AVANTIRecalls & reported problems across all models
Cordis US Corp. · 11 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 70
- Deaths reported
- 0
- Injuries reported
- 27
- Recalls
- 0
We found 11 models of AVANTI by Cordis US Corp. in the FDA's device registry (GUDID). Together they are named in 70 adverse event reports in the FDA's MAUDE database, received between Mar 24, 2017 and Sep 11, 2026 — classified as 27 injuries, 43 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported AVANTI problems
Device problems
- Detachment of Device or Device Component30
- Device Contamination with Chemical or Other Material28
- Manufacturing, Packaging or Shipping Problem8
- Difficult to Remove7
- Adverse Event Without Identified Device or Use Problem5
- Detachment Of Device Component2
- Device Markings/Labelling Problem2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Consequences Or Impact To Patient23
- No Patient Involvement5
- Insufficient Information3
- No Known Impact Or Consequence To Patient3
- Thrombosis/Thrombus2
- Vascular Dissection2
- Foreign Body In Patient1
AVANTI models
11 models, most reported first. Each model has its own page with its reports, recalls and device details.
- AVANTI 504607XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION20 reports
- AVANTI 504605XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION16 reports
- AVANTI 504611XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION14 reports
- AVANTI 504604XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION7 reports
- AVANTI 504605SDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION3 reports
- AVANTI 504609XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION3 reports
- AVANTI 402611XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION2 reports
- AVANTI 504657ADILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION2 reports
- AVANTI 402610XDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 report
- AVANTI 504604SDILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 report
- AVANTI 504606ADILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 report
Recent AVANTI adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Device Contamination with Chemical or Other Material
As reported, during the product nationalization process of a 5F 11cm Avanti Plus catheter sheath introducer (CSI), an unknown material was identified inside of the packaging, resembling a strand of hair. There were no reported patient injuries. The device will be returned for evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50734363 ↗MalfunctionApr 20, 2026Difficult to Remove; Detachment of Device or Device Component
AS REPORTED, A 9F 11CM AVANTI PLUS CATHETER SHEATH INTRODUCER (CSI) BROKE OFF WHILE INSIDE OF THE PATIENT. THERE WAS DIFFICULTY REMOVING THE SHEATH PRIOR TO OBSERVING THE SEPARATION, AND IT WAS REMOVED BY PULLING IT OUT MANUALLY. THE SEPARATED SEGMENT WAS SUBSEQUENTLY REMOVED FROM THE PATIENT BY THE PHYSICIAN, WHO RETRIEVED IT MANUALLY WITH HIS HANDS. THERE WERE NO REPORTED INJURIES TO THE PATIENT. ADDITIONAL DETAILS WERE REQUESTED BUT WERE NOT PROVIDED. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24926894 ↗MalfunctionApr 3, 2026Detachment of Device or Device Component
