Aqua-SealRecalls & reported problems across all models
Cardinal Health 200, LLC · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Aqua-Seal model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls
- 1
We found 5 models of Aqua-Seal by Cardinal Health 200, LLC in the FDA's device registry (GUDID). Together they are named in 40 adverse event reports in the FDA's MAUDE database, received between Jun 20, 2017 and Aug 26, 2025 — classified as 8 injuries, 32 malfunctions.
The most frequently reported specific device problem is “Decrease in Suction” (10 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Aqua-Seal recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Dec 26, 2025Open, Classified
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Z-1203-2026 · Root cause: Error in labeling · names 3 of these models · Recall details · FDA record ↗
Most reported Aqua-Seal problems
Device problems
- Decrease in Suction10
- Suction Problem10
- Suction Failure9
- Fluid/Blood Leak5
- Leak/Splash4
- Detachment of Device or Device Component3
- Insufficient Device Problem Information2
- Kinked2
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Known Impact Or Consequence To Patient8
- Pneumothorax6
- No Consequences Or Impact To Patient5
- No Patient Involvement4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Insufficient Information2
- No Code Available1
Aqua-Seal models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Aqua-Seal 8888571299APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Aqua-Seal 8888571315APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Aqua-Seal 8888571406APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Aqua-Seal 8888571299Bottle, Collection, Vacuum39 reports
- Aqua-Seal 8888571406Apparatus, Suction, Operating-Room, Wall Vacuum Powered1 report
Recent Aqua-Seal adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryAug 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED AN INFANT PATIENT WHO HAD PNEUMOTHORAX WHERE A DRAIN WAS INSERTED BUT AIR REACCUMULATED INTO THE CHEST. THE ONLY WAY TO RESOLVE THIS WAS TO MANUALLY SUCTION WITH A SYRINGE. THE CUSTOMER ASSUMED A BRONCHOPULMONARY FISTULA WAS PRESENT LEADING TO A CONTINUOUS AIR LEAK BUT DUE TO THE UNIQUE CONDITION THAT THE PATIENT HAD, THE PATIENT HAD THEIR CHEST OPENED AS PART OF THEIR MANAGEMENT. AT OPERATION, THE PATIENT DID NOT HAVE A PLEURAL LEAK IDENTIFIED AND YET AIR COULD BE FREELY ASPIRATED FROM THE CHEST. THE CUSTOMER STATED THAT THEY ARE CONCERNED THAT THERE IS A PROBLEM WITH THE UNDERWATER SEAL UNIT AS IT WAS REPLACED BY A DIFFERENT UNIT FROM ANOTHER MANUFACTURER AND THE AIR IN THE CHEST RESOLVED BOTH CLINICALLY AND ON X-RAY. THE DRAIN WAS REVIEWED ON MULTIPLE OCCASIONS BY MULTIPLE PROFESSIONALS DURING WARD ROUNDS TO ENSURE CORRECT USE AND THEN FINALLY BY 4 CONSULTANTS (ANAESTHETICS, PAEDIATRIC SURGERY X2 AND NEONATAL) AND NO ERRORS IN SET UP OR SUCTION APPLICATION WERE IDENTIFIED. ADDITIONAL INFORMATION RECEIVED ON 21AUG2025 STATED THAT THE PATIENT'S AGE IS UNKNOWN BUT UNDER 28 DAYS OLD. THE PATIENT HAD A CONGENITAL HEART DEFECT. THE OPENING OF THE CHEST WAS EXPLOR
