Acumatch L-SeriesRecalls & reported problems across all models
Exactech, Inc. · 9 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Acumatch L-Series model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 1
We found 9 models of Acumatch L-Series by Exactech, Inc. in the FDA's device registry (GUDID). Together they are named in 3 adverse event reports in the FDA's MAUDE database, received between Oct 15, 2019 and Feb 12, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Acumatch L-Series recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 26, 2024Open, Classified
The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.
Z-2236-2024 · Root cause: Other · names 6 of these models · Recall details · FDA record ↗
Most reported Acumatch L-Series problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Naturally Worn1
Patient problems
- Bone Fracture(s)1
- Failure of Implant1
- Fall1
- Joint Dislocation1
Acumatch L-Series models
9 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Acumatch L-Series 100-22-19Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 recall0 reports
- Acumatch L-Series 100-22-20Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 recall0 reports
- Acumatch L-Series 100-22-21Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 recall0 reports
- Acumatch L-Series 100-22-22Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 recall0 reports
- Acumatch L-Series 100-22-23Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 recall0 reports
- Acumatch L-Series 100-22-24Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 recall0 reports
- Acumatch L-Series BP-2844Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 report
- Acumatch L-Series BP-2850Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 report
- Acumatch L-Series BP-2853Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented1 report
Recent Acumatch L-Series adverse event reports
The 3 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THIS PATIENT'S HIP WAS REVISED APPROXIMATELY 3.5 YEARS POST INITIAL OPERATION. THE PATIENT FELL DOWN AND HAD DISLOCATION OF BIPOLAR CUP. NEXT DAY, THIS PATIENT WENT TO SEE A DOCTOR AND SURGEON FOUND OUT DISASSOCIATION OF FEMORAL HEAD FROM BIPOLAR CUP VIA IMAGING SYSTEM WHEN HE PERFORMED A CLOSED/MANUAL REDUCTION BY USING IMAGING SYSTEM. IN THIS SITUATION, HE DECIDED TO PERFORM REVISION SURGERY AND FEMORAL HEAD AND BIPOLAR CUP WERE REPLACED TO NEW ONES. THIS FEMORAL HEAD CAME OFF THIS BIPOLAR CUP WITHOUT DISASSOCIATION OF LOCKING RING.
Patient problems: Fall; Joint Dislocation
MAUDE report 21364026 ↗InjuryJan 15, 2020Naturally Worn; Naturally Worn
AS REPORTED, THIS (B)(6) Y/O ACTIVE FEMALE EXPERIENCED A IMPLANTATION OF THE FEMORAL HEAD REPLACEMENT FOR LEFT FEMORAL NECK FRACTURE WAS PERFORMED USING NOVATION TAPERED STEM, BIPOLAR CUP, AND METAL HEAD ON (B)(6) 2014. THE PATIENT IS NOW UNDERGOING A RIGHT TOTAL HIP ARTHROPLASTY REVISION DUE TO PROXIMAL MIGRATION OF THE BIPOLAR CUP WAS OBSERVED BY X-RAY REVIEW. THE PROCEDURE WAS ON (B)(6) 2019 AT POST-OPERATIVE 5-YEARS AND 9-MONTHS. AT THE TIME OF THE RE-OPERATION, THE BIPOLAR CUP AND THE METAL HEAD WAS REMOVED BUT A REMARKABLE DELAMINATION WEAR THAT MIGHT BE CAUSED BY OXIDATION WAS OBSERVED AROUND THE LOCKING RING OF THE POLYETHYLENE BIPOLAR LINER. PATIENT¿S WEIGHT (B)(6) LBS, HEIGHT 4¿7¿. DEVICES ARE TO BE RETURNED FOR INVESTIGATION.
Patient problems: Failure of Implant
MAUDE report 9591528 ↗InjuryOct 15, 2019Adverse Event Without Identified Device or Use Problem
REVISION OF A BIPOLAR HEAD DUE TO THE PATIENT WAS SUPER ACTIVE AND FRACTURED THEIR PELVIS. THE PATIENT WAS STABLE AS THEY LEFT THE OPERATING ROOM AND THIS DID NOT CAUSE A DELAY. IN ACCORDANCE WITH HOSPITAL POLICY WE ARE UNABLE TO RETURN THE EXPLANTS. NO OTHER INFORMATION AVAILABLE AT THIS TIME. UPON REVIEW OF THE AVAILABLE INFORMATION, THERE IS NO EVIDENCE THAT THIS IS A DEVICE RELATED PROBLEM AND THERE IS NO ALLEGATION AGAINST THE DEVICE. THE MOST LIKELY CAUSE FOR THE BONE FRACTURE WAS THE PATIENT¿S ACTIVITY.
Patient problems: Bone Fracture(s)
MAUDE report 9195304 ↗
Other Exactech, Inc. products
- Equinoxe305 models · 2,076 reports · 37 recalls
- Optetrak Logic408 models · 3,306 reports · 3 recalls
- Novation184 models · 1,145 reports · 25 recalls
- Optetrak336 models · 1,773 reports · 6 recalls
- Truliant414 models · 1,010 reports · 2 recalls
- Acumatch A Series94 models · 54 reports · 15 recalls
- MCS25 models · 7 reports · 5 recalls
- Alteon135 models · 100 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page