Skip to content

Ovation iX Iliac Stent Graft Iliac Extension (REF TV-EX181845-J)Recalls & reported problems

Endologix LLC · The TriVascular® Ovation Prime® Abdominal Stent Graft System with the Ovation iX Iliac Stent Graft is an endovascular device delivered via a small diameter catheter to treat abdominal aortic aneurysms (AAAs). The stent graft is designed to reline the diseased vasculature, providing an endovascular blood conduit for isolating the aneurysm from the high pressure flow of blood, thereby reducing the risk of rupture. The stent graft is a modular configuration comprised of an aortic body section, iliac limbs, and iliac extensions as required. The TriVascular Ovation Prime Abdominal Stent Graft System with the Ovation iX Iliac Stent Graft includes, an Aortic Body Stent Graft and delivery catheter, Iliac Limb Stent Grafts and delivery catheters, Iliac Extension Stent Grafts and delivery catheters, as required, a Fill Polymer Kit, and an Autoinjector. · Class 3 · Implant

All models of Ovation iX Iliac Stent Graft (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M701TVEX181845J1), received between Mar 22, 2024 and Mar 22, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Lack of Effect” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Lack of Effect1

Patient problems

  • Clinical Endoleak1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 22, 2024Lack of Effect; Lack of Effect

    THE PATIENT WAS INITIALLY IMPLANTED WITH AN AFX BIFURCATED STENT GRAFT TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA) ON (B)(6) 2015. ADDITIONALLY, (3) THREE OVATION IX ILIAC EXTENDERS WERE IMPLANTED ON (B)(6) 2016 FOR AN UNKNOWN REASON. THIS PROCEDURE IS OUTSIDE THE INDICATIONS OF USE (OFF-LABEL) DUE TO THE USE OF ADJUNCTIVE DEVICES NOT COMPATIBLE WITH AFX SYSTEM PER THE IFU. APPROXIMATELY NINE (9) YEARS POST INITIAL PROCEDURE, DURING A ROUTINE FOLLOW UP, THE LEFT LEG OF THE BIFURCATED STENT GRAFT WAS FOUND TO HAVE SEPARATED FROM ONE OF THE OVATION IX ILIAC EXTENDERS (CLASSIFIED AS A TYPE 3A ENDOLEAK). THE PHYSICIAN ELECTED TO RELINE THE PREVIOUSLY IMPLANTED STENT GRAFTS WITH AN ADDITIONAL AFX2 BIFURCATED STENT GRAFT, AN OVATION IX ILIAC LIMB AND AN OVATION IX ILIAC EXTENDER ON (B)(6)2024 TO SUCCESSFULLY RESOLVE THIS EVENT. THE PATIENT WAS REPORTED AS STABLE POST-SECONDARY PROCEDURE. THIS EVENT WAS PREVIOUSLY REPORTED UNDER MDR#: 3011063223-2024-00008. NOW APPROXIMATELY SEVEN (7) YEARS POST IMPLANT OF THE OVATION IX EXTENDERS, AN INTERNAL CLINICAL ASSESSMENT DETERMINED THAT THERE WAS EVIDENCE TO REASONABLY SUGGEST A TYPE 1B ENDOLEAK OF THE RIGHT COMMON ILIAC ARTERY OCCURRED. THE TYPE

    Patient problems: Clinical Endoleak

    MAUDE report 18958025 ↗

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page