FLASH MINI Ostial System OCB4508BARecalls & reported problems
Verge Medical, Inc. · FLASH MINI Ostial System, Dual Balloon Angioplasty Catheter, 4.5mm x 8mm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M917OCB4508BA0), received between Oct 19, 2017 and Mar 27, 2025 — classified as 1 death, 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method2
- Use of Device Problem2
- Activation, Positioning or Separation Problem1
- Device Handling Problem1
- Material Rupture1
Patient problems
- Air Embolism1
- Death1
- No Clinical Signs, Symptoms or Conditions1
- Vascular Dissection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2025Use of Device Problem; Use of Device Problem
THE EVENT DESCRIPTION WAS PROVIDED TO OSTIAL VIA EMAIL BY THEIR SALES REPRESENTATIVE, ALONG WITH USER FACILITY REPORT MDR # (B)(4) FROM THE FDA. THE DESCRIPTION STATED THAT DURING A COMPLEX INTERVENTION, ATTEMPTS WERE MADE TO ADVANCE AN OSTIAL FLASH BALLOON INTO THE CORONARY ARTERY FOR ANGIOPLASTY. THE BALLOON WOULD NOT ADVANCE ALL THE WAY THROUGH THE GUIDE AND HAD TO BE REMOVED. THE FLASH DEVICE WAS PROMPTLY REMOVED FROM SERVICE, AND THERE WAS NO HARM TO THE PATIENT. THE USER FACILITY REPORT MDR # (B)(4) INDICATED A MALFUNCTION OF DEVICES. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR PRODUCT EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21710485 ↗InjuryJan 14, 2023Material Rupture; Use of Device Problem; Improper or Incorrect Procedure or Method; Material Rupture; Use of Device Problem; Improper or Incorrect Procedure or Method
THE EVENT DESCRIPTION WAS PROVIDED TO OSTIAL IN AN EMAIL RECEIVED FROM FDA FOR THE USER FACILITY REPORT MDR#: 1001760000-2022-8008. THE EVENT DESCRIPTION IN THE MDR INDICATED THE PROXIMAL BALLOON BURST DUE TO THE USE OF A STANDARD MANIFOLD SYRINGE FOR INJECTION (10 ML) INSTEAD OF THE SUPPLIED SYRINGE CALLED FOR IN THE IFU (1 ML) WHICH MAY HAVE INADVERTENTLY CAUSED THE BALLOON TO BURST. ADDITIONALLY, AN "AIR EMBOLISM" WAS OBSERVED BY THE PHYSICIAN TRAVELING DOWN THE CORONARY ARTERY. PER THE MDR REPORT, THIS DID NOT CAUSE ANY PERMANENT ADVERSE OUTCOMES AND REQUIRED ONLY INITIAL MONITORING TO PREVENT HARM.
Patient problems: Air Embolism
MAUDE report 16169484 ↗DeathOct 19, 2017Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Device Handling Problem; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Device Handling Problem
THE FOLLOWING INFORMATION REGARDING THE REPORTED EVENT WAS OBTAINED FROM THE HOSPITAL AND LOCAL SALES REPRESENTATIVES: THE PATIENT WAS ADMITTED TO THE ER AND THEN TAKEN TO THE CATH LAB FOR TREATMENT OF STEMI. THE PHYSICIAN DEPLOYED A 3.0MM DIAMETER STENT IN A CALCIFIED CHRONIC TOTAL OCCLUSION OF THE RCA AND THEN REQUESTED A FLASH DEVICE FOR USE IN POST-DILATION OF THE STENT. THE CATH LAB DID NOT HAVE ANY 3.0MM DIAMETER FLASH DEVICES AT THE TIME, ONLY A 4.5MM DIAMETER FLASH DEVICE WAS AVAILABLE. THE PHYSICIAN DECIDED TO USE THE HIGHER DIAMETER 4.5MM DIAMETER FLASH DEVICE TO POST DILATE THE 3.0MM STENT. FOLLOWING INFLATION OF THE FLASH DEVICE, A DISSECTION OCCURRED AND THE VESSEL SHUT DOWN. CPR WAS IMMEDIATELY STARTED; HOWEVER, THE PATIENT SUBSEQUENTLY EXPIRED.
Patient problems: Death; Vascular Dissection
MAUDE report 6962212 ↗
Other Catheters, Transluminal Coronary Angioplasty, Percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page