AccuStick IIRecalls & reported problems across all models
Boston Scientific Corporation · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 51
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls
- 0
We found 5 models of AccuStick II by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 51 adverse event reports in the FDA's MAUDE database, received between Jan 19, 2017 and Jul 16, 2026 — classified as 12 injuries, 39 malfunctions.
The most frequently reported specific device problem is “Defective Device” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported AccuStick II problems
Device problems
- Defective Device20
- Detachment of Device or Device Component11
- Adverse Event Without Identified Device or Use Problem6
- Material Separation5
- Break3
- Material Deformation3
- Nonstandard Device3
- Physical Resistance/Sticking3
Patient problems
- No Clinical Signs, Symptoms or Conditions31
- Foreign Body In Patient5
- No Consequences Or Impact To Patient5
- Device Embedded In Tissue or Plaque4
- No Patient Involvement3
- No Known Impact Or Consequence To Patient2
- Discomfort1
- Internal Organ Perforation1
AccuStick II models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- AccuStick™ II M001207020ACCESSORIES, CATHETER15 reports
- AccuStick™ II M001207100ACCESSORIES, CATHETER12 reports
- AccuStick™ II M001207050ACCESSORIES, CATHETER11 reports
- AccuStick™ II M001207040ACCESSORIES, CATHETER10 reports
- AccuStick™ II M001207030ACCESSORIES, CATHETER3 reports
Recent AccuStick II adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryJul 16, 2026Fracture
It was reported that the microwire fractured, requiring additional intervention.During placement of a left renal abscess drain, an Accustick II system was selected for use. While manipulating the microwire, a portion of the microwire sheared off and remained within the patient. Because infected urine was present, only minimal retrieval attempts were made during the initial procedure. Nine days later, the patient returned for percutaneous retrieval of the retained wire fragment during removal of the indwelling abscess drain. Retrieval using a snare was unsuccessful, and the retained fragment could not be removed percutaneously. Further management with retrograde ureteroscopy or surgery was recommended if clinically indicated. The patient was expected to fully recover, and no patient complications were reported as a result of this event.
Patient problems: Foreign Body In Patient
MAUDE report 25847611 ↗MalfunctionMay 15, 2026Nonstandard Device; Deformation Due to Compressive Stress
IT WAS REPORTED THAT DISTAL TIP DETACHMENT OCCURRED AND REMAINED INSIDE THE PATIENT. THE VASCULAR ACCESS WAS THROUGH THE LEFT HEPATIC DUCT. A 0.038 J ACCUSTICK II WAS SELECTED FOR USE IN A PERCUTANEOUS TRANSHEPATIC CHOLANGIOGRAPHY (PTC) WITH PLACEMENT OF LEFT INTERNAL/EXTERNAL BILIARY DRAIN. UPON ACCESS INTO THE BILIARY TREE AND DURING MANIPULATION OF THE WIRE, IT WAS DISCOVERED THAT THE WIRE HAD KINKED. CONSEQUENTLY, THE DISTAL TIP SHEARED OFF DURING CONTINUED MANIPULATION. THE TIP REMAINED IN THE BILIARY SYSTEM AND WAS NOT RETRIEVED.
