Alto™ Abdominal Stent Graft System TV-AB2980-N (UDI M701TVAB2980N1)Recalls & reported problems
Endologix LLC · The Alto™ Abdominal Stent Graft System is an endovascular device delivered via a low-profile catheter to treat abdominal aortic aneurysms (AAAs). The stent graft is designed to reline the diseased vasculature, providing an endovascular blood conduit for isolating the aneurysm from the high-pressure flow of blood, thereby reducing the risk of rupture. The stent graft is a modular configuration comprised of an aortic body section, iliac limbs, and iliac extensions as required. The Alto™ Abdominal Stent Graft System includes: an Aortic Body Stent Graft and delivery catheter, Ovation iX™ Iliac Limb Stent Grafts and delivery catheters, Ovation iX™ Iliac Extension Stent Grafts and delivery catheters (as required), the CustomSeal™ Polymer Fill Kit, and the Autoinjector 2. · Class 3 · Implant
All models of Alto Abdominal Stent Graft System (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 2
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI M701TVAB2980N1), received between Feb 3, 2021 and Nov 10, 2022 — classified as 2 deaths, 5 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Patient-Device Incompatibility2
- Adverse Event Without Identified Device or Use Problem1
- Improper or Incorrect Procedure or Method1
- Malposition of Device1
- Material Puncture/Hole1
- Positioning Failure1
- Use of Device Problem1
Patient problems
- Failure of Implant6
- Aneurysm1
- Clinical Endoleak1
- Death1
- Rupture1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryNov 10, 2022Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY ONE AND A HALF (1.5) YEARS POST INITIAL PROCEDURE A POSSIBLE TYPE 3B ENDOLEAK AND SAC GROWTH WERE NOTED. THE PATIENT IS CURRENTLY BEING MONITORED WHILE AN INTERVENTION IS BEING DISCUSSED.
Patient problems: Aneurysm; Clinical Endoleak; Failure of Implant
MAUDE report 15775512 ↗InjuryAug 10, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). DURING THIS INITIAL PROCEDURE, THERE WAS A TYPE 1A ENDOLEAK. THE PHYSICIAN ELECTED TO IMPLANT A NON ENDOLOGIX (PALMAZ) STENT. THE ENDOLEAK IMPROVED BUT IT DID NOT COMPLETELY RESOLVE. THE PATIENT IS CURRENTLY BEING MONITORED AND THE ENDOLEAK WILL BE RE-EVALUATED AT THE 30 DAY FOLLOW UP VISIT. SUBSEQUENT TO THE INITIAL REPORT, ADDITIONAL INFORMATION WAS PROVIDED STATING THAT THE TYPE IA ENDOLEAK HAS RESOLVED ON ITS OWN AND REINTERVENTION WAS NOT REQUIRED.
Patient problems: Failure of Implant
MAUDE report 12305342 ↗InjuryJul 2, 2021Positioning Failure; Positioning Failure
THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). IT WAS REPORTED THAT WHEN THE POLYMER WAS DELIVERED, THE CONTRALATERAL LIMB DID NOT DEPLOY. ALL TROUBLESHOOTING STEPS WERE PERFORMED; HOWEVER, THE CONTRALATERAL LIMB DID NOT DEPLOY SO THE POLYMER WAS NOT ABLE TO FILL THE LIMB. DEPLOYMENT OF THE IPSILATERAL LIMB WAS COMPLETED AND THE ANEURYSM WAS SEALED WITHOUT AN ENDOLEAK. THE PHYSICIAN THEN PERFORMED A FEMORAL-FEMORAL BYPASS TO PROVIDE BLOOD FLOW TO BOTH LOWER EXTREMITIES. THE PATIENT WAS REPORTED AS DOING WELL AND BOTH LOWER EXTREMITIES HAVE PALPABLE PULSES.
