OptiMesh Expandable Interbody Fusion System 10-10-100-2Recalls & reported problems
Spineology Inc. · OptiMesh · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M7403003032), received between Jul 26, 2024 and Jul 26, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 26, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE PATIENT UNDERWENT AN INTERBODY FUSION PROCEDURE WITH PLACEMENT OF AN INTERBODY FUSION DEVICE ON (B)(6) 2024. DURING THE SURGICAL PROCEDURE, THE DISTAL TIP OF A BONE FILL TUBE BENT WHEN FILLING THE IMPLANT AND COULD NOT BE EASILY REMOVED FROM THE INSERTER. UPON FORCEFUL REMOVAL, THE INSERTER ASSEMBLY WITHDREW FROM THE ACCESS TUBE SHEARING OFF THE BENT DISTAL TIP OF THE BONE FILL TUBE AND UNINTENTIONALLY RELEASING THE MESH IMPLANT. A SEPARATE INCISION AND SURGICAL APPROACH WERE NECESSARY TO REMOVE THE SHEARED TIP AND REPLACE THE PARTIALLY FILLED IMPLANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19839210 ↗
Other Intervertebral Body Graft Containment Device devices
- OptiMesh Expandable Interbody Fusion System 10-10-100-3Spineology Inc. · Intervertebral Body Graft Containment Device2 reports
- OptiMesh Expandable Interbody Fusion System 10-10-100-4Spineology Inc. · Intervertebral Body Graft Containment Device2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page