Patient problems: Foreign Body In Patient
MAUDE report 25186854 ↗MalfunctionMar 24, 2026Nonstandard Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE RADIO OPAQUE TIP FELL OFF. AN ACCUSTICK WAS SELECTED FOR USE. DURING THE PROCEDURE, WHEN REMOVING THE DEVICE, IT WAS NOTED THAT THE RADIO OPAQUE TIP FELL OFF AND WAS RETRIEVED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTER THAT THE RADIO OPAQUE TIP WHEN RETRIEVED WAS BARELY ON WHEN REMOVED AND THEN FELL OFF WHEN IT WAS OUT OF THE BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24676234 ↗MalfunctionMar 2, 2026Nonstandard Device
IT WAS REPORTED THAT THE MARKER BAND WAS NOT FIXED IN PLACE AND WAS MOVING. DURING THE PROCEDURE WHILE USING THIS ACCUSTICK II DEVICE, IT WAS NOTED THAT THE MARKER BAND WAS NOT FIXED IN PLACE AND WAS MOVING. THEREFORE, ANOTHER ACCUSTICK DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490073 ↗MalfunctionFeb 12, 2026Entrapment of Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEVICE GOT STUCK WITH THE MICROWIRE. THE TARGET LESION WAS LOCATED IN THE LEFT KIDNEY. THE PATIENT HAD A LARGE RENAL PELVIS STONE AND NEEDED A NEPHROURETERAL CATHETER PLACEMENT FOR A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) WITH UROLOGY. AN .038 ACCUSTICK II INTRODUCER WAS SELECTED FOR USE. DURING THE PROCEDURE, THE ACCUSTICK SET WAS USED TO GAIN ACCESS STARTING WITH THE INTRODUCER NEEDLE WITH STYLET. ONCE PROPER ACCESS WAS ACHIEVED, THE STYLET WAS REMOVED AND THE MICRO WIRE WAS INSERTED. THE NEEDLE WAS THEN EXCHANGED FOR THE CO-AXIAL INTRODUCER. HOWEVER, UPON INSERTION OF THE INTRODUCER INTO THE PATIENT'S RENAL PELVIS, THE MICRO WIRE BECAME STUCK WITHIN THE INTRODUCER, PREVENTING REMOVAL OF THE INNER DILATOR AND MICRO WIRE. AS A RESULT, THE ACCESS SITE WAS ABANDONED, AND A NEW ACCUSTICK II SET WAS USED TO RESTART THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24327772 ↗MalfunctionJan 14, 2026Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT HUB DETACHMENT OCCURRED. A ACCUSTICK II WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE WIRE GOT BROKEN. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24070324 ↗InjuryJun 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT DEVICE FRACTURE OCCURRED REQUIRING ADDITIONAL INTERVENTION. AN ACCUSTICK II WAS SELECTED FOR USE. DURING THE PROCEDURE, THE DEVICE CAME APART WHILE INSIDE THE PATIENT. A BALLOON WAS PLACED PAST THE BROKEN ACCUSTICK TO RETRIEVE, HOWEVER THE BALLOON POPPED. A SECOND BALLOON WAS USED AND WAS ABLE TO PULL IT BACK OVER THE WIRE THAT WAS IN PLACE. THE DEVICE WAS REMOVED IN 2 (TWO) PIECES AND PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22186948 ↗InjuryAug 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE GUIDEWIRE DETACHED INSIDE THE PATIENT. ON (B)(6) 2024, A PATIENT UNDERWENT A RIGHT NEPHROSTOMY TUBE PLACEMENT PROCEDURE DURING WHICH A GUIDEWIRE BROKE, LEAVING A 30.2 CM PORTION RETAINED IN THE PATIENT'S RIGHT KIDNEY PELVIS. THE RADIOLOGIST REPORTED THAT NO UNDUE STRESS, SUCH AS TUGGING, PUSHING, OR PULLING, WAS APPLIED TO THE GUIDEWIRE PRIOR TO ITS BREAKAGE, AND THE BROKEN SEGMENT COULD NOT BE RETRIEVED DURING THE INITIAL PROCEDURE. SUBSEQUENT TO THE INCIDENT, THE PATIENT REQUIRED TWO ADDITIONAL PROCEDURES TO ADDRESS THE RETAINED GUIDEWIRE. ON (B)(6) 2024, A RIGHT URETEROSCOPIC LASER LITHOTRIPSY WAS PERFORMED WITH THE AIM OF REMOVING THE FOREIGN BODY. DURING THIS PROCEDURE, A 1.6 CM PIECE OF THE GUIDEWIRE WAS SUCCESSFULLY EXTRACTED FROM THE RIGHT KIDNEY PELVIS; HOWEVER, THE REMAINING PORTION COULD NOT BE REMOVED. A SECOND PROCEDURE WAS PERFORMED ON (B)(6) 2024, INVOLVING A RIGHT PARTIAL URETERECTOMY AND URETEROPYELOSTOMY WITH DOUBLE-J STENT PLACEMENT. THIS PROCEDURE SUCCESSFULLY REMOVED THE REMAINING 28.6 CM PORTION OF THE GUIDEWIRE. THE PATIENT CONDITION REMAINED STABLE, AND THERE WERE NO ADVERSE EFFECTS FROM THE RETAINED GUIDEWIRE. THE PATIENT WAS DISCHA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20015520 ↗MalfunctionFeb 9, 2024Defective Device; Defective Device