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22898828 ↗InjuryAug 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED AN INFANT PATIENT WHO HAD PNEUMOTHORAX WHERE A DRAIN WAS INSERTED BUT AIR REACCUMULATED INTO THE CHEST. THE ONLY WAY TO RESOLVE THIS WAS TO MANUALLY SUCTION WITH A SYRINGE. THE CUSTOMER ASSUMED A BRONCHOPULMONARY FISTULA WAS PRESENT LEADING TO A CONTINUOUS AIR LEAK BUT DUE TO THE UNIQUE CONDITION THAT THE PATIENT HAD, THE PATIENT HAD THEIR CHEST OPENED AS PART OF THEIR MANAGEMENT. AT OPERATION, THE PATIENT DID NOT HAVE A PLEURAL LEAK IDENTIFIED AND YET AIR COULD BE FREELY ASPIRATED FROM THE CHEST. THE CUSTOMER STATED THAT THEY ARE CONCERNED THAT THERE IS A PROBLEM WITH THE UNDERWATER SEAL UNIT AS IT WAS REPLACED BY A DIFFERENT UNIT FROM ANOTHER MANUFACTURER AND THE AIR IN THE CHEST RESOLVED BOTH CLINICALLY AND ON X-RAY. THE PATIENT AND DRAIN WERE TRANSPORTED TO ANOTHER HOSPITAL TO FACILITATE ECMO. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2025 STATED THAT THE PATIENT'S AGE IS UNKNOWN BUT UNDER 28 DAYS OLD. THE PATIENT HAD A CONGENITAL HEART DEFECT. THE OPENING OF THE CHEST WAS EXPLORATORY SURGERY BASED ON THE ASSUMPTION THAT THERE WAS A BRONCHOPULMONARY FISTULA. THE PATIENT WAS TRANSFERRED TO AN ECMO SITE DUE TO THE UNRESOLVING PNEUMOTHORAX. THE PATIENT RECOVE
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22898727 ↗MalfunctionJul 24, 2024Suction Problem; Suction Problem
THE CUSTOMER REPORTED THAT THE DOCTOR FROM THE HOSPITAL STATED THAT WHEN MAKING THE CONNECTION AFTER THE PROCEDURE, HE COULDN¿T PERFORM A SUCTION, SO HE HAD TO USE A NEW EQUIPMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19820102 ↗MalfunctionJun 19, 2024Suction Failure; Suction Failure
CUSTOMER REPORTS THAT THEY RECEIVED CALL FROM THE MOH NEONATAL ICU DOCTOR THAT THE AQUA SEAL DRAIN WAS NOT WORKING WELL AND STOP BUBBLING OR TIDAL LING AFTER A FEW MINUTES, THE PATIENT WAS ON SUCTION 20CM, WHEN THEY DO READJUST WORKS FOR FEW MINUTES THEN AGAIN STOPPED AND DON¿T DRAIN FLUID WELL. THE PATIENT WAS DIAGNOSED WITH PLEURAL EFFUSIONS ON A MECHANICAL VENTILATOR. THE CUSTOMER REQUESTED ANOTHER BRAND TO BE SUPPLIED FROM THE MOH STORE AND ONCE THEY REPLACED IT THERE WAS NO ISSUE AND WORKED VERY WELL. ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO SUCTION. THE VOLUME OF WATER IN THE SEAL CHAMBER WAS 2CM AND THE SUCTION WAS SET TO 20 CM. THERE WERE NO KINKS IN THE TUBING AND THE PRODUCT WAS REPLACED WITH AN ATRIUM BRAND. THE PATIENT IS FINE WITH NO COMPLICATION AT THAT TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19567398 ↗MalfunctionJun 19, 2024Suction Failure; Suction Failure
CUSTOMER REPORTS: THE PEDIATRIC DOCTOR USED THE AQUA SEAL FOR NEONATAL PATIENTS WITH PNEUMOTHORAX AND HAEMOTHORAX, WITH SUCTION FOR FLUID DRAIN. AFTER FEW MINUTES, THEY COULD NOT IDENTIFY ANY BUBBLING OR TIDAL LING AND THEY DO READJUSTMENT SEVERAL TIMES BUT AGAIN IT STOPPED WORKING WHICH AFFECTS THE WORKFLOW AND CONSUMES THE TIMING AND PROLONGS RECOVERY TIME FOR THE PATIENT. WHEN ANOTHER BRAND WAS USED, IT WAS WORKING WELL WITHOUT ANY ISSUE; THE BUBBLING, TIDAL LING AND DRAIN WORKING WELL. ADDITIONAL INFORMATION WAS RECEIVED THAT THE BUBBLING DID WORK FOR A FEW MINUTES. THERE WAS NO SUCTION. THE VOLUME OF WATER IN THE SEAL CHAMBER WAS 2CM AND THE SUCTION WAS SET TO 20 CM. THERE WERE NO KINKS IN THE TUBING AND THE PRODUCT WAS REPLACED WITH AN ATRIUM BRAND. THERE WERE NO COMPLICATIONS AND PATIENT IS FINE AT THAT TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19567274 ↗MalfunctionJun 18, 2024Suction Failure; Suction Failure
