Ovation iX Iliac Stent Graft Iliac Limb (REF TV-IL1416100-J)Recalls & reported problems
Endologix LLC · The TriVascular® Ovation Prime® Abdominal Stent Graft System with the Ovation iX Iliac Stent Graft is an endovascular device delivered via a small diameter catheter to treat abdominal aortic aneurysms (AAAs). The stent graft is designed to reline the diseased vasculature, providing an endovascular blood conduit for isolating the aneurysm from the high pressure flow of blood, thereby reducing the risk of rupture. The stent graft is a modular configuration comprised of an aortic body section, iliac limbs, and iliac extensions as required. The TriVascular Ovation Prime Abdominal Stent Graft System with the Ovation iX Iliac Stent Graft includes, an Aortic Body Stent Graft and delivery catheter, Iliac Limb Stent Grafts and delivery catheters, Iliac Extension Stent Grafts and delivery catheters, as required, a Fill Polymer Kit, and an Autoinjector. · Class 3 · Implant
All models of Ovation iX Iliac Stent Graft (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M701TVIL1416100J1), received between Oct 22, 2016 and Sep 14, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Occlusion Within Device2
- Adverse Event Without Identified Device or Use Problem1
- Collapse1
- Failure To Adhere Or Bond1
- Kinked1
- Lack of Effect1
Patient problems
- Failure of Implant2
- Occlusion2
- Aneurysm1
- Claudication1
- Clinical Endoleak1
- Death1
- Pain1
- Stenosis1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 14, 2023Adverse Event Without Identified Device or Use Problem; Lack of Effect; Adverse Event Without Identified Device or Use Problem; Lack of Effect
THE PATIENT WAS INITIALLY IMPLANTED WITH THE OVATION IX STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY ONE (1) YEAR POST INITIAL PROCEDURE, DURING A ROUTINE FOLLOW UP, A JUNCTION ENDOLEAK (CATEGORIZED AS A TYPE 3A ENDOLEAK) WAS IDENTIFIED. THE PHYSICIAN ELECTED TO IMPLANT AN OVATION IX ILIAC LIMB STENT GRAFT TO SUCCESSFULLY RESOLVE THIS EVENT. THE PATIENT STATUS WAS REPORTED AS STABLE. THIS EVENT WAS PREVIOUSLY REPORTED UNDER MDR # 3008011247-2021-00140. ON 28 AUGUST 2023, DURING ROUTINE FOLLOW UP, ANOTHER JUNCTIONAL ENDOLEAK WAS SUSPECTED AT THE PROXIMAL PART OF THE LIMB. THE PATIENT WAS BROUGHT BACK ON (B)(6) 2023 AND AN ANGIOGRAM WAS PERFORMED. AN ENDOLEAK WAS DIFFICULT TO VISUALIZE; HOWEVER, THE PHYSICIAN DETERMINED THIS TO BE A TYPE 3A ENDOLEAK AND ELECTED TO TREAT THE PATIENT BY PLACING 14X45 OVATION LIMBS TO EACH OF THE GATES RAISING THE BIFURCATION. THE PATIENT WAS REPORTED AS DOING FINE POST THIS TERTIARY PROCEDURE.
Patient problems: Clinical Endoleak
MAUDE report 17749389 ↗InjuryApr 4, 2018Failure To Adhere Or Bond; Leak/Splash; Failure To Adhere Or Bond; Leak/Splash
AN OVATION IX ILIAC LIMB STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE PATIENT RETURNED FOR AN UNKNOWN REASON APPROXIMATELY 2 YEARS POST INDEX PROCEDURE WITH A TYPE IB ENDOELAK OF THE RIGHT ILIAC LIMB. THE PHYSICIAN IS PLANNING TO RE-INTERVENE BY EXTENDING THE RIGHT ILIAC LIMB, HOWEVER, A DATE FOR THE RE-INTERVENTION HAS NOT BEEN SCHEDULED. AS OF THE DATE OF THIS REPORT, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED.
Patient problems: Failure of Implant
MAUDE report 7396061 ↗InjuryJan 17, 2018Collapse; Kinked; Leak/Splash; Occlusion Within Device; Collapse; Kinked; Leak/Splash; Occlusion Within Device
AN OVATION IX ILIAC LIMB STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. APPROXIMATELY 3 WEEKS POST INDEX PROCEDURE, THE PATIENT PRESENTED WITH CLAUDICATION AND A COLD LEFT FOOT. A CT WAS DONE WHICH SHOWED ONE LIMB BLOCKING THE FLOW OF THE OTHER LIMB LEADING TO AN OCCLUSION. A RE-INTERVENTION WAS PERFORMED WHERE THE PHYSICIAN ATTEMPTED TO BALLOON THE OCCLUDED LIMB, HOWEVER THE PATIENT COULD NOT TOLERATE THE PROCEDURE AND EXPIRED.
Patient problems: Aneurysm; Claudication; Death; Failure of Implant; Occlusion; Pain; Stenosis
MAUDE report 7199542 ↗InjuryOct 22, 2016Occlusion Within Device; Occlusion Within Device
AN OVATION IX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE 2 MONTH POST-OPERATIVE CT SHOWED THE PRESENCE OF A COMPLETE THROMBOSIS OF THE LEFT ILIAC LIMB STENT GRAFT. THE PATIENT WAS STARTED ON DRUG THERAPY, AND A RE-INTERVENTION WAS PERFORMED TO PLACE AN ADDITIONAL ILIAC LIMB STENT GRAFT TO RE-LINE THE PREVIOUSLY PLACED ILIAC LIMB, AND FLOW WAS SUCCESSFULLY RESTORED. AS OF THE DATE OF THIS REPORT, THERE HAVE BEEN NO ADDITIONAL PATIENT SEQUELAE REPORTED.
Patient problems: Occlusion
MAUDE report 6049532 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page