IT WAS REPORTED THAT MARKERBAND DISPOSITION OCCURRED. A: 038 ACCUSTICK II WAS SELECTED FOR USE IN THE NEPHROSTOMY CATHETER PLACEMENT. DURING THE PROCEDURE, THE RADIOPAQUE (RO) MARKER STARTED IN ITS ORIGINAL POSITION. HOWEVER, UPON INSERTION INTO THE PATIENT, IT WAS NOTED THAT THE RO SLID BACK ON THE INTRODUCER. THE DEVICE WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18676151 ↗MalfunctionJan 3, 2024Defective Device; Adverse Event Without Identified Device or Use Problem; Defective Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE MARKERBAND WAS DISLODGED. THE TARGET LESION WAS LOCATED IN THE KIDNEY. AN ACCUSTICK II W/O GUIDEWIRE WAS SELECTED FOR USE. AFTER WITHDRAWING THE DEVICE FROM THE PATIENT, IT WAS NOTED THAT THE SILVER MARKERBAND WAS FOUND IN THE UPPER PART OF THE DILATOR AND NOT AT THE END OF THE INTRODUCER. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18438597 ↗MalfunctionDec 11, 2023Defective Device; Defective Device
IT WAS REPORTED THAT MARKERBAND DISPOSITION OCCURRED. THE TARGET LESION WAS LOCATED IN THE BILE DUCT. AN ACCUSTICK II WAS SELECTED FOR USE IN THE PERCUTANEOUS TRANSHEPATIC CHOLANGIOGRAM. DURING THE PROCEDURE, IT WAS NOTED THAT THE RADIOPAQUE (RO) MARKER IS VERY NEAR THE SHAFT. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A NEW ACCUSTICK. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18303800 ↗MalfunctionSep 18, 2023Defective Device; Defective Device
IT WAS REPORTED THAT THE RO MARKER MOVED ALONG THE CATHETER SHAFT. THIS ACCUSTICK II INTRODUCER CATHETER WAS SELECTED FOR USE TO TREAT BILIARY DUCT OBSTRUCTION. DURING PERCUTANEOUS INSERTION, THE RO MARKER SLID BACK ALONG THE SHEATH PROXIMALLY TOWARD THE PHYSICIAN'S HANDS. IT WAS CONFIRMED THAT THE RO MARKER DIDN'T DETACH/SEPARATE FROM THE TIP OF THE DEVICE (ONLY SLID BACK). THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17764702 ↗MalfunctionSep 14, 2023Defective Device; Adverse Event Without Identified Device or Use Problem; Defective Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE RO MARKER WAS PUSHED. THE TARGET LESION WAS LOCATED IN THE BILIARY. A ACCUSTICK II W/O GUIDIWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, THE DILATOR WAS INSERTED INTO THE PATIENT, BUT COULD NOT SEE THE ANGIOMARK. THE DEVICE WAS REMOVED FROM THE PATIENT AND FOUND THAT THE ANGIOMARK WAS PUSHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17745337 ↗InjurySep 13, 2023Defective Device; Adverse Event Without Identified Device or Use Problem; Defective Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE RADIOPAQUE MARKER WAS LEFT IN THE PATIENT. THE TARGET LESION WAS LOCATED IN THE LEFT KIDNEY. A NEW NEPHOSTROMY TUBE IS BEING PLACED AND A .038 ACCUSTICK II INTRODUCER SYSTEM WAS SELECTED FOR USE. A .018 WIRE AND AMPLATZ GUIDEWIRE WAS PLACED THROUGH THE 0.38 LUMEN. DURING REMOVAL OF THE DILATOR FROM THE PATIENT, THE RADIOPAQUE MARKER WAS RETAINED IN THE SKIN. THE RADIOPAQUE MARKER WAS ATTEMPTED TO RETRIEVE PERCUTANEOUSLY WITH HEMOSTATS, A DILATOR AND A NON-BOSTON SCIENTIFIC BALLOON CATHETER, BUT IT WAS UNSUCCESSFUL. EVENTUALLY, A NEPHROSTOMY TUBE WAS THEN PLACED, AND THE RADIOPAQUE MARKER WAS LEFT IN THE RENAL PELVIS. THE PATIENT IS DOING FINE.
Patient problems: Foreign Body In Patient
MAUDE report 17741274 ↗MalfunctionAug 24, 2023Defective Device; Defective Device
IT WAS REPORTED THAT THE RADIOPAQUE MARKER WAS IN THE WRONG SPOT. A .038 ACCUSTICK II WAS SELECTED FOR USE. DURING THE COURSE OF THE PROCEDURE, IT WAS NOTED THAT THE RADIOPAQUE MARKER WAS IN THE WRONG SPOT. THERE WAS NO HARM TO THE PATIENT, AS IT WAS DETECTED BEFORE USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17618461 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page