CUSTOMER REPORTS: SOME OF OUR CUSTOMERS HAVE USED THE AQUA SEAL CDU FOR THE NEONATAL PATIENTS, AND THEY COULD NOT IDENTIFY ANY TODDLING OR BUBBLING DURING THE OPERATION. THE DOCTORS ARE SUSPECTING THAT THIS MIGHT BE DUE TO A DEVICE FAILURE. THE CUSTOMERS ARE CONCERNED IF THE AQUA SEAL CDU IS DESIGNED TO BE USED FOR NEONATAL PATIENTS AS WELL AND THEY ARE LOOKING FOR AN OFFICIAL CONFIRMATION IN WRITING FROM THE MANUFACTURER. THEREFORE, YOU ARE KINDLY REQUESTED TO CLARIFY THE ABOVE FOR OUR CUSTOMERS ACKNOWLEDGEMENT AND LATER USAGE. ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT IS 9 MONTHS OLD. THE VOLUME OF WATER IN THE SEAL CHAMBER WAS 2CM AND THE SUCTION WAS SET TO 20 CM. THERE WERE NO KINKS IN THE TUBING AND THE PRODUCT WAS REPLACED WITH AN ATRIUM BRAND. THE PATIENT IS FINE WITH NO COMPLICATION. THERE WAS BUBBLING AND TIDELING WHICH PROVIDED SUCTION FOR A FEW MINUTES AND IT STOPPED WORKING COMPLETELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19563773 ↗MalfunctionAug 11, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THE DOCTOR ASKED FOR ADVICE WITH REGARDS TO STEPPING DOWN THE SUCTION FROM THE SUCTION CONTROL CHAMBER OF THE AQUA-SEAL CHEST DRAIN UNIT - HOW TO REDUCE THE FLUID IN THE SUCTION CONTROL CHAMBER. THERE IS NO INFORMATION IN THE IFU ABOUT THIS. WITHIN CONVERSATION, IT WAS MENTIONED THAT BLUE DYE WAS NOTICED IN THE SUCTION CONTROL CHAMBER. THIS WAS NOTICED AS THE CDU WAS BEING DISCONNECTED DURING CLAMPING OFF. IT IS UNKNOWN IF THE WALL SUCTION WAS LEFT ON. BLUE DYE FROM THE WATER SEAL CHAMBER WAS NOTICED IN THE SUCTION CONTROL CHAMBER AFTER USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17509469 ↗MalfunctionJun 20, 2023Decrease in Suction; Decrease in Suction
THE CUSTOMER REPORTED THAT THE PACU RN PAGED TO MAKE THEM AWARE THAT THE ¿SUCTION PORT WAS LOOSE AND HAD TO BE TAPED DOWN TO HOLD SUCTION¿. THIS WAS THE FIRST OCCURRENCE. THE SCRUB TECH DID NOT NOTICE ANY IRREGULARITIES/DEFECTS UPON FILLING THE SUCTION CHAMBER PRIOR TO CONNECTING THE CHEST TUBES. UPON REMOVING THE SUCTION TUBING FROM THE PORT THE NEXT MORNING (ASSESSING FOR AIR LEAK), THE PORT WAS DISLODGING AND ATTACHED TO THE SUCTION TUBING IN WHICH THE DOCTOR DECIDED TO LEAVE THE PATIENT ON WATER SEAL ONLY INSTEAD OF CHANGING THE COMPLETE SYSTEM. ADDITIONAL INFORMATION RECEIVED ON 14JUN2023 STATED THAT THE ISSUE OCCURRED WHILE IN USE AND CAUSED DIMINISHED SUCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17164583 ↗MalfunctionMar 22, 2023Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT ON ONE OCCASION LAST WEEK (W/C 20TH FEBRUARY) THE SUCTION CONTROL CHAMBER CONTINUED BUBBLING AFTER THE SUCTION HAD BEEN TURNED OFF. THEY REPLACED THE CDU WITH A NEW AQUASEAL UNIT AND THE ISSUE RESOLVED. THERE WAS NO PROBLEM FOR THE PATIENT BECAUSE OF THIS ISSUE. ADDITIONAL INFORMATION RECEIVED: THERE WAS BUBBLING IN BOTH THE WATER SEAL CHAMBER AND THE SUCTION CONTROL CHAMBER. THE CUSTOMER MENTIONED THAT WHEN THE SUCTION WAS TURNED OFF, BUBBLING CONTINUED IN THE SUCTION CONTROL CHAMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16589965 ↗MalfunctionSep 14, 2022Suction Problem; Suction Problem
