Alexis Laparoscopic SystemRecalls & reported problems across all models
Applied Medical Resources Corporation · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 2 models of Alexis Laparoscopic System by Applied Medical Resources Corporation in the FDA's device registry (GUDID). Together they are named in 27 adverse event reports in the FDA's MAUDE database, received between Sep 13, 2016 and Apr 29, 2026 — classified as 27 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Alexis Laparoscopic System problems
Device problems
- Material Split, Cut or Torn10
- Material Perforation5
- Device Contaminated During Manufacture or Shipping4
- Detachment of Device or Device Component2
- Material Separation2
- Detachment Of Device Component1
- Entrapment of Device1
- Incomplete or Inadequate Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Consequences Or Impact To Patient6
- No Patient Involvement2
- No Known Impact Or Consequence To Patient1
Alexis Laparoscopic System models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Alexis Laparoscopic System C8501Ring (Wound Protector), Drape Retention, Internal18 reports
- Alexis Laparoscopic System C8502Ring (Wound Protector), Drape Retention, Internal9 reports
Recent Alexis Laparoscopic System adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2026Material Split, Cut or Torn
PROCEDURE PERFORMED: ANTERIOR RESECTION. EVENT DESCRIPTION: "IT WAS TORN." PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL ON 07APR2026 FROM [NAME], AMK TERRITORY MANAGER. "IT'S UNKNOWN DURING WHAT STEP IN THE PROCEDURE THE MALFUNCTION OCCURRED. IT'S UNKNOWN WHERE THE SHEATH SEPARATED/TORE FROM. IT'S IMPOSSIBLE TO PROVIDE THE PICTURES. IT'S UNKNOWN WHETHER ANY INSTRUMENTS WERE USED WITHIN THE ALEXIS DURING PLACEMENT. IT'S UNKNOWN WHETHER ANY INSTRUMENTS WERE USED WITHIN THE ALEXIS DURING THE PROCEDURE. IT'S UNKNOWN WHETHER IT PARTICULATED." INTERVENTION: NI. PATIENT STATUS: THERE WAS NO PROBLEM WITH THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25035360 ↗MalfunctionMar 5, 2026Device Contaminated During Manufacture or Shipping
NAME OF PROCEDURE BEING PERFORMED: NI (BEFORE USE). DETAILED DESCRIPTION OF EVENT: UPON OPENING PACKAGE DURING SURGERY PREP, THE NURSE FOUND A HAIR INSIDE THE PACKAGING. ADDITIONAL INFORMATION PROVIDED BY [NAME] ON 20FEB2026 VIA EMAIL: THANK YOU FOR THE HELP, I HAVE TAKEN PHOTOS AND ATTACHED AS ASKED. IF THERE IS ANYTHING I CAN DO TO HELP, JUST LET ME KNOW. TYPE OF INTERVENTION: NI (BEFORE USE). PATIENT STATUS: NI (BEFORE USE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24515492 ↗MalfunctionNov 20, 2025Entrapment of Device; Entrapment of Device
