Ovation iX Iliac Stent Graft Iliac Limb (REF TV-IL1414140-J)Recalls & reported problems
Endologix LLC · The TriVascular® Ovation Prime® Abdominal Stent Graft System with the Ovation iX Iliac Stent Graft is an endovascular device delivered via a small diameter catheter to treat abdominal aortic aneurysms (AAAs). The stent graft is designed to reline the diseased vasculature, providing an endovascular blood conduit for isolating the aneurysm from the high pressure flow of blood, thereby reducing the risk of rupture. The stent graft is a modular configuration comprised of an aortic body section, iliac limbs, and iliac extensions as required. The TriVascular Ovation Prime Abdominal Stent Graft System with the Ovation iX Iliac Stent Graft includes, an Aortic Body Stent Graft and delivery catheter, Iliac Limb Stent Grafts and delivery catheters, Iliac Extension Stent Grafts and delivery catheters, as required, a Fill Polymer Kit, and an Autoinjector. · Class 3 · Implant
All models of Ovation iX Iliac Stent Graft (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI M701TVIL1414140J1), received between Jul 9, 2016 and Feb 13, 2020 — classified as 8 injuries.
The most frequently reported specific device problem is “Occlusion Within Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Occlusion Within Device3
- Leak/Splash2
- Use of Device Problem2
- Failure To Adhere Or Bond1
- Insufficient Device Problem Information1
- Malposition of Device1
- Obstruction of Flow1
- Partial Blockage1
Patient problems
- Failure of Implant4
- Occlusion4
- Blood Loss1
- Pain1
- Perforation of Vessels1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryFeb 13, 2020Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE OVATION IX ABDOMINAL AORTIC ANEURYSM STENT. DURING THE INITIAL PROCEDURE, THE PHYSICIAN ALSO REPAIRED THE PATIENT'S RIGHT ILIAC ARTERY (DISTAL TO THE OVATION LIMB), AND A PERFORATION WAS SEEN ON ANGIOGRAM. THEREFORE, THE PHYSICIAN IMPLANTED TWO (2) VBX (NON-ENDOLOGIX) STENTS; THIS IS AN OFF-LABEL PROCEDURE DUE TO NONCOMPATIBLE USE OF OVATION WITH A NON-ENDOLOGIX DEVICE. THIS REPORT IS TO ADDRESS THE INTRAOPERATIVE PERFORATION EVENT. A FEW HOURS POST INITIAL PROCEDURE, THE PATIENT WAS NOTED TO STILL BE BLEEDING. THE PHYSICIAN ELECTED TO TREAT THE PATIENT BY IMPLANTING AN ADDITIONAL OVATION IX ILIAC LIMB DEVICE. THE PATIENT REPORTEDLY TOLERATED THE PROCEDURE WELL. AS OF DATE, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED. REFERENCE MFR. REPORT # 3008011247-2020-00023 FOR POSTOPERATIVE BLEEDING EVENT.
Patient problems: Perforation of Vessels
MAUDE report 9709176 ↗InjuryFeb 13, 2020Use of Device Problem; Use of Device Problem
THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE OVATION IX ABDOMINAL AORTIC ANEURYSM STENT. DURING THE INITIAL PROCEDURE, THE PHYSICIAN ALSO REPAIRED THE PATIENT'S RIGHT ILIAC ARTERY (DISTAL TO THE OVATION LIMB), AND A PERFORATION WAS SEEN ON ANGIOGRAM. THEREFORE, THE PHYSICIAN IMPLANTED TWO (2) VBX (NON-ENDOLOGIX) STENTS; THIS IS AN OFF-LABEL PROCEDURE DUE TO NONCOMPATIBLE USE OF OVATION WITH A NON-ENDOLOGIX DEVICE. A FEW HOURS POST INITIAL PROCEDURE, THE PATIENT WAS NOTED TO STILL BE BLEEDING. THE PHYSICIAN ELECTED TO TREAT THE PATIENT BY IMPLANTING AN ADDITIONAL OVATION IX ILIAC LIMB DEVICE. THE PATIENT REPORTEDLY TOLERATED THE PROCEDURE WELL. AS OF DATE, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED.
Patient problems: Blood Loss; Failure of Implant
MAUDE report 9709173 ↗InjuryAug 21, 2018Obstruction of Flow; Obstruction of Flow
AN OVATION IX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM IN 2017. RECENTLY, ON AN UNKNOWN DATE, IT WAS REPORTED THAT THE PATIENT PRESENTED AT THE HOSPITAL WITH OCCLUDED ENDOGRAFT WITH BILATERAL LIMB ISCHEMIA. RE-INTERVENTION WAS PERFORMED ON (B)(6) 2018 AT A HOSPITAL NOT AFFILIATED WITH ENDOLOGIX; THEREFORE, NO INFORMATION COULD BE OBTAINED REGARDING THE SECONDARY PROCEDURE.
Patient problems: Failure of Implant; Occlusion
MAUDE report 7801100 ↗InjuryMay 30, 2018Failure To Adhere Or Bond; Leak/Splash; Failure To Adhere Or Bond; Leak/Splash
AN OVATION IX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE PHYSICIAN REPORTED THAT AT THE ROUTINE 2 MONTH FOLLOW-UP, THE PATIENT PRESENTED WITH A POTENTIAL TYPE IB ENDOLEAK. A RE-INTERVENTION WAS PERFORMED WHERE AN ADDITIONAL EXTENSION WAS PLACED TO EXTEND THE DISTAL SEAL. AS OF THE DATE OF THIS REPORT, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED AND THE PATIENT WILL CONTINUE TO BE MONITORED.