Patient problems: Failure of Implant
MAUDE report 12110984 ↗InjuryJun 15, 2021Malposition of Device; Patient-Device Incompatibility; Malposition of Device; Patient-Device Incompatibility
THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). IT WAS NOTED THAT THE PATIENT PRESENTED WITH A SHORT ANGLED AORTIC NECK THAT WAS FUNNEL IN SHAPE. IT WAS REPORTED THAT WHEN THE PROXIMAL FIXATION OPENED THE PHYSICIAN PUSHED UP ON THE DELIVERY SYSTEM TO BALLOON THE RING; HOWEVER, THE RINGS PUSHED UP, COVERED THE RIGHT RENAL ARTERY AND PARTIALLY COVERED THE LEFT. THE PHYSICIAN WAS ABLE TO STENT THE LEFT RENAL BUT THE GRAFT SLIPPED INTO THE RENAL ARTERY MORE THAN PLANNED. THE PHYSICIAN WAS UNABLE TO ADVANCE A SECOND RENAL STENT. THE PATIENT IS STABLE AND PRODUCING URINE WHEN HE LEFT THE OPERATING ROOM AND CURRENTLY BEING MONITORED WHILE ADDITIONAL INTERVENTION IS BEING DISCUSSED.
Patient problems: Failure of Implant
MAUDE report 12005738 ↗InjuryJun 3, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
THE PATIENT WAS INITIALLY BEING IMPLANTED WITH AN ALTO ABDOMINAL STENT GRAFT SYSTEM. DURING THE PROCEDURE, THE INTEGRATED BALLOON RUPTURED UPON INFLATION. THE ILIAC LIMBS WERE SUCCESSFULLY DEPLOYED USING THE UP AND OVER LUMEN AND, A PALMAZ P4010 (NON-ENDOLOGIX DEVICE) WAS PLACED AT THE LEVEL OF THE SEALING RING. A TYPE 1A ENDOLEAK WAS IDENTIFIED SO THE PHYSICIAN CHOSE TO STOP THE PROCEDURE. THE PHYSICIAN WILL BE GETTING A (CT) COMPUTED TOMOGRAPHY SCAN TO ASSESS THE LEAK LOCATION AND OPTIONS.
Patient problems: Failure of Implant
MAUDE report 11935306 ↗DeathFeb 17, 2021Improper or Incorrect Procedure or Method; Insufficient Device Problem Information; Improper or Incorrect Procedure or Method; Insufficient Device Problem Information
THE PATIENT WAS INITIALLY IMPLANTED WITH AN OVATION ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). AN INTRAOPERATIVE TYPE 1A ENDOLEAK WAS OBSERVED. THE PHYSICIAN ATTEMPTED BALLOONING AND IMPLANTED A NON-ENDOLOGIX (PALMAZ) BARE METAL STENT. THE NON-ENDOLOGIX (PALMAZ) BARE METAL STENT WAS MOUNTED ON A TYSHAK (NON - ENDOLOGIX) BALLOON. THE BALLOON WAS OVERINFLATED AND DISSECTED THE AORTA ABOVE THE RENAL ARTERIES. THE INTRAOPERATIVE TYPE 1A ENDOLEAK WAS RESOLVED. TWO (2) NON-ENDOLOGIX (COOK) ENDOGRAFTS WERE IMPLANTED TO CORRECT THE DISSECTION OF THE AORTA AND A RUPTURE THE RUPTURE APPEARED TO BE CONTAINED AND THE PATIENT¿S ACCESS SITES WERE CLOSED. THE PROCEDURE HAS CONCLUDED HOWEVER, IT WAS LATER REPORTED THAT THE PATIENT EXPIRED LATER THAT DAY. THE EXACT CAUSE OF DEATH WAS NOT PROVIDED.
Patient problems: Death; Rupture
MAUDE report 11340864 ↗DeathFeb 3, 2021Use of Device Problem; Patient-Device Incompatibility; Insufficient Device Problem Information; Use of Device Problem; Patient-Device Incompatibility; Insufficient Device Problem Information
DURING THE INITIAL IMPLANT OF THE OVATION ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA), THE PHYSICIAN WAS UNABLE TO CANNULATE THE CONTRALATERAL LIMB. AFTER SEVERAL HOURS OF TRYING MULTIPLE TECHNIQUES THE PHYSICIAN OPTED TO ABORT THE PROCEDURE. ONLY THE AORTIC MAIN BODY AND ONE (1) LIMB WAS IMPLANTED. AN INTERVENTION IS BEING PLANNED AND THE PATIENT IS BEING MONITORED.
Patient problems: Failure of Implant
MAUDE report 11273086 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.
- Z-2263-2020 · May 6, 2020 · TerminatedOvation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Z-0125-2015 · Sep 20, 2014 · TerminatedOvation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
- Z-0288-2013 · Oct 19, 2012 · TerminatedTriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page