE SCREWS. THE EMAIL CONTAINED SIX X-RAYS. THERE WERE NO ADDITIONAL DETAILS IN THE EMAIL COMMUNICATION. X-SPINE MADE THREE ATTEMPTS TO OBTAIN ADDITIONAL COMPLAINT DETAILS. ADDITIONAL DETAILS WERE RECEIVED BY THE COMPLAINANT ON (B)(6) 2018. THE COMPLAINANT STATED THAT "THE DOCTOR SENT THE PATIENT FOR A CONTROL X-RAY AND HE FOUND THAT THE SCREW WAS BROKEN, THE PATIENT NEVER HAD AN INCIDENT THAT MIGHT CAUSE THE BREAKAGE OF THE SCREW." THESE DETAILS CORRESPONDED WITH ONE OF THE X-RAYS. THE REMAINING X-RAYS DEPICT MULTIPLE CASES AND POSSIBLY MULTIPLE PATIENTS, HOWEVER, THE COMPLAINANT ONLY INDICATED THAT IT WAS ONE CASE/ONE PATIENT. X-SPINE WAS UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE MULTIPLE CASES/PATIENTS. A COMPREHENSIVE REVIEW OF ALL SUPPLIED DATA WAS COMPLETED ON (B)(6) 2018. THE INVESTIGATION INCLUDED A REVIEW OF THE DHF FOR THE SINGLE LOT NUMBER PROVIDED. THE REVIEW WAS UNREMARKABLE AND THERE WERE NO INDICATIONS THAT THE PART WOULD NOT PERFORM AS INTENDED. THE ASSESSMENT OF THE CORRESPONDING X-RAY (#1) SHOWS A BROKEN PEDICLE SCREW SHANK ON THE CAUDAL END OF A 1 LEVEL CONSTRUCT.
Patient problems: No Code Available
MAUDE report 7244246 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K152132. One submission often covers many models, so check the product and lot details in each record.
- Z-1773-2026 · Feb 18, 2026 · Open, ClassifiedXTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page