AS REPORTED, DURING PREPARATION THE LUER HUB OF THE VESSEL DILATOR FOR A 5F 7.5CM AVANTI PLUS CATHETER SHEATH INTRODUCER (CSI) BROKE OFF IN THE SHEATH WHEN REMOVING THE DILATOR. A NEW 5F 7.5CM AVANTI PLUS CSI WAS USED AS A REPLACEMENT. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) AND WAS STORED IN THE PROCEDURE ROOM PRIOR TO THE PROCEDURE. THE FACILITY DOES NOT USE ROOM STERILIZATION METHODS INVOLVING ULTRAVIOLET (UV) LIGHT OR OZONE. THERE WAS NO VISIBLE DAMAGE TO THE PACKAGE PRIOR TO USE, NO DIFFICULTY INSERTING THE DILATOR INTO THE SHEATH, AND THE LUER HUB OF THE DILATOR DID NOT KINK OR BEND DURING INSERTION. THE DILATOR WAS SNAPPED INTO PLACE AT THE HUB PRIOR TO INSERTION OF THE SHEATH, AND THERE WAS NO DIFFICULTY WITHDRAWING THE DILATOR FROM THE SHEATH PRIOR TO THE SEPARATION OCCURRING. THERE WERE NO REPORTED PATIENT INJURIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777213 ↗MalfunctionApr 3, 2026Detachment of Device or Device Component
AS REPORTED, DURING PREPARATION THE PROXIMAL END OF THE VESSEL DILATOR FOR A 5F 7.5CM AVANTI PLUS CATHETER SHEATH INTRODUCER (CSI) BROKE OFF IN THE SHEATH WHEN REMOVING THE DILATOR. THERE WERE NO REPORTED PATIENT INJURIES. THE DEVICE WAS STORED AND PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), AND IT WAS KEPT IN THE PROCEDURE ROOM PRIOR TO THE PROCEDURE. THE FACILITY DOES NOT USE STERILIZATION METHODS INVOLVING ULTRAVIOLET LIGHT OR O-ZONE. THERE WAS NO VISIBLE DAMAGE TO THE PACKAGING BEFORE USE. NO DIFFICULTY WAS ENCOUNTERED WHEN INSERTING THE DILATOR INTO THE SHEATH, AND THE LUER HUB DID NOT KINK OR BEND DURING INSERTION. THE DILATOR WAS SNAPPED INTO PLACE AT THE HUB. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777127 ↗MalfunctionApr 3, 2026Detachment of Device or Device Component
AS REPORTED, THE PROXIMAL END OF THE VESSEL DILATOR FOR A 5F 7.5CM AVANTI PLUS CATHETER SHEATH INTRODUCER (CSI) BROKE OFF IN THE SHEATH WHEN REMOVING THE DILATOR. THERE WERE NO REPORTED PATIENT INJURIES. THE DEVICE WAS STORED AND PREPARED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU) AND WAS KEPT IN THE PROCEDURE ROOM PRIOR TO THE PROCEDURE. THE FACILITY DOES NOT USE ROOM STERILIZATION METHODS INVOLVING ULTRAVIOLET LIGHT OR O ZONE. THERE WAS NO VISIBLE DAMAGE TO THE PACKAGING PRIOR TO USE. THERE WAS NO DIFFICULTY INSERTING THE DILATOR INTO THE SHEATH, THE LUER HUB DID NOT KINK OR BEND DURING INSERTION, AND THE DILATOR WAS SNAPPED INTO PLACE AT THE HUB PRIOR TO INSERTION OF THE SHEATH. THE ACCESS VESSEL WAS THE RIGHT FEMORAL VEIN, WHICH WAS NOT CALCIFIED OR TORTUOUS AND HAD NO OTHER NOTED CHARACTERISTICS. NO DIFFICULTY WAS EXPERIENCED DURING INSERTION OF THE SHEATH AND DILATOR INTO THE PATIENT, AND THERE WAS NO DIFFICULTY WITHDRAWING THE DILATOR FROM THE SHEATH PRIOR TO THE SEPARATION OCCURRING. THE DEVICE WILL BE RETURNED FOR EVALUATION ALONG WITH 27 STERILE UNITS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777075 ↗MalfunctionDec 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE HEMOSTASIS VALVE OF AN 11F 11CM AVANTI+ CATHETER SHEATH INTRODUCER (CSI) SEPARATED FROM THE CATHETER AND WAS NOTED ON THE CORONARY SINUS CATHETER AFTER REMOVAL AND THERE WAS NO WAY TO PREVENT BACK BLEEDING AND SUBSTANTIAL BLOOD LOSS OTHER THAN QUICKLY REMOVING THE REST OF THE CATHETER PREMATURELY FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY. THIS FAILURE OCCURRED WHEN REMOVING CORONARY SINUS CATHETERS AT THE CONCLUSION OF THE CASE. THE CATHETER WAS PREPPED CORRECTLY BY THE OR STAFF. THE PROCEDURE WAS ALREADY COMPLETE WHEN THIS EVENT OCCURRED. THE DEVICE WAS DISCARDED AND WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23709074 ↗MalfunctionDec 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE HEMOSTASIS VALVE OF AN 11F 11CM AVANTI+ CATHETER SHEATH INTRODUCER (CSI) SEPARATED FROM THE CATHETER AND WAS NOTED ON THE CORONARY SINUS CATHETER AFTER REMOVAL AND THERE WAS NO WAY TO PREVENT BACK BLEEDING AND SUBSTANTIAL BLOOD LOSS OTHER THAN QUICKLY REMOVING THE REST OF THE