CUSTOMER REPORTS: DURING USE ON A PATIENT, THE HOSE THAT CONNECTS THE EQUIPMENT COLLAPSED SEVERAL TIMES. PER ADDITIONAL INFORMATION RECEIVED ON (B)(6)2022 THERE WAS A LEAK THAT DID NOT ALLOW THE DEVICE TO BE USED PROPERLY BUT THE USER DID NOT IDENTIFY THE EXACT LOCATION OF THE LEAK. THERE WAS DIMINISHED SUCTION AND WHEN THE USER SAW THAT THE WATER SEAL WAS NOT WORKING PROPERLY, THEY TRIED TO FIX IT BUT DECIDED TO USE A NEW DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15422337 ↗MalfunctionMay 12, 2022Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT IT WAS NECESSARY TO CHANGE THE ADULT THORACIC DRAINAGE SYSTEM FOR ONE OF THE SAME CAPACITY BECAUSE THE COLUMNS WERE MIXED WITHOUT THE EQUIPMENT FALLING. IT CAUGHT THEIR ATTENTION TO SEE THAT THE WATER FROM THE WATER SEAL THAT IS COLORED BLUE WAS IN THE LAST COLUMN OF THE DRAIN. THERE WAS NO PATIENT INJURY. THE DEVICE WAS BEING OPERATED IN THE HOSPITAL AT THE TIME OF THIS EVENT. THERE WAS NO RECIDIVISM. THERE IS NO PHOTOGRAPHIC OR PHYSICAL EVIDENCE OF THE MEDICAL DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14372294 ↗MalfunctionApr 21, 2022Decrease in Suction; Leak/Splash; Decrease in Suction; Leak/Splash
THE CUSTOMER REPORTED AN ISSUE IN THE SEALING CAP RESULTING IN THE LOSS OF NEGATIVE PRESSURE CAUSING LOSS OF DRAINAGE OF THE PATIENT'S LEFT PLEURAL EFFUSION.
Patient problems: Insufficient Information
MAUDE report 14171665 ↗InjuryApr 21, 2022Decrease in Suction; Leak/Splash; Decrease in Suction; Leak/Splash
THE CUSTOMER REPORTED THAT THE PLEUROVAC CAP DOES NOT SEAL, CAUSING LOSS OF NEGATIVE PRESSURE IN THE RIGHT LUNG TUBE, LEADING TO DECREASED DRAINAGE OF THE HEMOTHORAX CAUSING A RESIDUAL RIGHT PNEUMOTHORAX OF THE PATIENT.
Patient problems: Pneumothorax
MAUDE report 14171339 ↗MalfunctionDec 10, 2021Suction Problem; Suction Problem
THE CUSTOMER REPORTED THAT THE DEVICE HAD AN ISSUE WITH THE PRESSURE REGULATING KEY OF THE PLEUROVAC. PER ADDITIONAL INFORMATION RECEIVED FROM THE REPORTER, THE PRESSURE REGULATOR WAS NOT WORKING PROPERLY. THERE WERE SUCTION ISSUES AT THE BEGINNING OF USE UNTIL THE PRESSURE REGULATOR DID NOT WORK WELL. AS A RESULT, THE DEVICE WAS REPLACED WITH THE SAME MODEL. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT HAS LEFT THE INSTITUTION SO THERE IS NO CURRENT STATUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12977593 ↗MalfunctionJul 1, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT ONE OF THE HOSES DISCONNECTED FROM THE DRAIN AND CANNOT BE RECONNECTED AS THERE IS NO VACUUM GENERATION IN THE CIRCUIT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12100516 ↗
Other Cardinal Health 200, LLC products
- JACKSON-PRATT20 models · 13 reports · 19 recalls
- Monoject35 models · 96 reports · 11 recalls
- CONVERTORS38 models · 40 reports · 11 recalls
- Argyle33 models · 58 reports · 7 recalls
- Salem Sump31 models · 40 reports · 7 recalls
- Dover45 models · 241 reports · 2 recalls
- Kendall31 models · 51 reports · 3 recalls
- Kangaroo35 models · 130 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page