PROCEDURE PERFORMED: ROBOT-ASSISTED LOW ANTERIOR RESECTION EVENT DESCRIPTION: THIS IS A CER WHICH CAME FROM FEEDBACK. HOSPITAL [NAME]. COMPLAINT; WHAT HAPPENED. IN EARLY (B)(6) 2025, A PROBLEM HAPPENED DURING A ROBOT-ASSISTED LOW ANTERIOR RESECTION AT [FACILITY] (GASTROINTESTINAL SURGERY). THE SURGEON USED C8501 AS THE CAMERA PORT OF [ROBOT]. DURING THE OPERATION, A CLIP APPLIER INSERTED FROM ANOTHER PORT GOT CAUGHT ON THE TETHER CORD OF C8501. THE CORD WAS ENTANGLED WITH THE IMA (INFERIOR MESENTERIC ARTERY). THIS HAPPENED OUTSIDE THE CAMERA VIEW, SO IT WAS VERY DANGEROUS. THE SURGEON, DR. [NAME], SAID THERE MIGHT HAVE BEEN SLACK IN THE CORD INSIDE THE ABDOMEN. ADDITIONAL INFORMATION PROVIDED BY [NAME] ON 20AUG2025: THE [ROBOT] PORT SETUP WAS: PORT 1: ASSIST PORT (RIGHT) PORT 2: CAMERA PORT 3: SURGEON¿S LEFT HAND PORT 4: SURGEON¿S RIGHT HAND THREE ARMS WERE USED. THE CLIP APPLIER WAS INSERTED FROM PORT 1 (ASSIST PORT), WHICH IS LOCATED TO THE RIGHT OF THE CAMERA PORT (PORT 2). THE SURGEON PLACED THE ALEXIS LAPAROSCOPIC SYSTEM (C8501). NOTE: THE CAP WAS NOT USED. THE TETHER WAS KEPT OUTSIDE THE PATIENT. WE COULD NOT CONFIRM WHETHER THE SURGEON SWEEPED THE INNER RING AFTER DEPLOYING
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23612482 ↗MalfunctionNov 1, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NAME OF PROCEDURE BEING PERFORMED: UNKNOWN. DETAILED DESCRIPTION OF EVENT: AI TRANSLATED: 1. REF C8502: ALEXIS LAPAROSCOPIC SYSTEM M 5-9CM, A. UNSTERILE: HAIR FOUND BACK IN THE SEALED PACKAGE. PLEASE RETURN WHAT TO DO WITH THESE PIECES. OUR OPERATIONS QUARTER HAS NOTICED THAT A HAIR GOT CAUGHT IN A STERILE PACKED PRODUCT. EVENT DESCRIPTION IN PCF: WHEN TAKING THE ALEXIS, THEY SAW A HAIR IN THE PACKAGE (SEE PICTURES). PATIENT STATUS: PATIENT IS OK. TYPE OF INTERVENTION: THEY TOOK A NEW ALEXIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20587419 ↗MalfunctionMar 29, 2024Tear, Rip or Hole in Device Packaging; Material Split, Cut or Torn; Tear, Rip or Hole in Device Packaging; Material Split, Cut or Torn
PROCEDURE PERFORMED: RIGHT HEMICOLECTOMY. DESCRIPTION OF EVENT: [NAME] MEDICAL CENTER. "THE PACKAGING WAS FOUND TO BE OPENED BEFORE USE." PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS RECEIVED VIA PHONE ON 27MAR2024 FROM [NAME] AMK TERRITORY MANAGER "THERE WAS MISCOMMUNICATION BETWEEN THE HOSPITAL AND THE DEALER. THE EVENT IS NOT ABOUT THE OPEN PACKAGE. THE EVENT IS THAT THE SHEATH WAS FOUND TORN." TYPE OF INTERVENTION: THE CASE WAS COMPLETED WITH THE NEW DEVICE. PATIENT STATUS: NA (NOT USED ON PATIENT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19006986 ↗MalfunctionFeb 2, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
PROCEDURE PERFORMED: NI. DESCRIPTION OF EVENT: ALEXIS IN USE & SPLIT. TYPE OF INTERVENTION: NI. PATIENT STATUS: NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18634104 ↗MalfunctionDec 11, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
