FLASH MINI Ostial System OCB4008BARecalls & reported problems
Verge Medical, Inc. · FLASH MINI Ostial System, Dual Balloon Angioplasty Catheter, 4.0mm x 8mm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI M917OCB4008BA0), received between Jan 20, 2017 and Mar 27, 2025 — classified as 2 deaths, 3 malfunctions.
The most frequently reported specific device problem is “Device Handling Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Handling Problem3
- Activation, Positioning or Separation Problem2
- Improper or Incorrect Procedure or Method2
- Compatibility Problem1
- Deflation Problem1
- Occlusion Within Device1
- Torn Material1
- Use of Device Problem1
Patient problems
- Death2
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- Perforation of Vessels1
- Vascular Dissection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2025Use of Device Problem; Use of Device Problem
THE EVENT DESCRIPTION WAS PROVIDED TO OSTIAL VIA EMAIL BY THEIR SALES REPRESENTATIVE, ALONG WITH USER FACILITY REPORT MDR # (B)(4) FROM THE FDA. THE DESCRIPTION STATED THAT DURING A COMPLEX INTERVENTION, ATTEMPTS WERE MADE TO ADVANCE AN OSTIAL FLASH BALLOON INTO THE CORONARY ARTERY FOR ANGIOPLASTY. THE BALLOON WOULD NOT ADVANCE ALL THE WAY THROUGH THE GUIDE AND HAD TO BE REMOVED. THE FLASH DEVICE WAS PROMPTLY REMOVED FROM SERVICE, AND THERE WAS NO HARM TO THE PATIENT. THE USER FACILITY REPORT MDR # (B)(4) INDICATED A DEVICE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21710525 ↗MalfunctionMar 2, 2024Device Handling Problem; Device Handling Problem
THE USER FACILITY REPORTED DIFFICULTY PREPARING THE FLASH DEVICE. NO ADDITIONAL DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18823349 ↗DeathAug 10, 2017Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Device Handling Problem; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Device Handling Problem
THE FOLLOWING INFORMATION REGARDING THE REPORTED EVENT WAS OBTAINED FROM THE HOSPITAL: FOLLOWING TREATMENT OF A RAMUS INTERMEDIUS ARTERY, THE PHYSICIAN ATTEMPTED TO USE THE FLASH MINI OSTIAL SYSTEM IN THE PATIENT'S RCA. WHEN INFLATING THE PROXIMAL BALLOON, BOTH ENDS OF THE BALLOON WERE NOTED AS EXPANDING (DOG BONE), RESULTING IN A VESSEL DISSECTION. THE PATIENT SUBSEQUENTLY CODED AND EXPIRED.
Patient problems: Death; Vascular Dissection
MAUDE report 6787283 ↗MalfunctionFeb 10, 2017Deflation Problem; Occlusion Within Device; Compatibility Problem; Torn Material; Deflation Problem; Occlusion Within Device; Compatibility Problem; Torn Material
IT WAS REPORTED THAT ONE OF THE BALLOONS ON A 4.0X8MM FLASH MINI DEVICE COULD NOT BE DEFLATED AND THAT WHEN ATTEMPTING TO WITHDRAW THE STILL INFLATED BALLOON INTO THE GUIDE CATHETER THAT THE CATHETER SHAFT BROKE. THE USER WAS ABLE TO SUCCESSFULLY REMOVE BOTH PIECES OF THE FLASH MINI DEVICE BY RETRACTING THE DEVICE AND GUIDE CATHETER AS A SINGLE UNIT. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6324593 ↗DeathJan 20, 2017Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Device Handling Problem; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Device Handling Problem
A 4.0X12MM XIENCE STENT WAS DEPLOYED AT THE OSTIUM OF THE PATIENT'S RCA. THE PHYSICIAN ATTEMPTED TO POST DILATE THE STENT USING A 4.0X8MM FLASH MINI OSTIAL SYSTEM DEVICE. THE DISTAL BALLOON INFLATED AS EXPECTED; HOWEVER, WHEN THE PHYSICIAN ATTEMPTED TO INFLATE THE PROXIMAL BALLOON, IT WAS NOTED THAT THE SHAPE ON THE CINE DID NOT LOOK CORRECT. THE BALLOONS WERE DEFLATED AND IT WAS FOUND THAT VESSEL HAD TORN/RUPTURED AND THE PATIENT BECAME UNRESPONSIVE AND WENT INTO RESPIRATORY DISTRESS. ATTEMPTS TO REVIVE THE PATIENT WERE UNSUCCESSFUL AND THE PATIENT PASSED AWAY.
Patient problems: Death; Perforation of Vessels
MAUDE report 6268842 ↗
Other Catheters, Transluminal Coronary Angioplasty, Percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page