Alto™ Abdominal Stent Graft System TV-AB2980-N (UDI M701TVAB2980N1)Recalls & reported problems

Endologix LLC · The Alto™ Abdominal Stent Graft System is an endovascular device delivered via a low-profile catheter to treat abdominal aortic aneurysms (AAAs). The stent graft is designed to reline the diseased vasculature, providing an endovascular blood conduit for isolating the aneurysm from the high-pressure flow of blood, thereby reducing the risk of rupture. The stent graft is a modular configuration comprised of an aortic body section, iliac limbs, and iliac extensions as required. The Alto™ Abdominal Stent Graft System includes: an Aortic Body Stent Graft and delivery catheter, Ovation iX™ Iliac Limb Stent Grafts and delivery catheters, Ovation iX™ Iliac Extension Stent Grafts and delivery catheters (as required), the CustomSeal™ Polymer Fill Kit, and the Autoinjector 2. · Class 3 · Implant

All models of Alto Abdominal Stent Graft System (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
2
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI M701TVAB2980N1), received between Feb 3, 2021 and Nov 10, 2022 — classified as 2 deaths, 5 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).

3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Insufficient Device Problem Information4
  • Patient-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Improper or Incorrect Procedure or Method1
  • Malposition of Device1
  • Material Puncture/Hole1
  • Positioning Failure1
  • Use of Device Problem1

Patient problems

  • Failure of Implant6
  • Aneurysm1
  • Clinical Endoleak1
  • Death1
  • Rupture1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 10, 2022Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem; Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY ONE AND A HALF (1.5) YEARS POST INITIAL PROCEDURE A POSSIBLE TYPE 3B ENDOLEAK AND SAC GROWTH WERE NOTED. THE PATIENT IS CURRENTLY BEING MONITORED WHILE AN INTERVENTION IS BEING DISCUSSED.

    Patient problems: Aneurysm; Clinical Endoleak; Failure of Implant

    MAUDE report 15775512 ↗
  • InjuryAug 10, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). DURING THIS INITIAL PROCEDURE, THERE WAS A TYPE 1A ENDOLEAK. THE PHYSICIAN ELECTED TO IMPLANT A NON ENDOLOGIX (PALMAZ) STENT. THE ENDOLEAK IMPROVED BUT IT DID NOT COMPLETELY RESOLVE. THE PATIENT IS CURRENTLY BEING MONITORED AND THE ENDOLEAK WILL BE RE-EVALUATED AT THE 30 DAY FOLLOW UP VISIT. SUBSEQUENT TO THE INITIAL REPORT, ADDITIONAL INFORMATION WAS PROVIDED STATING THAT THE TYPE IA ENDOLEAK HAS RESOLVED ON ITS OWN AND REINTERVENTION WAS NOT REQUIRED.

    Patient problems: Failure of Implant

    MAUDE report 12305342 ↗
  • InjuryJul 2, 2021Positioning Failure; Positioning Failure

    THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). IT WAS REPORTED THAT WHEN THE POLYMER WAS DELIVERED, THE CONTRALATERAL LIMB DID NOT DEPLOY. ALL TROUBLESHOOTING STEPS WERE PERFORMED; HOWEVER, THE CONTRALATERAL LIMB DID NOT DEPLOY SO THE POLYMER WAS NOT ABLE TO FILL THE LIMB. DEPLOYMENT OF THE IPSILATERAL LIMB WAS COMPLETED AND THE ANEURYSM WAS SEALED WITHOUT AN ENDOLEAK. THE PHYSICIAN THEN PERFORMED A FEMORAL-FEMORAL BYPASS TO PROVIDE BLOOD FLOW TO BOTH LOWER EXTREMITIES. THE PATIENT WAS REPORTED AS DOING WELL AND BOTH LOWER EXTREMITIES HAVE PALPABLE PULSES.