PROCEDURE PERFORMED: LAP. RESECTION RECTOPEXY. DESCRIPTION OF EVENT: THE ALEXIS WAS PLACED INTO AN EXTENDED TROCAR INCISION, THE OUTER RETRACTION RING WAS ROLLED SEVERAL TIMES OUTWARDS FOR RETRACTION, THEN THE CAP WAS PLACED ONTO THE OUTER EXTRACTION RING AND TROCAR HAS BEEN INSERTED THROUGH THE CAP. WHEN TRYING TO REESTABLISH PNEUMOPERITONEUM A LEAKAGE WAS DETECTED. WHEN LOOKING FOR THE SOURCE THE ALEXIS SHEATH WAS IDENTIFIED AS THE PROBLEM. THE MATERIAL HAD A LARGE ALMOST 180° RIFT. NO CLINICAL IMPACT TO THE PATIENT, THERE WAS ONLY A LOSS OF TIME. THE SURGEON EXTRACTED THE ALEXIS LAP. SYSTEM, LIGATED THE FASCIA AND REINSERTED A TROCAR THROUGH A SMALL LEFTOVER GAP. NO OTHER INSTRUMENTS USED- THE SURGEON ALSO POINTED OUT THAT NO SHARP INSTRUMENTS COULD HAVE CAUSED THE DAMAGE. IMAGES AVAILABLE IN ATTACHMENTS. INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 21NOV23: THE SURGEON USED A REUSABLE TROCAR WITHOUT OBTURATOR. THE INSUFFLATION-PRESSURE WAS SET TO 12 MMHG. IT WAS A FIRST INCISION. NO SUTURES OR STAPLES, NO PREVIOUS SURGERIES. TYPE OF INTERVENTION: THE SURGEON EXTRACTED THE ALEXIS LAP. SYSTEM, LIGATED THE FASCIA AND REINSERTED A TROCAR THROUGH A SMALL LEFTOV
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18302898 ↗MalfunctionApr 11, 2023Material Fragmentation; Material Fragmentation
PROCEDURE PERFORMED: UNKNOWN. AI TRANSLATION: EDGE OF THE LID OF THE GEL PORT ALEXIS HAS BEEN TORN OFF AND ENTERED THE PATIENT. INTERVENTION: N/I. PATIENT STATUS: NO PATIENT INJURY HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16718650 ↗MalfunctionNov 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
PROCEDURE PERFORMED: "LAPAROSCOPIC ABDOMINAL PROCEDURE". EVENT DESCRIPTION: COMPLAINT CREATED BASED ON MEDWATCH RECEIVED BY EMAIL ON 05OCT2022. REPORT NUMBER MW5112223. "THE ALEXIS LAPAROSCOPIC SMALL WOUND RETRACTOR C8501 HAS A TRIANGULAR SHAPED PIECE THAT IS ATTACHED TO THE MAIN BODY OF THE RETRACTOR BY A [INVALID]. DURING A LAPAROSCOPIC ABDOMINAL PROCEDURE, IT WAS RECOGNIZED THAT THE TRIANGULAR PIECE WAS MISSING FROM THE [INVALID]. THE KNOT ATTACHING THE TRIANGULAR PIECE HAD APPARENTLY CAME UNDONE. AT THE TIME IT WAS RECOGNIZED, THE RETRACTOR WAS STILL IN THE BODY. THERE WAS CONCERN FOR A RETAINED FOREIGN BODY IN THE SURGICAL SITE. THE ITEM IS NOT RADIOPAQUE. THE PIECE WAS EVENTUALLY FOUND IN THE DRAPES BUT CONCERN FOR FUTURE USE AND RISK OF FOREIGN BODY RETENTION, ESPECIALLY WHEN NOT RADIOPAQUE, REMAINS. (NOTE THAT THE REPRESENTATIVE DID SAY IT WOULD BE SEEN ON MRI.) FDA SAFETY REPORT ID #(B)(4)." ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL ON 11OCT2022 FROM [NAME] THE TETHER DID NOT APPEAR TO BE BROKEN- IT APPEARED THAT THE SLIP KNOT HOLDING THE TETHER TAG CAME UNDONE. THE TETHER TAG WAS FOND IN THE DRAPES AT THE CONCLUSION OF THE PROCEDURE- IT IS NOT KNOWN AT EXACTLY WHAT PO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15712349 ↗MalfunctionSep 22, 2022Detachment of Device or Device Component; Incomplete or Inadequate Connection
