Fortress Pedicle Screw 10-10-48-16Recalls & reported problems
Spineology Inc. · Pedicle Screw · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M7406406540), received between Sep 8, 2016 and Sep 8, 2016 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
Patient problems
- Fall3
- Pain3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 8, 2016Fracture; Fracture
THE PATIENT UNDERWENT A 2-LEVEL SPINAL SURGICAL PROCEDURE ON (B)(6) 2014 WITH PLACEMENT OF SIX (6) FORTRESS PEDICLE SCREWS, SIX (6) FORTRESS SET SCREWS, AND TWO (2) FORTRESS RODS. AN UNSPECIFIED ANTERIOR INTERBODY FUSION WAS ALSO PLACED. AT AN UNSPECIFIED TIME FOLLOWING THE INITIAL IMPLANT PROCEDURE, THE PATIENT SUFFERED A FALL FOR UNKNOWN REASONS AND REPORTED "POPPING AND PAIN". CT IMAGING REVEALED THAT FOUR OF THE SIX PEDICLE SCREWS SUSTAINED SCREW SHANK FRACTURES. THE IMAGING ALSO REVEALED THAT THE INTERBODY DEVICE APPEARED TO BE "VERY WELL FUSED". ON (B)(6) 2016, THE PATIENT RETURNED TO SURGERY AT WHICH TIME ALL FORTRESS DEVICES WERE EXPLANTED. THE DISTAL SHANK ENDS OF THE FOUR BROKEN PEDICLE SCREWS WERE UNABLE TO BE RETRIEVED. SPINEOLOGY MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT; HOWEVER, ALL ATTEMPTS WERE UNSUCCESSFUL.
Patient problems: Fall; Pain
MAUDE report 5936490 ↗InjurySep 8, 2016Fracture; Fracture
THE PATIENT UNDERWENT A 2-LEVEL SPINAL SURGICAL PROCEDURE ON (B)(6) 2014 WITH PLACEMENT OF SIX (6) FORTRESS PEDICLE SCREWS, SIX (6) FORTRESS SET SCREWS, AND TWO (2) FORTRESS RODS. AN UNSPECIFIED ANTERIOR INTERBODY FUSION WAS ALSO PLACED. AT AN UNSPECIFIED TIME FOLLOWING THE INITIAL IMPLANT PROCEDURE, THE PATIENT SUFFERED A FALL FOR UNKNOWN REASONS AND REPORTED "POPPING AND PAIN". CT IMAGING REVEALED THAT FOUR OF THE SIX PEDICLE SCREWS SUSTAINED SCREW SHANK FRACTURES. THE IMAGING ALSO REVEALED THAT THE INTERBODY DEVICE APPEARED TO BE "VERY WELL FUSED". ON (B)(6) 2016, THE PATIENT RETURNED TO SURGERY AT WHICH TIME ALL FORTRESS DEVICES WERE EXPLANTED. THE DISTAL SHANK ENDS OF THE FOUR BROKEN PEDICLE SCREWS WERE UNABLE TO BE RETRIEVED. SPINEOLOGY MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT; HOWEVER, ALL ATTEMPTS WERE UNSUCCESSFUL.
Patient problems: Fall; Pain
MAUDE report 5936488 ↗InjurySep 8, 2016Fracture; Fracture
THE PATIENT UNDERWENT A 2-LEVEL SPINAL SURGICAL PROCEDURE ON (B)(6) 2014 WITH PLACEMENT OF SIX (6) FORTRESS PEDICLE SCREWS, SIX (6) FORTRESS SET SCREWS, AND TWO (2) FORTRESS RODS. AN UNSPECIFIED ANTERIOR INTERBODY FUSION WAS ALSO PLACED. AT AN UNSPECIFIED TIME FOLLOWING THE INITIAL IMPLANT PROCEDURE, THE PATIENT SUFFERED A FALL FOR UNKNOWN REASONS AND REPORTED "POPPING AND PAIN". CT IMAGING REVEALED THAT FOUR OF THE SIX PEDICLE SCREWS SUSTAINED SCREW SHANK FRACTURES. THE IMAGING ALSO REVEALED THAT THE INTERBODY DEVICE APPEARED TO BE "VERY WELL FUSED". ON (B)(6) 2016, THE PATIENT RETURNED TO SURGERY AT WHICH TIME ALL FORTRESS DEVICES WERE EXPLANTED. THE DISTAL SHANK ENDS OF THE FOUR BROKEN PEDICLE SCREWS WERE UNABLE TO BE RETRIEVED. SPINEOLOGY MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT; HOWEVER, ALL ATTEMPTS WERE UNSUCCESSFUL.
Patient problems: Fall; Pain
MAUDE report 5936486 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page