OptiMesh 1500 E 10-10-04-15Recalls & reported problems
Spineology Inc. · Surgical Mesh · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M7403302705), received between Jul 24, 2018 and Oct 28, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Migration” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Migration1
Patient problems
- Numbness1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 28, 2018Appropriate Device Problem Term/Code Not Available
PATIENT AWOKE WITH NEW LEG PAIN FOLLOWING LUMBAR FUSION SURGERY. A FOLLOW-UP CT IDENTIFIED BONE GRAFT IN THE CENTRAL CANAL AT L3-L4. A SECOND SURGERY WAS PERFORMED TWO DAYS LATER. A LAMINECTOMY WAS PERFORMED AT L2-L3 AND BONE GRAFT WAS REMOVED FROM THE CANAL AT L2-L3 AND AT L3-L4, AND INTRAOPERATIVE NEUROMONITORING SHOWED IMMEDIATE IMPROVEMENT. THE IMPLANT AND A LARGE PIECE OF DISC MATERIAL WERE BOTH REMOVED FROM THE L3-L4 DISC SPACE AND REPLACED WITH AN EXPANDABLE FUSION CAGE AND ALLOGRAFT. THE SYMPTOMS ARE REPORTED TO BE RESOLVED.
Patient problems: Pain
MAUDE report 8012455 ↗InjuryJul 24, 2018Migration; Migration
PATIENT UNDERWENT AN INSTRUMENTED 3-LEVEL LUMBAR INTERBODY FUSION PROCEDURE AT L2-3, L3-4 AND L4-5. PATIENT AWOKE POST-OP WITH BILATERAL NUMBNESS IN LOWER EXTREMITIES. IMMEDIATE CT SHOWED MIGRATION OF GRAFT/IMPLANT AT L2-3 AND L4-5, WITH ALLOGRAFT BONE IN CENTRAL CANAL. IT WAS NOTED THAT ROD REDUCTION CAUSED STRAIGHTENING OF THE CURVED RODS, INDICATING SIGNIFICANT MECHANICAL FORCE HAD BEEN APPLIED. PATIENT WAS RETURNED TO THE OR FOR DECOMPRESSION OF THE CENTRAL CANAL AND REMOVAL OF INTERBODY IMPLANTS AT L2-3 AND L4-5. ALLOGRAFT BONE WAS PLACED IN THE DISC SPACES AT THOSE LEVELS AND ADDITIONAL DORSAL GRAFTING WAS PERFORMED. PATIENT WAS DISCHARGED ON POD #5 WITH RESIDUAL NUMBNESS IN BOTH LOWER EXTREMITIES AND IS RECEIVING PHYSICAL THERAPY. NUMBNESS IS REPORTED TO BE IMPROVING.
Patient problems: Numbness
MAUDE report 7716285 ↗
Other Mesh, surgical, metal devices
- OptiMesh 1500 E 10-10-04-6Spineology Inc. · Mesh, surgical, metal4 reports
- OptiMesh 1500 E 10-10-04-8Spineology Inc. · Mesh, surgical, metal1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page