THE ALEXIS LAPAROSCOPIC SMALL WOUND RETRACTOR C8501 HAS A TRIANGULAR SHAPED PIECE THAT IS ATTACHED TO THE MAIN BODY OF THE RETRACTOR BY A [INVALID]. DURING A LAPAROSCOPIC ABDOMINAL PROCEDURE, IT WAS RECOGNIZED THAT THE TRIANGULAR PIECE WAS MISSING FROM THE [INVALID]. THE KNOT ATTACHING THE TRIANGULAR PIECE HAD APPARENTLY CAME UNDONE. AT THE TIME IT WAS RECOGNIZED, THE RETRACTOR WAS STILL IN THE BODY. THERE WAS CONCERN FOR A RETAINED FOREIGN BODY IN THE SURGICAL SITE. THE ITEM IS NOT RADIOPAQUE. THE PIECE WAS EVENTUALLY FOUND IN THE DRAPES BUT CONCERN FOR FUTURE USE AND RISK OF FOREIGN BODY RETENTION, ESPECIALLY WHEN NOT RADIOPAQUE, REMAINS. (NOTE THAT THE REPRESENTATIVE DID SAY IT WOULD BE SEEN ON MRI.) FDA SAFETY REPORT ID #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15470454 ↗MalfunctionAug 10, 2022Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NAME OF PROCEDURE BEING PERFORMED: NA (INCOMING INSPECTION). DETAILED DESCRIPTION OF EVENT: [FACILITY] INCOMING INSPECTION (B)(4). WE WOULD LIKE TO INFORM YOU THAT SOME OF THE PRODUCTS DIDN'T PASS OUR RECEIVING INSPECTION. PLEASE FIND ATTACHED SPREADSHEET. THERE ARE THE ITEMS REQUIRE CER IN THIS REPORT. C8502 LOT #1448279 - PARTICULATE: SUSPECTED HAIR (1 EA.). PATIENT STATUS: NA (INCOMING INSPECTION). TYPE OF INTERVENTION: NA (INCOMING INSPECTION).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15204397 ↗MalfunctionAug 10, 2022Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
NAME OF PROCEDURE BEING PERFORMED: NA (INCOMING INSPECTION). DETAILED DESCRIPTION OF EVENT: [FACILITY] INCOMING INSPECTION (B)(4). WE WOULD LIKE TO INFORM YOU THAT SOME OF THE PRODUCTS DIDN'T PASS OUR RECEIVING INSPECTION. PLEASE FIND ATTACHED SPREADSHEET. THERE ARE THE ITEMS REQUIRE CER IN THIS REPORT. C8501 LOT #1444855 - PARTICULATE: SUSPECTED HAIR (2 EA.). PATIENT STATUS: NA (INCOMING INSPECTION). TYPE OF INTERVENTION: NA (INCOMING INSPECTION).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15204342 ↗MalfunctionJun 20, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
PROCEDURE PERFORMED: ROBOTIC SIGMOIDECTOMY. EVENT DESCRIPTION: TWO COMPLAINTS OCCURRED IN THE SAME CASE. COMPLAINT 1 OF 2: #(B)(4). COMPLAINT 2 OF 2: #(B)(4). FIELD TEAM WAS NOT PRESENT DURING DEVICE USE. THE FIRST C8501 WAS PLACED WITHIN PATIENT AND APPROXIMATELY 30 MINUTES AFTER DEPLOYMENT, THE SURGICAL TEAM FOUND THAT THE SHEATH WAS SEPARATED FROM THE GREEN INNER RING AND A PORTION OF THE RING WAS HANGING INSIDE THE PATIENT. A SECOND C8501 WAS USED AND THE SHEATH SEPARATED FROM THE GREEN INNER RING AGAIN 30 MINUTES AFTER DEPLOYMENT, SIMILAR TO THE FIRST C8501. THE CASE WAS COMPLETED WITH THE THIRD C8501. NO PATIENT INJURY OCCURRED. THE SURGICAL TEAM ALSO NOTE THAT THEY WOULD PULL OFF LOOPED CORD FROM DEVICE BEFORE USE, USING THE SAME PROCEDURE FOR 15 YEARS. IMAGES OF DEVICES ARE AVAILABLE PRODUCTS ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022 VIA EMAIL FROM ACCOUNT MANAGER. THE GREEN PLACEMENT RING DETACHED FROM THE SHEATH 30 MINUTES AFTER DEPLOYMENT FOR BOTH DEVICES. PATIENT STATUS: NO PATIENT INJURY. TYPE OF INTERVENTION: CASE WAS COMPLETED WITH THE THIRD DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14748733 ↗MalfunctionJun 20, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
