Skip to content

i-FACTOR™ Peptide Enhanced Bone Graft 700-050 (UDI M8517000501)Recalls & reported problems

Cerapedics Inc. · i-FACTOR™ Putty, 5.0 cc Syringe · Class 3 · Implant

All models of i-FACTOR Peptide Enhanced Bone Graft (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M8517000501), received between Nov 14, 2019 and Nov 14, 2019 — classified as 1 injury.

1 other recall cover devices cleared under the same FDA submission (P140019); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Nerve Damage1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2019Appropriate Device Problem Term/Code Not Available

    INTERBODY FUSION WAS PERFORMED USING I-FACTOR + ALLOGRAFT. IT WAS PACKED INSIDE THE DISC SPACE FROM RIGHT SIDE WITHOUT A CAGE AS DISC SPACE WAS NARROW AND THERE WAS OSTEOPOROSIS. 30 HOURS AFTER SURGERY, THE PATIENT WALKED IN THE HALLWAY AND SHE DEVELOPED PROGRESSIVELY WORSENING FOOT DROP. CT AND MRI SHOWED THIS WAS SECONDARY TO I-FACTOR AS IT RETROPULSED INTO THE SPINAL CANAL ON THE LEFT SIDE (OPPOSITE THE SIDE IT WAS INTRODUCED). THE ALLOGRAFT WAS AN UNKNOWN BRAND OF CANCELLOUS ALLOGRAFT. REMEDIAL ACTION BY THE SURGEON INCLUDED REVISION SURGERY FOR REMOVAL OF THE I-FACTOR FROM THE SPINAL CANAL AROUND THE LEFT L5 NERVE ROOT. THE SURGERY RESULTED WITH PROMPT SENSORY RETURN AND SLOW IMPROVEMENT OF FOOT DROP.

    Patient problems: Nerve Damage

    MAUDE report 9321330 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140019. One submission often covers many models, so check the product and lot details in each record.

  • Z-1056-2022 · Apr 14, 2022 · Terminated
    putty i-FACTOR Peptide Enhanced Bone Graft 1.0cc Syringe REF 700-010, PUTTY

Other Filler, Bone Void, Synthetic Peptide devices

All Filler, Bone Void, Synthetic Peptide devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page