Alto™ Abdominal Stent Graft System TV-AB2380-N (UDI M701TVAB2380N1)Recalls & reported problems
Endologix LLC · The Alto™ Abdominal Stent Graft System is an endovascular device delivered via a low-profile catheter to treat abdominal aortic aneurysms (AAAs). The stent graft is designed to reline the diseased vasculature, providing an endovascular blood conduit for isolating the aneurysm from the high-pressure flow of blood, thereby reducing the risk of rupture. The stent graft is a modular configuration comprised of an aortic body section, iliac limbs, and iliac extensions as required. The Alto™ Abdominal Stent Graft System includes: an Aortic Body Stent Graft and delivery catheter, Ovation iX™ Iliac Limb Stent Grafts and delivery catheters, Ovation iX™ Iliac Extension Stent Grafts and delivery catheters (as required), the CustomSeal™ Polymer Fill Kit, and the Autoinjector 2. · Class 3 · Implant
All models of Alto Abdominal Stent Graft System (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M701TVAB2380N1), received between Oct 5, 2020 and Apr 16, 2021 — classified as 1 death, 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Patient-Device Incompatibility2
- Device-Device Incompatibility1
- Difficult to Advance1
- Failure to Advance1
- Improper or Incorrect Procedure or Method1
- Material Puncture/Hole1
Patient problems
- Failure of Implant3
- Aneurysm1
- Death1
- Heart Failure1
- Pseudoaneurysm1
- Rupture1
- Stenosis1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 16, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
THE PATIENT WAS INITIALLY IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY FIVE MONTHS POST INITIAL PROCEDURE DURING ROUTINE FOLLOW UP, AN ULTRASOUND REVEALED SAC GROWTH. AN ANGIOGRAM WAS PERFORMED AND REVEALED A PSEUDO ANEURYSM ABOVE THE RENAL ARTERIES THAT CONTINUED BELOW THE RINGS OF THE ALTO DEVICE. THE PSEUDO ANEURYSM DID COMMUNICATE WITH THE PRIMARY FUSIFORM ANEURYSM, HOWEVER IT IS NOT CATEGORIZED AS A TYPE 1A ENDOLEAK. THE PHYSICIAN ELECTED TO TREAT THIS EVENT BY PLACING A NUMBER OF PNUMBRA (NON-ENDOLOGIX) COILS. THE PATIENT IS CURRENTLY BEING MONITORED TO CONFIRM THE EVENT HAS BEEN RESOLVED.
Patient problems: Aneurysm; Pseudoaneurysm
MAUDE report 11682740 ↗InjuryJan 28, 2021Patient-Device Incompatibility; Insufficient Device Problem Information; Patient-Device Incompatibility; Insufficient Device Problem Information
THE PATIENT WAS INITIALLY IMPLANTED WITH AN OVATION STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY (4) WEEKS POST INITIAL IMPLANT THE PATIENT PRESENTED FOR A ROUTINE VISIT AND A TYPE 1A ENDOLEAK WAS IDENTIFIED. AN INTERVENTION IS BEING PLANNED AND THE PATIENT IS BEING MONITORED.
Patient problems: Failure of Implant; Stenosis
MAUDE report 11245628 ↗DeathDec 3, 2020Material Puncture/Hole; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Device-Device Incompatibility; Insufficient Device Problem Information; Material Puncture/Hole; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Device-Device Incompatibility; Insufficient Device Problem Information
THE PATIENT WAS IMPLANTED WITH THE OVATION ALTO ABDOMINAL STENT GRAFT SYSTEM AND THREE (3) OVATION IX ILIAC LIMBS TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). IT SHOULD BE NOTED THAT THERE WAS A PREVIOUSLY PLACED GORE (NON-ENDOLOGIX) LIMB IN THE RIGHT COMMON COVERING THE COILED HYPO AND EXTENDING INTO THE EXTERNAL. FINAL ANGIOGRAPHY IDENTIFIED AN INTRAOPERATIVE TYPE 1B ENDOLEAK ON THE RIGHT. THIS WAS EXTENDED WITH A VIABHON (NON-ENDOLOGIX) STENT. SUBSEQUENT ANGIOGRAPHY SHOWED A BLUSH, POSSIBLE INTRAOPERATIVE TYPE IIIA ENDOLEAK, IN ILIAC LIMB. THIS WAS RESOLVED WITH IMPLANT OF AN OVATION IX ILIAC LIMB. DURING ADVANCEMENT OF A 12X4 BALLOON IN THE LEFT COMMON ILIAC, THE ILIAC LIMB THAT WAS IN THE EXTERNAL ON THE LEFT (COILED HYPO AND EXTENDED INTO EXTERNAL ON THE LEFT) WAS PUSHED UP BY THE SHEATH AND INTO THE COMMON AS THE SHEATH ON THE LEFT HAD RETRACTED BACK PUSHED INTO THE ARTERY AND CATCHING THE BOTTOM OF LIMB GRAFT IN THE LEFT COMMON THIS WAS BALLOONED, SEAL WAS RE-ATTAINED. THE FINAL ANGIOGRAPHY SHOWED A NEW INTRAOPERATIVE TYPE 1A ENDOLEAK. PERFORMED BALLOONING WITH A 20MM ZMED SUBSEQUENT ANGIOGRAPHY SHOWED THE LEAK WAS STILL FAINT. AN ADDITIONAL PALMAZ (NON-ENDOLOGIX) STENT WAS
Patient problems: Death; Failure of Implant; Heart Failure; Rupture
MAUDE report 10942430 ↗InjuryOct 5, 2020Failure to Advance; Difficult to Advance; Insufficient Device Problem Information; Failure to Advance; Difficult to Advance; Insufficient Device Problem Information
THE PATIENT WAS BEING TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH IMPLANT OF AN OVATION ALTO ABDOMINAL STENT GRAFT AND TWO OVATION IX ILIAC LIMBS. POST DEPLOYMENT AND POLYMER FILL AN INTRAOPERATIVE TYPE 1A ENDOLEAK WAS IDENTIFIED. AN ATTEMPT WAS MADE TO BALLOON THE TYPE 1A ENDOLEAK; HOWEVER, THE INTEGRAL BALLOON OF THE OVATION ALTO DELIVERY SYSTEM WOULD NOT ADVANCE PAST THE SIGNIFICANT ANGLE AT THE GRAFT BIFURCATION. PATIENT IS STABLE AND WILL CONTINUE TO BE MONITORED.
Patient problems: Failure of Implant
MAUDE report 10627665 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.
- Z-2263-2020 · May 6, 2020 · TerminatedOvation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Z-0125-2015 · Sep 20, 2014 · TerminatedOvation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
- Z-0288-2013 · Oct 19, 2012 · TerminatedTriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page