CATHETER FROM THE PATIENT PREMATURELY. THERE WAS NO REPORTED PATIENT INJURY. THIS FAILURE OCCURRED WHEN REMOVING CORONARY SINUS CATHETERS AT THE CONCLUSION OF THE CASE. THE CATHETER WAS PREPPED CORRECTLY BY THE OR STAFF. THE DEVICE WAS DISCARDED AND WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23709016 ↗MalfunctionDec 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, TWO PARTS OF AN 11F 11CM AVANTI+ CATHETER SHEATH INTRODUCER (CSI) SEPARATED FROM EACH OTHER AND THE CSI REPORTEDLY FAILED PART WAY THROUGH ROBOTIC CARDIAC SURGERY PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. PACKAGING AND LOT NUMBERS WERE NOT AVAILABLE FOR RETRIEVAL. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23697501 ↗InjuryAug 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE WAS NOT RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22755553 ↗InjuryJun 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE 7F AVANTI+ SHEATH (THIRD SHEATH) HAD SIGNIFICANT RESISTANCE WHEN ATTEMPTED TO BE REMOVED. ON CLOSER INSPECTION, IT APPEARED TO HAVE SPLIT ROUGHLY 1 CM DOWN FROM THE VALVE. BLOOD WAS COMING OUT THE SPLIT. AT THE SPLIT, A SMALL PORTION OF THE PLASTIC WAS STICKING OUT WHICH WAS CATCHING ON THE SKIN (HOOK LIKE EFFECT) AND PREVENTED THE PHYSICIAN FROM REMOVING THE SHEATH. THE PHYSICIAN ADVANCED THE SHEATH A FEW TIMES WHILE ATTEMPTING THE TURN THE SHEATH HOPING TO ¿UNHOOK¿ THIS PORTION OF THE SHEATH. THIS WAS UNSUCCESSFUL. THE PHYSICIAN THEN PULLED A LITTLE HARDER IN THE HOPES THAT THE HOOK WOULD BEND AND ALLOW THE SHEATH TO BE REMOVED. THE SHEATH UNFORTUNATELY BROKE IN HALF. IT WAS SUBSEQUENTLY SUCCESSFULLY SNARED FROM THE NECK. THE CASE WAS UNEVENTFUL TILL THIS POINT. TWO OF THE THREE SHEATHS CAME OUT EASILY. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22183007 ↗InjuryNov 27, 2024Difficult to Remove; Stretched; Unraveled Material; Detachment of Device or Device Component; Difficult to Remove; Stretched; Unraveled Material; Detachment of Device or Device Component
AS REPORTED, A 7F 11CM AVANTI PLUS CATHETER SHEATH INTRODUCER (CSI) WAS INSERTED INTO THE FEMORAL VEIN TO PERFORM AN ATRIAL FIBRILLATION (AF) ABLATION. THE CSI WAS PREPARED AND DE-AIRED AS PER INSTRUCTIONS FOR USE (IFU) USING HEPARINIZED SALINE, AND IT WAS INSERTED USING ULTRASOUND GUIDANCE BY AN EXPERIENCED OPERATOR. THERE WAS NO RESISTANCE DURING INSERTION. UPON SUCCESSFUL COMPLETION OF THE AF ABLATION THE 7F 11CM AVANTI PLUS CSI WAS REMOVED. INITIALLY, THERE WAS NO RESISTANCE WHILE REMOVING THE SHEATH, BUT A SMALL AMOUNT OF RESISTANCE WAS NOTED TOWARDS THE END OF THE REMOVAL AND THE USERS NOTICED THE SHEATH HAD STRETCHED AND THE DISTAL END OF THE SHEATH HAD SPLIT. A PORTION OF THE DISTAL TIP WAS RETAINED IN THE PATIENT WHICH WAS CONFIRMED VIA COMPUTED TOMOGRAPHY (CT). A SURGICAL CUTDOWN TO RETRIEVE THE TIP WAS PERFORMED THE SAME DAY AFTER CONFIRMING THE SHEATH¿S TIP POSITION ON CT, BUT IT WAS UNSUCCESSFUL. NO EVIDENCE OF THE SEPARATED PORTION OF THE SHEATH WAS FOUND. FOLLOWING THE UNSUCCESSFUL CUTDOWN PROCEDURE, THE PATIENT WAS TRANSFERRED TO A TERTIARY CENTER FOR REMOVAL, WHERE THE TIP, WHICH HAD MIGRATED TO THE PULMONARY VEIN, WAS REMOVED USING AN UNKNOWN SNARE DURING A SEPARA
Patient problems: No Clinical Signs, Symptoms or Conditions; Thrombosis/Thrombus