Patient problems: Failure of Implant
MAUDE report 7553682 ↗InjuryMay 17, 2018Partial Blockage; Occlusion Within Device; Partial Blockage; Occlusion Within Device
THE PATIENT WAS INITIALLY TREATED WITH OVATION IX. ENDOLOGIX WAS RECENTLY MADE AWARE FROM THE PHYSICIAN THAT THE PATIENT UNDERWENT A SECONDARY PROCEDURE FOR AN OCCLUDED LIMB. THE PATIENT WAS REPORTEDLY LOST THE PATIENT TO FOLLOW-UP HOWEVER, WAS CALLED TO THE ER FOR A LEFT LIMB OCCLUSION WHICH MAY HAVE BEEN OCCLUDED FOR AN EXTENDED LENGTH OF TIME BECAUSE THE PATIENT REVEALED THAT THERE HAD BEEN PAIN IN THE LEFT LEG SOON AFTER THE INITIAL PROCEDURE. THE RADIOLOGIST CONFIRMED SEEING THE PATIENT ON (B)(6) 2017; APPROXIMATELY 14 MONTHS POST-OP. AT THE RECENT ER VISIT, THE IMPLANTING PHYSICIAN ELECTED TO TREAT THE PATIENT BY FEM/FEM AND RELINE FLOW TO THE PATIENT'S LEFT LEG. THIS SECONDARY WAS REPORTED TO HAVE OCCURRED ON (B)(6) 2018. THE CURRENT PATIENT CONDITION IS REPORTED AS FINE WITH NO PAIN IN EITHER LEG, NO EVIDENCE OF ANY ENDOLEAK. THE PHYSICIAN HAS SHARED THAT THE PATIENT WILL BE MONITORED UNDER HIS CARE FOR FUTURE FOLLOW-UPS.
Patient problems: Occlusion; Pain
MAUDE report 7523976 ↗InjuryFeb 13, 2018Leak/Splash; Malposition of Device; Leak/Splash; Malposition of Device
AN OVATION IX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE PATIENT WAS ON-LABEL FOR THE OVATION IX SYSTEM. THE STENT MID-CROWN WAS DEPLOYED AND PARALLAX WAS CORRECTED; A NEW ANGIO WAS DONE TO RETRACE THE INFRARENAL ARTERY (IR) ON THE SCREEN. THE PATIENT HAD A RAPID CONICAL NECK AND NEEDED A PRECISE PLACEMENT TO OBTAIN A SEAL. IT WAS NOTED THAT THE PATIENT WAS NOT ASLEEP AND COOPERATION FROM THE PATIENT FOR A BREATH HOLD WAS REQUESTED BY THE VASCULAR SURGEON. THE PATIENT DID NOT COMPLETELY STOP BREATHING AND THERE WAS MOVEMENT EVIDENT ON THE IMAGING. THE INTERVENTIONAL RADIOLOGIST REQUESTED A SECOND ANGIO FOR MORE PRECISION AND THE VASCULAR SURGEON DECLINED AND SAID THAT THE IMAGES OBTAINED WOULD SUFFICE. THE PIG TAIL WAS REMOVED AND THE PROXIMAL CROWN WAS DEPLOYED; THE POLYMER WAS MIXED AND INJECTED. THE CONTRA GATE WAS CANNULATED AND CONTRA LIMB WAS DEPLOYED. AT 20 MINUTES POST POLYMER INJECTION A CONTROL ANGIO WAS SUGGESTED BY THE REP TO CONFIRM SEAL AND DECIDE ON THE NEED FOR A LARGE DIAMETER RING BALLOONING (SINCE THERE WAS PRESENCE OF THROMBUS IN THE SEAL ZONE); THE CONTROL ANGIO SHOWED THAT THE GRAFT WAS APPROXIMATELY 10 MM BELOW THE F
Patient problems: Failure of Implant
MAUDE report 7266526 ↗InjuryAug 4, 2016Occlusion Within Device; Occlusion Within Device
AN OVATION IX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE 5 MONTH POST-OPERATIVE ANGIOGRAPHIC IMAGING SHOWED THE PRESENCE OF AN OCCLUSION OF THE ILIAC LIMB STENT GRAFT. A THROMBECTOMY RE-INTERVENTION WAS PERFORMED FOLLOWED BY THE PLACEMENT OF A BALLOON EXPANDABLE STENT WITHIN THE ILIAC LIMB STENT GRAFT TO MAINTAIN PATENCY. AS OF THE DATE OF THIS REPORT, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED.
Patient problems: Occlusion
MAUDE report 5848945 ↗InjuryJul 9, 2016Occlusion Within Device; Occlusion Within Device
AN OVATION IX ABDOMINAL STENT GRAFT WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. IT WAS NOTED THAT THE NATIVE AORTIC BIFURCATION WAS NARROW; HOWEVER, THE BILATERAL ILIAC LIMB STENT GRAFTS WERE PATENT AT THE COMPLETION OF THE INDEX PROCEDURE. LESS THAN 24 HOURS POST-OP, THE PATIENT EXPERIENCED A "COLD LEFT LEG" AND A POTENTIAL OCCLUSION OF THE LEFT ILIAC LIMB STENT GRAFT. A RE-INTERVENTION WAS PERFORMED SUCCESSFULLY RESOLVING THE OCCLUSION. AS OF THE DATE OF THIS REPORT, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED.
Patient problems: Occlusion
MAUDE report 5780929 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page