    Patient problems: Failure of Implant

    MAUDE report 12110984 ↗
  • InjuryJun 15, 2021Malposition of Device; Patient-Device Incompatibility; Malposition of Device; Patient-Device Incompatibility

    THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). IT WAS NOTED THAT THE PATIENT PRESENTED WITH A SHORT ANGLED AORTIC NECK THAT WAS FUNNEL IN SHAPE. IT WAS REPORTED THAT WHEN THE PROXIMAL FIXATION OPENED THE PHYSICIAN PUSHED UP ON THE DELIVERY SYSTEM TO BALLOON THE RING; HOWEVER, THE RINGS PUSHED UP, COVERED THE RIGHT RENAL ARTERY AND PARTIALLY COVERED THE LEFT. THE PHYSICIAN WAS ABLE TO STENT THE LEFT RENAL BUT THE GRAFT SLIPPED INTO THE RENAL ARTERY MORE THAN PLANNED. THE PHYSICIAN WAS UNABLE TO ADVANCE A SECOND RENAL STENT. THE PATIENT IS STABLE AND PRODUCING URINE WHEN HE LEFT THE OPERATING ROOM AND CURRENTLY BEING MONITORED WHILE ADDITIONAL INTERVENTION IS BEING DISCUSSED.

    Patient problems: Failure of Implant

    MAUDE report 12005738 ↗
  • InjuryJun 3, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT WAS INITIALLY BEING IMPLANTED WITH AN ALTO ABDOMINAL STENT GRAFT SYSTEM. DURING THE PROCEDURE, THE INTEGRATED BALLOON RUPTURED UPON INFLATION. THE ILIAC LIMBS WERE SUCCESSFULLY DEPLOYED USING THE UP AND OVER LUMEN AND, A PALMAZ P4010 (NON-ENDOLOGIX DEVICE) WAS PLACED AT THE LEVEL OF THE SEALING RING. A TYPE 1A ENDOLEAK WAS IDENTIFIED SO THE PHYSICIAN CHOSE TO STOP THE PROCEDURE. THE PHYSICIAN WILL BE GETTING A (CT) COMPUTED TOMOGRAPHY SCAN TO ASSESS THE LEAK LOCATION AND OPTIONS.

    Patient problems: Failure of Implant

    MAUDE report 11935306 ↗
  • DeathFeb 17, 2021Improper or Incorrect Procedure or Method; Insufficient Device Problem Information; Improper or Incorrect Procedure or Method; Insufficient Device Problem Information

    THE PATIENT WAS INITIALLY IMPLANTED WITH AN OVATION ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). AN INTRAOPERATIVE TYPE 1A ENDOLEAK WAS OBSERVED. THE PHYSICIAN ATTEMPTED BALLOONING AND IMPLANTED A NON-ENDOLOGIX (PALMAZ) BARE METAL STENT. THE NON-ENDOLOGIX (PALMAZ) BARE METAL STENT WAS MOUNTED ON A TYSHAK (NON - ENDOLOGIX) BALLOON. THE BALLOON WAS OVERINFLATED AND DISSECTED THE AORTA ABOVE THE RENAL ARTERIES. THE INTRAOPERATIVE TYPE 1A ENDOLEAK WAS RESOLVED. TWO (2) NON-ENDOLOGIX (COOK) ENDOGRAFTS WERE IMPLANTED TO CORRECT THE DISSECTION OF THE AORTA AND A RUPTURE THE RUPTURE APPEARED TO BE CONTAINED AND THE PATIENT¿S ACCESS SITES WERE CLOSED. THE PROCEDURE HAS CONCLUDED HOWEVER, IT WAS LATER REPORTED THAT THE PATIENT EXPIRED LATER THAT DAY. THE EXACT CAUSE OF DEATH WAS NOT PROVIDED.

    Patient problems: Death; Rupture

    MAUDE report 11340864 ↗
  • DeathFeb 3, 2021Use of Device Problem; Patient-Device Incompatibility; Insufficient Device Problem Information; Use of Device Problem; Patient-Device Incompatibility; Insufficient Device Problem Information

    DURING THE INITIAL IMPLANT OF THE OVATION ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA), THE PHYSICIAN WAS UNABLE TO CANNULATE THE CONTRALATERAL LIMB. AFTER SEVERAL HOURS OF TRYING MULTIPLE TECHNIQUES THE PHYSICIAN OPTED TO ABORT THE PROCEDURE. ONLY THE AORTIC MAIN BODY AND ONE (1) LIMB WAS IMPLANTED. AN INTERVENTION IS BEING PLANNED AND THE PATIENT IS BEING MONITORED.

    Patient problems: Failure of Implant

    MAUDE report 11273086 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.

  • Z-2263-2020 · May 6, 2020 · Terminated
    Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
  • Z-0125-2015 · Sep 20, 2014 · Terminated
    Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
  • Z-0288-2013 · Oct 19, 2012 · Terminated
    TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The