MAUDE report 20796273 ↗MalfunctionJun 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE ENTIRE HUB OF A 5F 11CM AVANTI PLUS CATHETER SHEATH INTRODUCER (CSI) SNAPPED OFF THE CANNULA OF THE SHEATH WHILE BEING USED FOR INSERTION IN THE FEMORAL ARTERY. A NEW 5F AVANTI PLUS CSI WAS USED AS A REPLACEMENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY EXPERIENCED DURING THE INSERTION PROCESS. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDENDUM: THE DILATOR FOR THE 5F 11CM AVANTI PLUS CSI WAS RETURNED WITH A SEPARATED HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19632735 ↗MalfunctionMay 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, AFTER INSERTION, THE HUB ON A 5F AVANTI SHEATH INTRODUCER POPPED OFF, AND WHILE PUTTING THE CAP BACK ON, THE CAP SPLIT. THERE WERE NO REPORTS OF PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19382429 ↗InjuryMar 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THREE 5F AVANTI + SHEATH INTRODUCERS, IN A ROW, WERE FOUND TO BE VERY BRITTLE AND SNAPPED OFF INSIDE THE PATIENT. THE HUB BROKE OF THE DILATOR OF A 5F AVANTI + SHEATH INTRODUCER. IT WAS THEN THROWN AWAY. A NEW 5F AVANTI SHEATH WAS USED AND PROCEEDED WITH THE CASE, HOWEVER, WHEN THE PHYSICIAN WENT TO REMOVE THE DILATOR FROM THE SHEATH, THE HUB BROKE OFF. THEY WERE ABLE TO INSERT A WIRE AND REMOVE THE SHEATH AND DILATOR ENTIRELY. AN ADDITIONAL SHEATH WAS OPENED TO CHECK INTEGRITY AND IT ALSO SNAPPED AT THE DILATOR. THERE WERE NO REPORTS OF PATIENT INJURY. THREE CATHETERS WERE AT THE TABLE BUT ONLY ONE WAS INSERTED IN PATIENT. ALL 5 DEVICES FROM LOT WERE PULLED. THIS DEVICE WAS USED FOR AN EP ABLATION. THE DEVICE WAS FLUSHED PRIOR TO USE. THE VESSEL DILATOR WAS SNAPPED INTO PLACE AT THE HUB. THERE WERE NO DIFFICULTIES ENCOUNTERED WHILE INSERTING THE DILATOR. HOWEVER, THERE WAS DIFFICULTIES WHILE REMOVING THE DILATOR. WHEN THE HUB OF THE VESSEL DILATOR OF THE SHEATH BROKE OFF IN THE PATIENT, THE PHYSICIAN WAS ABLE TO ADVANCE A WIRE AND REMOVE IN WHOLE. THERE WERE NO DIFFICULTIES ENCOUNTERED WHILE REMOVING DEVICES THROUGH THE SHEATH. THERE WAS NO EXCESS FORCE USED WITH THE
Patient problems: Insufficient Information
MAUDE report 18959204 ↗InjuryMar 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THREE 5F AVANTI + SHEATH INTRODUCERS, IN A ROW, WERE FOUND TO BE VERY BRITTLE AND SNAPPED OFF INSIDE THE PATIENT. THE HUB BROKE OF THE DILATOR OF A 5F AVANTI + SHEATH INTRODUCER. IT WAS THEN THROWN AWAY. A NEW 5F AVANTI SHEATH WAS USED AND PROCEEDED WITH THE CASE, HOWEVER, WHEN THE PHYSICIAN WENT TO REMOVE THE DILATOR FROM THE SHEATH, THE HUB BROKE OFF. THEY WERE ABLE TO INSERT A WIRE AND REMOVE THE SHEATH AND DILATOR ENTIRELY. AN ADDITIONAL SHEATH WAS OPENED TO CHECK INTEGRITY AND IT ALSO SNAPPED AT THE DILATOR. THERE WERE NO REPORTS OF PATIENT INJURY. THREE CATHETERS WERE AT THE TABLE BUT ONLY ONE WAS INSERTED IN PATIENT. ALL 5 DEVICES FROM LOT WERE PULLED. THIS DEVICE WAS USED FOR AN EP ABLATION. THE DEVICE WAS FLUSHED PRIOR TO USE. THE VESSEL DILATOR WAS SNAPPED INTO PLACE AT THE HUB. THERE WERE NO DIFFICULTIES ENCOUNTERED WHILE INSERTING THE DILATOR. HOWEVER, THERE WAS DIFFICULTIES WHILE REMOVING THE DILATOR. WHEN THE HUB OF THE VESSEL DILATOR OF THE SHEATH BROKE OFF IN THE PATIENT, THE PHYSICIAN WAS ABLE TO ADVANCE A WIRE AND REMOVE IN WHOLE. THERE WERE NO DIFFICULTIES ENCOUNTERED WHILE REMOVING DEVICES THROUGH THE SHEATH. THERE WAS NO EXCESS FORCE USED WITH THE
Patient problems: Insufficient Information
MAUDE report 18958976 ↗
Other Cordis US Corp. products
- MYNXGRIP2 models · 2,276 reports · 2 recalls
- Cordis ExoSeal Vascular Closure Device3 models · 1,279 reports
- MYNX CONTROL2 models · 534 reports
- POWERFLEX56 models · 306 reports
- SABER51 models · 270 reports
- Cordis S.M.A.R.T. Control Vascular Stent System30 models · 99 reports · 2 recalls
- INFINITI20 models · 107 reports
- AVIATOR19 models · 75 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page