PROCEDURE PERFORMED: ROBOTIC SIGMOIDECTOMY EVENT DESCRIPTION: TWO COMPLAINTS OCCURRED IN THE SAME CASE. COMPLAINT 1 OF 2: #(B)(6), COMPLAINT 2 OF 2: #(B)(6). FIELD TEAM WAS NOT PRESENT DURING DEVICE USE. THE FIRST C8501 WAS PLACED WITHIN PATIENT AND APPROXIMATELY 30 MINUTES AFTER DEPLOYMENT, THE SURGICAL TEAM FOUND THAT THE SHEATH WAS SEPARATED FROM THE GREEN INNER RING AND A PORTION OF THE RING WAS HANGING INSIDE THE PATIENT. A SECOND C8501 WAS USED AND THE SHEATH SEPARATED FROM THE GREEN INNER RING AGAIN 30 MINUTES AFTER DEPLOYMENT, SIMILAR TO THE FIRST C8501. THE CASE WAS COMPLETED WITH THE THIRD C8501. NO PATIENT INJURY OCCURRED. THE SURGICAL TEAM ALSO NOTE THAT THEY WOULD PULL OFF LOOPED CORD FROM DEVICE BEFORE USE, USING THE SAME PROCEDURE FOR 15 YEARS. IMAGES OF DEVICES ARE AVAILABLE PRODUCTS ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED ON 02JUN2022 VIA EMAIL FROM ACCOUNT MANAGER. THE GREEN PLACEMENT RING DETACHED FROM THE SHEATH 30 MINUTES AFTER DEPLOYMENT FOR BOTH DEVICES. PATIENT STATUS: NO PATIENT INJURY. TYPE OF INTERVENTION: CASE WAS COMPLETED WITH THE THIRD DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14748720 ↗MalfunctionJun 2, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
PROCEDURE PERFORMED: ROBOTIC COLECTOMY. SIMILAR C8501 MALFUNCTIONS OCCURRED IN 3 CASES. COMPLAINT 1 OF 3: #(B)(4). COMPLAINT 2 OF 3: #(B)(4). COMPLAINT 3 OF 3: #(B)(4) PRODUCT RETURNING. EVENT DESCRIPTION: EVENT DATE WAS SOMETIME IN THE PAST WEEK. C8501 WAS PLACED WITHIN PATIENT WITH 12MM TROCAR AND NO SHARP INSTRUMENTS INSERTED. APPROXIMATELY 1 HOUR INTO THE CASE, THE SURGICAL TEAM HEARD LEAKING AND FOUND THAT THE SHEATH WAS SEPARATED FROM THE GREEN INNER RING. A PORTION OF THE RING WAS HANGING FREELY FROM THE SHEATH INSIDE THE PATIENT. A NEW DEVICE WAS OPENED TO COMPLETE THE CASE. NO PATIENT INJURY OCCURRED. PRODUCT IS NOT AVAILABLE FOR RETURN. PATIENT STATUS: NO PATIENT INJURY OCCURRED. TYPE OF INTERVENTION: A NEW DEVICE WAS OPENED TO COMPLETE THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14581973 ↗
Other Applied Medical Resources Corporation products
- Inzii Retrieval System3 models · 796 reports
- Kii Fios First Entry18 models · 587 reports
- Kii Optical Access System18 models · 309 reports
- Kii Balloon Blunt Tip System3 models · 263 reports
- VOYANT 5mm Fusion Device4 models · 257 reports
- Epix Laparoscopic Scissors2 models · 245 reports
- VOYANT Maryland Fusion Device5 models · 132 reports
- Kii Sleeve7 models · 130 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page